Multiplex Adaptive Sampling (MAS_9238_v114_revA_11Nov2026)
PromethION: Protocol
Multiplex Adaptive Sampling V MAS_9238_v114_revA_11Nov2026
This protocol describes library preparation and Adaptive Sampling sequencing for human genomic DNA, from extracted gDNA to basecalled, barcode-split BAM files.
The protocol:
- Uses high molecular weight gDNA extracted from human whole blood
- Pools exactly three barcoded samples per PromethION Flow Cell
- Is intended mainly for germline applications
- Enriches for the targets in a BED file, either exported from the Digital Panel Designer for a custom panel or downloaded for a ready-to-use panel
- Requires no PCR, so native methylation is retained
- Is compatible with R10.4.1 flow cells on PromethION 24 and PromethION 2 Integrated
For Research Use Only
FOR RESEARCH USE ONLY.
Contents
Introduction to the protocol
- 1. Overview of the protocol
- 2. Leveraging ready-to-use digital panels
- 3. Overview of steps in the workflow
- 4. Before getting started
- 5. Equipment and consumables
Sample preparation
Library preparation
- 8. DNA repair and end-prep
- 9. Native barcode ligation
- 10. Sequencing adapter ligation and clean-up
- 11. Priming and loading the PromethION Flow Cell
Sequencing and data analysis
Troubleshooting
Overview
This protocol describes library preparation and Adaptive Sampling sequencing for human genomic DNA, from extracted gDNA to basecalled, barcode-split BAM files.
The protocol:
- Uses high molecular weight gDNA extracted from human whole blood
- Pools exactly three barcoded samples per PromethION Flow Cell
- Is intended mainly for germline applications
- Enriches for the targets in a BED file, either exported from the Digital Panel Designer for a custom panel or downloaded for a ready-to-use panel
- Requires no PCR, so native methylation is retained
- Is compatible with R10.4.1 flow cells on PromethION 24 and PromethION 2 Integrated
For Research Use Only
1. Overview of the protocol
Introduction to the Multiplex Adaptive Sampling protocol
This protocol describes the process for library preparation and Adaptive Sampling sequencing for three samples per PromethION Flow Cell, using the Native Barcoding Kit and the Adaptive Sampling target enrichment function in MinKNOW. All Oxford Nanopore Technologies supplied reagents required for this protocol are included in the Multiplex Adaptive Sampling bundles available in 18-sample or 72-sample formats. These bundles include the Native Barcoding Kit and all necessary sequencing components, and is the standard configuration for implementing this workflow.
You can use an Oxford Nanopore ready-to-use, research-use-only digital panel as your BED-file-defined panel, or create a custom digital panel using the Adaptive Sampling, Digital Panel Designer. If you would like to create a custom panel, refer to the Adaptive Sampling technical guide - Digital Panel Designer for more information.
Data analysis is not part of this protocol. The Data outputs and analysis (section 13) describes which exported BED file to use, and the workflow for each panel: wf-hereditary-cancer for the Hereditary Cancer Panel and wf-human-variation for custom germline panels.
Intended use and expected performance
The Multiplex Adaptive Sampling protocol is intended mainly for germline applications, in which each sample needs moderate coverage across the panel, rather than the depth needed to detect low-frequency variants.
For panels covering 0.1–5% of the genome, we expect each of the three samples on a flow cell to reach a mean coverage of about 30x across the targets, with ~90% of target bases covered at more than 20x. Custom panels differ in size, gene content and sequence context, so the coverage your panel reaches may differ. Coverage also depends on the number of available pores on the flow cell, the fragment length of the library, and how evenly the three samples are represented in the pool.
We recommend running a pilot with your panel and confirming the coverage you need, before processing large numbers of samples.
2. Leveraging ready-to-use digital panels
Oxford Nanopore digital panel specification
Ready-to-use digital panels:
| Hereditary Cancer | Pan-Caner | Pan-Haematological Cancer | |
|---|---|---|---|
| Enriched genes | 258 inherited cancer risk genes | 723 cancer associated genes | 144 haematological malignancies associated genes |
| Validated Adaptive Sampling protocols | Multiplex Adaptive Sampling | Ultra-rapid singleplex Adaptive Sampling Singleplex Adaptive Sampling | Singleplex Adaptive Sampling |
| Input amount† | 1 µg | Ultra-rapid singleplex: 1 µg Singleplex: 500 ng to 1 µg* | 1 µg |
| Input sample type | Blood | Solid tumour tissue | Blood Bone marrow |
| Recommended extraction kits | Genomic DNA extraction from human blood using the Qiagen Puregene kit‡ | 1. QIAamp Fast DNA Tissue Kit (Qiagen, 51404) Recommended for Ultra-rapid singleplex Adaptive Sampling 2. Puregene Tissue Kit (Qiagen, 158063) 3. MagAttract HMW DNA Kit (Qiagen, 67563) | 1. Genomic DNA extraction from human blood using the Qiagen Puregene kit‡ 2. QIAamp DSP DNA Mini Kit (Qiagen, 61304) 3. GenFind V3 Genomic DNA Extraction Kit (Beckman Coulter, C34880) |
| QC info (Femto Pulse) | Extracted gDNA samples may be assessed on the Femto Pulse System to visualise the unsheared/sheared gDNA profiles. For example profiles refer to the Input DNA section. | Extracted gDNA samples may be assessed on the Femto Pulse System to visualise the unsheared/sheared gDNA profiles. For example profiles refer to the DNA QC section. For the Ultra-rapid singleplex AS protocol, perform gDNA quality control assessment after completion of the sequencing run using aliquots reserved during the relevant sample preparation steps. | Extracted gDNA samples may be assessed on the Femto Pulse System to visualise the unsheared/sheared gDNA profiles. For example profiles refer to the DNA QC section. |
| Shearing conditions (speed and time) | Three centrifugation steps at ~3,381 RCF for 1 minute each. Refer to the corresponding protocol step for further details. | Two centrifugation steps at ~3,381 RCF for 30 seconds each. Refer to the corresponding protocol step for further details. | Two centrifugation steps at ~4,868 RCF for 1 minute each. Refer to the corresponding protocol step for further details. |
| Reference genomes | GRCh38_GIABv3 | GRCh38_GIABv3 | GCA_000001405.15_GRCh38_no_alt_analysis_set.fna |
| Enrichment BED file (links to the BED files can be found on the Adaptive Sampling page) | ONT_AS_HCP_GRCh38_enrichment_v1.bed | ONT_AS_PanCancer_GRCh38_enrichment_v1.bed | ONT_AS_PanHaem_GRCh38_enrichment_v1.bed |
*A lower input limit of 500 ng can be used; however 1 µg is recommended whenever possible to improve pore occupancy.
†A maximum of 2 µg input can be used to maximise pore occupancy.
‡This Oxford Nanopore adapted extraction method uses the Puregene Blood Kit (Qiagen, 158023). It has gentler mixing conditions and resuspends the final DNA at 50°C rather than 65°C.
3. Overview of steps in the workflow
Prepare for your experiment
You will need to:
- Extract HMW gDNA from each whole blood sample. You can use the following recommended protocol for blood extraction, Genomic DNA extraction from human blood using the Qiagen Puregene kit.
- Design a custom panel in the Digital Panel Designer and export the BED files, or download the BED files for a ready-to-use panel and download the matching reference FASTA file (section 4).
- Ensure you have your sequencing kit, the correct equipment, and third-party reagents.
- Check your flow cell to ensure it has sufficient pores for a good sequencing run.
Sample preparation, library preparation, and sequencing timing overview
| Step in workflow | Process | Time | Stop option |
|---|---|---|---|
| Sample preparation | Extract gDNA from sample and quantify. Fragment the gDNA and quantify. Check the length, quantity and purity of your extracted material. The quality checks performed during the protocol are recommended to ensure experimental success. | Dependent on extraction kit used ~15 minutes for shearing | At this stage, the extracted gDNA or fragmented gDNA can be stored at –20°C for later use. |
| DNA repair and end-prep | Repair the gDNA, and prepare the DNA ends for adapter attachment. | 60 minutes | 4°C overnight |
| Native barcode ligation | Ligate the native barcodes to the DNA ends. | 70 minutes | 4°C overnight |
| Sequencing adapter ligation and clean-up | Ligate sequencing adapters to the DNA ends. | 60 minutes | 4°C overnight |
| Priming and loading the flow cell | Prime the flow cell, and load your DNA library into the flow cell. | 10 minutes | |
| Sequencing | Sequence your library using your sequencing device. | 72 hours | |
| Data analysis | Analyse using one of our EPI2ME workflows. |
Compatibility of this protocol
This protocol should only be used in combination with:
- Multiplex Adaptive Sampling bundles (MAS18 or MAS72)
- Native Barcoding Kit 24 V14 (SQK-NBD114.24)
- PromethION Flow Cell (FLO-PRO114M)
- PromethION 24: PromethION - device and IT specifications
- PromethION 2 Integrated: PromethION 2 Integrated - device and IT specifications
4. Before getting started
Panel BED files
For a successful Adaptive Sampling run, MinKNOW needs two files: an enrichment BED file, which lists the regions to enrich for, and a reference FASTA file for the same genome build. Enrichment BED files can be created in the Digital Panel Designer, or you can start from one of our ready-to-use panels. Instructions for using the Digital Panel Designer, including the design checks and buffer settings, are given in the Adaptive Sampling technical guide.
The export produces three BED files and a summary report. The files are named by date and panel name, in the form:
YYYY-MM-DD_panel-name_enrichment.bed
Only the enrichment BED file is loaded into MinKNOW.
| File | Contents | Used in |
|---|---|---|
| *_enrichment.bed | Two rows per target, one for the plus strand and one for the minus strand, each with buffer at the 5′ end. | MinKNOW in the Adaptive Sampling BED field (section 12). |
| *_analysis.bed | One row per target, with no buffer. | Coverage and variant calling over the targets (section 13). |
| *_exons.bed | One row per exon of the MANE Select transcript of each gene, or of the Ensembl canonical transcript where there is none. | Exon-level coverage analysis (section 13). |
| summary.html | Panel statistics, design checks, the target table and a genome browser view of the targets and buffers. | Your record of the panel design. |
Reference genome
MinKNOW needs a reference FASTA file to align reads and make Adaptive Sampling decisions. The reference must be the same genome build that was selected in the Digital Panel Designer when the panel was exported. We recommend the following files:
| Panel | Build | Reference FASTA file for MinKNOW | Source |
|---|---|---|---|
| Custom panel | GRCh38 | GCA_ 000001405. 15_ GRCh38_ no_ alt_ analysis_ set. fna. gz | NCBI |
| Custom panel | GRCh37 | hg19. fa. gz | UCSC |
| Hereditary Cancer Panel | GRCh38 | GRCh38_ GIABv3_ no_ alt_ analysis_ set_ maskedGRC_ decoys_ MAP2K3_ KMT2C_ KCNJ18. fasta. gz | Genome in a Bottle |
The chromosome names in the reference FASTA file must match the BED file.
The BED files exported from the Digital Panel Designer use UCSC-style chromosome names (chr1, chr2, chrX), and all three recommended reference files use the same names.
If you use a different reference file that names chromosomes without the "chr" prefix (1, 2, X), MinKNOW will not find any of the targets. The run will then proceed without enrichment and will look like a low-coverage whole genome run.
For the Hereditary Cancer Panel, we recommend the Genome in a Bottle (GIAB) v3 GRCh38 reference, which was used to establish the performance of the panel. It is the GRCh38 no-alt analysis set with falsely duplicated regions of the assembly masked, so it uses the same coordinates, and the panel BED files exported from the Digital Panel Designer can be used with it unchanged.
The GRCh38 files above contain no alternate locus contigs. If a reference that includes alternate contigs is used instead, reads from regions with alternate haplotypes, such as the HLA genes, can be split between the primary assembly and the alternate contigs. The hg19 file contains a small number of haplotype contigs in the MHC region, and the same applies to panels with targets in that region.
If you use another reference file whose chromosome names differ from the BED files, rename the chromosomes in the BED files to match the reference. Do not rename the reference. We do not advise pre-indexing the reference before uploading it, as MinKNOW indexes it automatically at the start of the run. Reference requirements are described in more detail in the Adaptive Sampling technical guide.
5. Equipment and consumables
Materials
- (FOR LIBRARY PREPARATION) ~1 µg of sheared gDNA
- Multiplex Adaptive Sampling bundles (containing the materials listed below)
- Native Barcoding Kit 24 V14 (SQK-NBD114.24)
- LFB Expansion (EXP-LFB001)
- SFB Expansion (EXP-SFB001) if using the 72 sample bundle
Consumables
- PromethION Flow Cell (FLO-PRO114M)
- g-TUBE™ (Covaris, 520079)
- NEBNext® FFPE Repair Mix (NEB, M6630)
- NEBNext® Ultra II End repair/dA-tailing Module (NEB, E7546)
- NEB Blunt/TA Ligase Master Mix (NEB, M0367)
- NEBNext® Quick Ligation Module (NEB, E6056)
- Nuclease-free water (e.g. Thermo Scientific, AM9937)
- Freshly prepared 80% ethanol in nuclease-free water
- Qubit™ dsDNA BR Assay Kit (Invitrogen, Q32850)
- Qubit™ dsDNA HS Assay Kit (Invitrogen, Q32851)
- Qubit™ Assay Tubes (Invitrogen, Q32856)
- 1.5 ml Eppendorf DNA LoBind tubes
- 0.2 ml thin-walled PCR tubes
- Genomic DNA 165 kb Analysis Kit (Agilent, FP-1002-0275) (optional)
Equipment
- PromethION device
- PromethION Flow Cell Light Shield
- Rotator mixer (e.g. Hula mixer)
- Magnetic separation rack, suitable for 1.5 ml Eppendorf tubes
- Microfuge
- Vortex mixer
- Thermal cycler with heated lid
- Heat block
- Wide-bore pipette tips
- P1000 pipette and tips
- P200 pipette and tips
- P100 pipette and tips
- P20 pipette and tips
- P10 pipette and tips
- P2 pipette and tips
- Multichannel pipette and tips
- Ice bucket with ice
- Timer
- Qubit™ fluorometer (or equivalent for QC check)
Optional equipment
- Femto Pulse System (Agilent Technologies) or equivalent
Oxford Nanopore Technologies Multiplex Adaptive Sampling bundles
All Oxford Nanopore Technologies reagents required for this protocol are included in the Multiplex Adaptive Sampling bundles, available in the Nanopore Store in 18-sample or 72-sample formats.
These bundles include the Native Barcoding Kit and all necessary sequencing components, and are the standard configuration for implementing this workflow.
For more information on specific kit components found within these bundles, please refer to the relevant products found in the Nanopore Store page:
The above list of materials, consumables, and equipment is for the shearing of the DNA samples, as well as library preparation.
The consumables, equipment and instructions for the recommended DNA extraction are provided in the protocol Genomic DNA extraction from human blood using the Qiagen Puregene kit.
For this protocol, the following input is required for library preparation:
- ~1 µg of sheared gDNA
Input DNA
This protocol has been validated with gDNA extracted from fresh human whole blood collected in EDTA K2 tubes, using the Puregene Blood Kit (Qiagen, 158023) from 1 ml of blood with an Oxford Nanopore adapted method. The adapted method uses gentler mixing and resuspends the final DNA at 50°C rather than 65°C. It is described in the Genomic DNA extraction from human blood using the Qiagen Puregene kit protocol.
Other extraction kits have not been tested with this protocol. gDNA from whole blood extracted with another kit may perform well, if it meets the requirements below, but meeting them does not guarantee the performance, as the fragment length and purity profile of the DNA can affect enrichment and output. If you use another kit, we recommend running a pilot and confirming the coverage you need before processing large numbers of samples.
| Requirement | Value | How to check |
|---|---|---|
| Quantity | >1 µg per sample, at a concentration of at least 20 ng/µl, so that 1 µg fits in the 50 µl g-TUBE input | Qubit dsDNA BR Assay, in triplicate |
| Purity | OD 260/280 ≥1.8 and OD 260/230 of 2.0–2.2 | NanoDrop or equivalent |
| Fragment length | HMW gDNA Figure 1 shows a typical Puregene extraction. | Femto Pulse or equivalent (optional) |
Figure 1: Example fragment length profile of gDNA extracted from human blood using the adapted Puregene method.
The g-TUBE shears DNA to a target length but cannot lengthen it. gDNA that is already shorter than expected after extraction will be shorter after shearing, and read length will be lower as a result.
How to QC your input DNA
It is important that the input DNA meets the quantity and quality requirements. Using too little or too much DNA, or DNA of poor quality (e.g. highly fragmented or containing RNA or chemical contaminants) can affect your library preparation.
For instructions on how to perform quality control of your DNA sample, please read the Input DNA/RNA QC protocol.
Chemical contaminants
Depending on how the DNA is extracted from the raw sample, certain chemical contaminants may remain in the purified DNA, which can affect library preparation efficiency and sequencing quality. Read more about contaminants on the Contaminants page of the Community.
Eppendorf tube orientation in centrifuge
For all centrifugation steps, ensure that tubes are loaded into the centrifuge with the hinge side of the tube facing outwards. This will assist in visual identification of the pellet.
Ensure gentle handling when removing the tubes from the centrifuge to avoid dislodging the pellet.

Third-party reagents
We have validated and recommend the use of all the third-party reagents used in this protocol. Alternatives have not been tested by Oxford Nanopore Technologies.
For all third-party reagents, we recommend following the manufacturer's instructions to prepare the reagents for use, unless specified in this protocol.
Check your flow cell
We highly recommend that you check the number of pores in your flow cell prior to starting a sequencing experiment, within the warranty period. Flow cells should be removed from refrigeration and their packaging and allowed to come to room temperature for 30 minutes before they are used.
To perform the flow cell check, please follow the instructions in the Flow Cell Check document. As part of your flow cell check, you will see the number of pores in your new flow cell. The warranty on your flow cell covers a minimum pore count of 5,000 pores for a PromethION Flow Cell. You can find more information about the warranty period, stability and storage conditions of flow cells on the Oxford Nanopore Store. If you have any concerns or questions, please contact Oxford Nanopore.
Oxford Nanopore will replace any unused PromethION Flow Cell with fewer than 5,000 pores, when the result is reported within two days of performing the flow cell check, and when the storage recommendations have been followed.
The Native Adapter (NA) used in this kit and protocol is not interchangeable with other sequencing adapters.
6. Guidance on recommended DNA extraction protocol
Prepare your sample(s) for DNA extraction.
A recommended protocol for gDNA extraction from blood is provided below:
Genomic DNA extraction from human blood using the Qiagen Puregene kit
Take forward your extracted gDNA samples into the next step, DNA shearing.
Alternatively, your sample(s) can be stored at 4°C overnight or at –20°C for later use.
7. DNA shearing using g-TUBE
Materials
- >1 µg of extracted gDNA for each sample (from previous step)
Consumables
- g-TUBE (Covaris, 520079)
- Nuclease-free water (e.g. ThermoFisher, AM9937)
- Qubit™ dsDNA BR Assay Kit (Invitrogen, Q32850)
- Qubit™ Assay Tubes (ThermoFisher, Q32856)
- 1.5 ml Eppendorf DNA LoBind tubes
- Genomic DNA 165 kb Analysis Kit (Agilent, FP-1002-0275) (optional)
Equipment
- Eppendorf 5424 centrifuge (or equivalent)
- Microfuge
- Qubit™ fluorometer (or equivalent for QC check)
- Ice bucket with ice
- Timer
- Wide-bore pipette tips
- P1000 pipette and tips
- P200 pipette and tips
- P100 pipette and tips
- P20 pipette and tips
- P10 pipette and tips
- P2 pipette and tips
Optional equipment
- Femto Pulse System (Agilent Technologies) or equivalent
Prepare the DNA in nuclease-free water:
1. Ensure you have 1 µg of extracted gDNA from sample extraction, and transfer this into a 1.5 ml Eppendorf tube.
2. Adjust the volume to 50 μl with nuclease-free water.
3. Mix thoroughly by pipetting up and down using a wide-bore pipette tip.
4. Spin down briefly in a microfuge.
Transfer 50 μl of the sample into the top of the g-TUBE and screw the cap firmly.
Load the sample tube into the centrifuge, ensuring the instrument is appropriately balanced according to the manufacturer's instructions.
Centrifuge your sample as follows:
Note: the speeds may vary depending on the type of centrifuge being used. Please refer to the manufacturer’s manual for further details of your equipment.
1. Centrifuge at 3,381 RCF (6,000 RPM) for 60 seconds.
2. Visually inspect to confirm the entire sample has passed through the upper chamber to the lower chamber of the g-TUBE. If the sample has not passed through, then repeat the step until the sample has fully passed into the lower chamber.
3. Remove the tube from the centrifuge, invert the g-TUBE and place it back in the centrifuge (ensuring the volume of sample is in the upper chamber.
4. Spin the g-TUBE again at the same speed and duration as above: 3,381 RCF (6,000 RPM) for 60 seconds.
5. Visually inspect to confirm the entire sample has passed through the upper chamber to the lower chamber of the g-TUBE. If the sample has not passed through, then repeat the step until the sample has fully passed into the lower chamber.
Repeat the previous step two more times, for a total of 3 passes through the g-TUBE in each direction.
With the sample volume in the lower chamber, unscrew the g-TUBE body, leaving the screw cap containing the sample volume.
Retrieve the sample from the g-TUBE screw cap and transfer it into a clean 1.5 ml Eppendorf tube.
Quantify your sample using the Qubit dsDNA BR Assay Kit.
Approximately 1 µg of gDNA is expected following shearing.
The Qubit measurements should be ~18–20 ng/μl.
The sheared gDNA can also be assessed for fragment size using Femto Pulse (Agilent).
The expected sheared length should be ~7.5–8.5 kb.
Take forward 48 µl of sheared gDNA into the library preparation section of this protocol. Alternatively, your sample(s) can be stored at 4°C overnight.
8. DNA repair and end-prep
Materials
- 1 µg of sheared gDNA in 48 µl for each sample (from previous step)
Consumables
- NEBNext® FFPE DNA Repair Mix (NEB, M6630)
- NEBNext® Ultra™ II End Repair/dA-Tailing Module (NEB, E7546)
- Nuclease-free water (e.g. Thermo Scientific, AM9937)
- Freshly prepared 80% ethanol in nuclease-free water
- Qubit™ dsDNA BR Assay Kit (Invitrogen, Q32850)
- Qubit™ Assay Tubes (Invitrogen, Q32856)
- 0.2 ml thin-walled PCR tubes
- 1.5 ml Eppendorf DNA LoBind tubes
Equipment
- Microfuge
- Vortex mixer
- Thermal cycler with heated lid
- Rotator mixer (e.g. Hula mixer)
- Magnetic separation rack, suitable for 1.5 ml Eppendorf tubes
- Qubit™ fluorometer (or equivalent for QC check)
- Ice bucket with ice
- P1000 pipette and tips
- P200 pipette and tips
- P100 pipette and tips
- P20 pipette and tips
- P10 pipette and tips
- P2 pipette and tips
Prepare the NEBNext FFPE DNA Repair Mix and NEBNext Ultra II End Repair / dA-tailing Module reagents in accordance with the manufacturer’s instructions and place on ice.
For optimal performance, NEB recommend the following:
Thaw all reagents on ice.
Flick and/or invert the reagent tubes to ensure they are well mixed.
Note: Do not vortex the FFPE DNA Repair Mix or Ultra II End Prep Enzyme Mix.Always spin down tubes before opening for the first time each day.
The Ultra II End Prep Reaction Buffer and FFPE DNA Repair Buffer may have precipitate. Allow the mixtures to come to room temperature and pipette the buffers up and down several times to break up the precipitate, followed by vortexing the tubes for 30 seconds to solubilise any precipitate.
Note: It is important the buffers are mixed well by vortexing.The FFPE DNA Repair Buffer may have a yellow tinge.
Do not vortex the NEBNext FFPE DNA Repair Mix or NEBNext Ultra II End Prep Enzyme Mix.
It is important that the NEBNext FFPE DNA Repair Buffer and NEBNext Ultra II End Prep Reaction Buffer are mixed well by vortexing.
Check for any visible precipitate; vortexing for at least 30 seconds may be required to solubilise any precipitate.
Prepare your DNA sample(s):
1. For each sample, ensure you have 48 µl (with ~1 µg) sample in a separate, clean 0.2 ml thin-walled PCR tube.
2. Mix each sample thoroughly by pipetting up and down, or by flicking the tube. Spin down briefly in a microfuge.
Combine the following components per tube:
| Reagent | Volume |
|---|---|
| DNA sample | 48 µl |
| Ultra II End-prep Reaction Buffer | 3.5 µl |
| Ultra II End-prep Enzyme Mix | 3 µl |
| NEBNext FFPE DNA Repair Buffer | 3.5 µl |
| NEBNext FFPE DNA Repair Mix | 2 µl |
| Total | 60 µl |
Ensure the components are thoroughly mixed by pipetting and spin down briefly.
Using a thermal cycler with a heated lid, incubate the reaction(s) at 20°C for 5 minutes, 65°C for 5 minutes, and hold at 4°C.
Remove the reaction from the thermal cycler and place the tubes on ice.
At this stage, you can spin down and transfer each sample reaction to a separate, clean 1.5 ml Eppendorf DNA LoBind tube for ease of handling.
Resuspend the AMPure XP Beads (AXP) by vortexing.
Add 1X volume (60 µl) of resuspended AMPure XP Beads (AXP) to each end-prep reaction and mix by flicking the tube.
Incubate the samples on a rotator mixer (e.g. Hula mixer) for 10 minutes at room temperature.
Freshly prepare 2 ml of 80% ethanol in nuclease-free water (this is sufficient to process three samples).
Prepare sufficient 80% ethanol for all your samples.
Spin down each sample and pellet on a magnetic rack for 10 minutes until the supernatant is clear and colourless. Keep the tube(s) on the magnetic rack and pipette off the supernatant.
Keep the tube(s) on the magnetic rack and wash the beads with 200 µl of freshly prepared 80% ethanol without disturbing the pellet. Remove the supernatant using a pipette and discard.
Repeat the previous step.
Spin down and place the tube back on the magnetic rack. Pipette off any residual ethanol. Allow the pellet to dry for ~30 seconds, but do not dry the pellet to the point of cracking.
For each sample, remove the tube from the magnetic rack and resuspend the pellet in 20 µl nuclease-free water. Incubate for 5 minutes at room temperature.
Pellet the beads on the magnetic rack for at least 1 minute, until the eluate is clear and colourless.
For each sample, remove and retain 20 µl of eluate into a separate, clean 1.5 ml Eppendorf DNA LoBind tube.
Make sure your samples are processed separately. At this stage, they are not yet barcoded.
Quantify 1 µl of each eluted sample using a Qubit fluorometer.
We recommend performing multiple (triplicate) Qubit readings of each of the 3 samples (3 samples per flow cell) to quantify them more accurately. This will be essential for further normalising of each sample before barcoding.
Note: You should expect to recover between 500–700 ng per sample after end-prep.
Using your quantification results, normalise your samples to the sample with the lowest yield as follows:
1. Take forward 16 μl of your lowest yield sample into a clean 1.5 ml Eppendorf DNA LoBind tube.
2. Take forward an equivalent mass of each of the other samples into separate, clean 1.5 ml Eppendorf DNA LoBind tubes.
3. Adjust the volume of each of the samples to 16 μl using nuclease-free water.
Take forward your end-prepped DNA samples into the native barcode ligation step.
Optional: At this point, you can also store the samples at 4°C overnight.
9. Native barcode ligation
Materials
- Native Barcodes (NB01-24)
- AMPure XP Beads (AXP)
- EDTA (EDTA)
- Short Fragment Buffer (SFB)
Consumables
- NEB Blunt/TA Ligase Master Mix (NEB, M0367)
- Nuclease-free water (e.g. Thermo Scientific, AM9937)
- Qubit™ dsDNA HS Assay Kit (Invitrogen, Q32851)
- Qubit™ Assay Tubes (Invitrogen, Q32856)
- 1.5 ml Eppendorf DNA LoBind tubes
Equipment
- Microfuge
- Vortex mixer
- Rotator mixer (e.g. Hula mixer)
- Magnetic separation rack, suitable for 1.5 ml Eppendorf tubes
- Heat block
- Qubit™ fluorometer (or equivalent for QC check)
- Ice bucket with ice
- Multichannel pipette and tips
- P1000 pipette and tips
- P200 pipette and tips
- P100 pipette and tips
- P20 pipette and tips
- P10 pipette and tips
- P2 pipette and tips
Briefly spin down the NEB Blunt/TA Ligase Master Mix and pipette mix. Then place the reagent on ice.
Thaw the AMPure XP Beads (AXP) at room temperature and mix by vortexing. Keep the beads at room temperature.
Thaw the EDTA at room temperature and mix by vortexing. Then spin down and place on ice.
Thaw the Short Fragment Buffer (SFB) at room temperature and mix by vortexing. Place on ice.
Thaw the Native Barcodes (NB01-24) at room temperature. Individually mix the barcodes by pipetting, spin down, and place them on ice.
The wells of the barcoding plate are intended for single use only. Make sure your barcode well is sealed before use. Do not reuse the barcode wells.
Select a unique barcode for every sample to be run together on the same flow cell. Up to 3 samples should be barcoded and combined onto one PromethION Flow Cell.
Only use one barcode per sample.
Into each 1.5 ml Eppendorf tube containing your normalised sample inputs, add the reagents in the following order below, mixing well by pipetting between each addition:
| Reagent | Volume |
|---|---|
| End-prepped DNA | 16 µl |
| Native Barcode (NB01-24) | 4 µl |
| Blunt/TA Ligase Master Mix | 20 µl |
| Total | 40 µl |
Thoroughly mix the reaction by gently pipetting and briefly spinning down.
Incubate for 20 minutes at room temperature.
Add 4 µl of EDTA to each sample tube. Mix thoroughly by pipetting and spin down briefly.
EDTA is added at this step to stop the reaction.
Pool the barcoded samples (up to three samples at a time) in a 1.5 ml Eppendorf DNA LoBind tube.
Resuspend the AMPure XP Beads (AXP) by vortexing.
Add 0.65X volume of AMPure XP Beads (AXP) to the pooled reaction, and mix by pipetting.
For 3 samples pooled (132 µl), the volume of AXP required will be 86 µl.
Incubate the samples on a rotator mixer (e.g. Hula mixer) for 10 minutes at room temperature.
The next clean-up step uses Short Fragment Buffer (SFB) rather than 80% ethanol to wash the beads.
Using 80% ethanol may result in sub-optimal results.
Spin down the sample and pellet on a magnetic rack for 5 minutes. Keep the tube on the magnetic rack until the eluate is clear and colourless. Then pipette off the supernatant.
Wash the beads with 500 µl of Short Fragment Buffer (SFB). Flick the beads to resuspend, spin down , then return the sample to the magnetic rack and allow the beads to pellet. Remove the buffer using a pipette and discard.
Repeat the previous step.
Spin down and place the tube back on the magnetic rack. Pipette off any residual buffer.
Remove the tube from the magnetic rack and resuspend the pellet in 32 µl nuclease-free water by gently flicking the tube.
Incubate for 10 minutes at 37°C. Every 2 minutes, agitate the sample by gently flicking for 10 seconds to encourage DNA elution.
Pellet the beads on the magnetic rack until the eluate is clear and colourless.
Remove and retain 32 µl of eluate into a clean 1.5 ml Eppendorf DNA LoBind tube.
Quantify 1 µl of eluted sample using a Qubit fluorometer.
You should expect to recover between 600–800 ng DNA after native barcoding.
Take forward the barcoded DNA library to the adapter ligation and clean-up step. However, you may store the sample at 4°C overnight.
10. Sequencing adapter ligation and clean-up
Materials
- Long Fragment Buffer (LFB)
- Elution Buffer (EB)
- Native Adapter (NA)
- AMPure XP Beads (AXP)
Consumables
- NEBNext® Quick Ligation Module (NEB, E6056)
- Qubit™ dsDNA HS Assay Kit (Invitrogen, Q32851)
- Qubit™ Assay Tubes (Invitrogen, Q32856)
- 1.5 ml Eppendorf DNA LoBind tubes
Equipment
- Microfuge
- Vortex mixer
- Rotator mixer (e.g. Hula mixer)
- Magnetic separation rack, suitable for 1.5 ml Eppendorf tubes
- Heat block
- Ice bucket with ice
- P1000 pipette and tips
- P200 pipette and tips
- P100 pipette and tips
- P20 pipette and tips
- P10 pipette and tips
- Qubit™ fluorometer (or equivalent for QC check)
The Native Adapter (NA) used in this kit and protocol is not interchangeable with other sequencing adapters.
Check your flow cell
We recommend performing a flow cell check before starting adapter ligation to ensure you have a flow cell with sufficient pores for a good sequencing run.
Flow cells should be removed from refrigeration and their packaging and allowed to come to room temperature for 30 minutes before they are used.
Condensation can form on the flow cell in humid environments. Inspect the gold connector pins on the top and underside of the flow cell for condensation and wipe off with a lint-free wipe if any is observed. Ensure the heat pad (dark grey) is present on the underside of the flow cell.
Refer to the flow cell check document for more information.
Prepare the NEBNext Quick Ligation Reaction Module according to the manufacturer's instructions, and place on ice:
Thaw the NEBNext Quick Ligation Reaction Buffer at room temperature.
Spin down the reagent tubes for 5 seconds.
Ensure the reagents are fully mixed by performing 10 full volume pipette mixes.
Place the Quick T4 DNA Ligase on ice. Do NOT vortex the Quick T4 DNA Ligase.
Note: The NEBNext Quick Ligation Reaction Buffer (5x) may have a little precipitate. Allow the mixture to come to room temperature and pipette the buffer up and down several times to break up the precipitate, followed by vortexing the tube for several seconds to ensure the reagent is thoroughly mixed.
Do not vortex the Quick T4 DNA Ligase.
Spin down the Native Adapter (NA), pipette mix and place on ice.
Thaw the Elution Buffer (EB) at room temperature and mix by vortexing. Then spin down and place on ice.
Thaw the Long Fragment Buffer (LFB) at room temperature and mix by vortexing. Then place on ice.
In a 1.5 ml Eppendorf LoBind tube, mix in the following order:
| Reagent | Volume |
|---|---|
| 3-plex pooled barcoded library | 30 µl |
| Native Adapter (NA) | 5 µl |
| NEBNext Quick Ligation Reaction Buffer (5X) | 10 µl |
| Quick T4 DNA Ligase | 5 µl |
| Total | 50 µl |
Thoroughly mix the reaction by gently pipetting and briefly spinning down.
Incubate the reaction for 20 minutes at room temperature.
The next clean-up step uses Long Fragment Buffer (LFB) rather than 80% ethanol to wash the beads. The use of ethanol will be detrimental to the sequencing reaction.
Resuspend the AMPure XP Beads (AXP) by vortexing.
Add 0.5X volume (25 µl) of resuspended AMPure XP Beads (AXP) to the reaction and mix by pipetting.
Incubate on a rotator mixer (e.g. Hula mixer) for 10 minutes at room temperature.
Spin down the sample and pellet on the magnetic rack. Keep the tube on the magnetic rack and pipette off the supernatant.
Wash the beads by adding 250 μl Long Fragment Buffer (LFB). Flick the beads to resuspend, spin down, then return the tube to the magnetic rack and allow the beads to pellet for at least 5 minutes. Remove the supernatant using a pipette and discard.
Note: Take care when removing the supernatant as the viscosity of the buffer can contribute to loss of beads from the pellet.
Repeat the previous step.
Spin down and place the tube back on the magnetic rack. Pipette off any residual supernatant. Allow the pellet to dry for ~30 seconds, but do not dry the pellet to the point of cracking.
Remove the tube from the magnetic rack and resuspend the pellet in 35 µl Elution Buffer (EB).
Spin down and incubate the sample for 10 minutes at 37°C. Every 2 minutes, agitate the sample by gently flicking for 10 seconds to encourage DNA elution.
Pellet the beads on the magnetic rack for at least 1 minute, until the eluate is clear and colourless.
Remove and retain 35 µl of eluate containing the DNA library into a clean 1.5 ml Eppendorf DNA LoBind tube.
Note: Dispose of the pelleted beads
Quantify 1 µl of eluted sample using a Qubit fluorometer.
You should expect to recover between 300–500 ng of DNA library after adapter ligation.
Note: Loading less than 250 ng of DNA library on your flow cell could result in lower pore occupancies and not meeting the required target output.
The prepared library is used for loading onto the flow cell. Store the library on ice or at 4°C until ready to load.
For this method we recommend loading your flow cell(s) as soon as possible following library preparation.
11. Priming and loading the PromethION Flow Cell
Materials
- Flow Cell Flush (FCF)
- Flow Cell Tether (FCT)
- Library Beads (LIB)
- Sequencing Buffer (SB)
Consumables
- PromethION Flow Cell (FLO-PRO114M)
- 1.5 ml Eppendorf DNA LoBind tubes
Equipment
- PromethION device
- PromethION Flow Cell Light Shield
- Vortex mixer
- Microfuge
- Ice bucket with ice
- P1000 pipette and tips
- P200 pipette and tips
- P100 pipette and tips
- P20 pipette and tips
This kit is only compatible with R10.4.1 flow cells (FLO-PRO114M).
We recommend removing your flow cell from refrigeration and its packaging and allowing it to come to room temperature for 30 minutes before it is inserted into the device.
Condensation can form on the flow cell in humid environments. Inspect the gold connector pins on the top and underside of the flow cell for condensation and wipe off with a lint-free wipe if any is observed. Ensure the heat pad (dark grey pad) is present on the underside of the flow cell.
Thaw the Sequencing Buffer (SB), Library Beads (LIB), Flow Cell Tether (FCT) and Flow Cell Flush (FCF) at room temperature before mixing by vortexing. Then spin down and store on ice.
Prepare the flow cell priming mix in a suitable tube for the number of flow cells required. Once combined, mix well by briefly vortexing.
| Reagent | Volume per flow cell |
|---|---|
| Flow Cell Tether (FCT) | 30 µl |
| Flow Cell Flush (FCF) | 1,170 µl |
| Total volume | 1,200 µl |
Load the flow cell(s) into the device docking ports.
Make sure that the dark grey heat pad on the underside of the flow cell is not torn or deformed.
For the PromethION 24:
Line up the flow cell with the connector horizontally and vertically before smoothly inserting into position.
Press down firmly onto the flow cell and ensure the latch engages and clicks into place.


For the PromethION 2 Integrated:
Place the flow cell flat on the metal plate.
Slide the flow cell into the docking port until the gold pins cannot be seen.
Press the flow cell firmly into the loader until a click is heard.

Insertion of the flow cells at the wrong angle can cause damage to the pins on the PromethION and affect your sequencing results. If you find the pins on a PromethION position are damaged, please contact support@nanoporetech.com for assistance.

Complete a flow cell check to assess the number of pores available before loading the library.
This step can be omitted if the flow cell has been checked previously.
Refer to the flow cell check document for more information.
Slide the inlet port cover clockwise to open.

Take care when drawing back buffer from the flow cell. Do not remove more than 20–30 µl, and make sure that the array of pores are covered by buffer at all times. Introducing air bubbles into the array can irreversibly damage pores.
After opening the inlet port, draw back a small volume to remove any air bubbles:
- Set a P1000 pipette tip to 200 µl.
- Insert the tip into the inlet port.
- Turn the wheel until the dial shows 220–230 µl, or until you see a small volume of buffer entering the pipette tip.

Load 500 µl of the priming mix into the flow cell via the inlet port, avoiding the introduction of air bubbles. Wait five minutes. During this time, prepare the library for loading using the next steps in the protocol.

Thoroughly mix the contents of the Library Beads (LIB) by pipetting.
The Library Beads (LIB) tube contains a suspension of beads. These beads settle very quickly. It is vital that they are mixed immediately before use.
We recommend using the Library Beads (LIB) for most sequencing experiments.
In a new 1.5 ml Eppendorf DNA LoBind tube, prepare the library for loading as follows:
| Reagent | Volume per flow cell |
|---|---|
| Sequencing Buffer (SB) | 100 µl |
| Library Beads (LIB) thoroughly mixed before use | 68 µl |
| DNA library | 32 µl |
| Total | 200 µl |
Note: The library loading volume has been increased to improve array coverage.
Complete the flow cell priming by slowly loading 500 µl of the priming mix into the inlet port.

Mix the prepared library gently by pipetting up and down, just prior to loading.
Load 200 µl of library into the inlet port using a P1000 pipette.

Close the valve to seal the inlet port.
For optimal sequencing output, install the light shield on your flow cell as soon as the library has been loaded.
We recommend leaving the light shield on the flow cell when the library is loaded, including during any washing and reloading steps. The shield can be removed when the library has been removed from the flow cell.
If the light shield has been removed from the flow cell, install the light shield as follows:
- Align the inlet port cut out of the light shield with the inlet port cover on the flow cell. The leading edge of the light shield should sit above the flow cell ID.
- Firmly press the light shield around the inlet port cover. The inlet port clip will click into place underneath the inlet port cover.


Close the PromethION lid when ready to start a sequencing run on MinKNOW.
Wait a minimum of 10 minutes after loading the flow cells onto the PromethION, before initiating any experiments. This will help to increase the sequencing output.
12. Data acquisition and basecalling
How to start sequencing
The sequencing device control, data acquisition and real-time basecalling are carried out by MinKNOW.
Refer to the PromethION 24 or PromethION 2 Integrated documentation for detailed instructions on setting up the device and the sequencing run. The same run settings are used on both devices.
Before you start
Before starting the run, ensure that:
The reference FASTA file for your panel (see Reference genome in section 4) and the enrichment BED file are available on the device.
You have enough storage space on the device for each flow cell.
You have decided whether to switch On Basecall on-target reads only, as described below.
Live basecalling and Basecall on-target reads only
We recommend basecalling in real-time with the high accuracy (HAC) model, with BAM as the output format. The Basecall on-target reads only setting in MinKNOW decides which reads are basecalled during the run. The right choice depends on whether you need the reads that Adaptive Sampling rejects.
| Basecall on-target reads only | Reads basecalled during the run | Flow cells at once |
|---|---|---|
| Off | All reads, including the short reads rejected by Adaptive Sampling. The rejected reads give a low-pass view of the whole genome and its methylation alongside the targets. | Fewer, as every read is basecalled |
| On | On-target reads only. Rejected reads are written to a pod5skip folder in the run folder and can be basecalled after the run if they are needed. | All device positions |
The number of flow cells each device can basecall live, with and without Basecall on-target reads only, is given in the Device limits section of the Adaptive Sampling technical guide. If you run more flow cells than your device can basecall live, basecalling falls behind sequencing and enrichment may be reduced.
To run more flow cells at once and still keep the off-target data, switch Basecall on-target reads only On and basecall the rejected reads after the run. This takes additional time on the device, once sequencing has finished. Instructions are given in Basecalling rejected reads after the run, in the Data outputs and analysis section below.
With POD5 output off, the raw signal of basecalled reads is not kept.
If you plan to basecall the rejected reads after the run, switch POD5 output On. Rejected reads skipped by Basecall on-target reads only are kept in the pod5skip folder.
Setting up your sequencing run in MinKNOW
In MinKNOW, select Start sequencing, then enter the flow cell position, experiment name and a sample ID.
The sample ID applies to the whole flow cell, not to each sample; the three samples are identified by their barcode numbers (for example, barcode01, barcode02 and barcode03) in the output folders and files. Record which sample has which barcode.
To name each sample instead, start from a sample sheet, as described below in Optional: Starting from a sample sheet.Apply the settings below. Leave all other settings at their defaults.
| Screen | Setting | Set to |
|---|---|---|
| Kit | Kit selection | Native Barcoding Kit 24 V14 (SQK-NBD114.24) or Native Barcoding Kit 96 V14 (SQK-NBD114.96) |
| Run configuration | Modified bases | On, with 5mC and 5hmC CG contexts selected |
| Alignment | On. Upload the reference FASTA file for your panel: GIAB v3 for the Hereditary Cancer Panel, or the file for your build for a custom panel (refer to Reference genome in section 4). | |
| Adaptive sampling | On, with Enrich selected. Upload the enrichment BED file (*_enrichment.bed) and the same reference FASTA file. | |
| Basecall on-target reads only | Off or On, as described above | |
| Output | FASTQ | Off |
| POD5 | Off, or On if you plan to basecall the rejected reads after the run |
- Start the run.
Optional: Starting from a sample sheet
Starting from a sample sheet allows you to give each of the three samples its own name, which is carried through to the output folders and files in place of the barcode number. Use the same run settings provided in the Table above.
Prepare a CSV file with one row per barcode. It needs the following columns:
- experiment_id (your experiment name, the same in every row)
- kit (SQK-NBD114.24)
- barcode (for example barcode01)
- alias (the sample name for that barcode)
Sample names in the alias column can contain only letters, numbers, hyphens, underscores and full stops, with no spaces, and must be 1–40 characters long. - either position_id (the slot on the device)
- or flow_cell_id (the ID printed on the flow cell)
Save the file with a visible .csv extension, or MinKNOW will not recognise it.
In MinKNOW, select Start from a sample sheet in place of Start sequencing, and select your CSV file. The flow cell position, experiment name and sample names are read from the file, so you do not need to enter them on screen. Then apply the settings provided in the Table above.
Further sample sheet options are described in the sample sheet section of the MinKNOW documentation.
We do not recommend sequencing and performing data analysis at the same time on your device.
To ensure the compute on your device can keep up with the requirements for sequencing and basecalling, we strongly recommend against running both processes at the same time. Make sure your sequencing run has completed before starting data analysis, and do not start a sequencing run while data analysis is running on the device.
13. Data outputs and analysis
Data output
At the end of the run, the data for each of the three samples is in its own barcode folder. Each BAM file contains basecalled reads aligned to the reference FASTA file uploaded in MinKNOW. Modified base calls (5mC and 5hmC in CpG context) are stored in the read records only if modified base calling was turned On, when the run was set up.
With Basecall on-target reads only switched Off, the BAM files also contain the short reads rejected by Adaptive Sampling. With it switched On, rejected reads are not basecalled and are written to the pod5skip folder.
| Output | Description |
|---|---|
| Aligned BAM files, one folder per barcode | Basecalled, aligned reads for each sample, with modified base calls if they were turned On. |
| Sequencing summary and run report | Run-level metrics, including read length, output and Adaptive Sampling decisions. |
Basecalling rejected reads after the run
If Basecall on-target reads only was switched On, the reads rejected by Adaptive Sampling are written to the pod5skip folder in the run folder. They can only be basecalled if POD5 output was switched On when the run was set up. Wait until the sequencing run has finished before basecalling them, as both processes compete for the compute on the device. Use the same basecalling model, modified base setting, reference FASTA file and barcode kit as the run, so that the rejected reads can be analysed alongside the BAM files from the run.
In the EPI2ME Desktop App, use the wf-basecalling workflow with the pod5skip folder as input, the HAC model and 5mC and 5hmC modified base model used for the run, the reference FASTA file, barcode kit SQK-NBD114.24 (or SQK-NBD114.96) and BAM output. The workflow writes aligned BAM files for each barcode.
On the command line, use the standalone Dorado software in two steps.
First basecall, align and identify the barcodes, replacing the file names with your own:
dorado basecaller hac,5mCG_5hmCG pod5skip/ --recursive --reference reference.fasta --kit-name SQK-NBD114.24 > rejected.bam
Then split the reads by barcode, keeping the alignment:
dorado demux --output-dir rejected_by_barcode --no-classify --no-trim --sort-bam rejected.bam
The --no-trim option is needed because dorado demux removes the alignment when it trims barcodes; the barcodes have already been trimmed during basecalling.
The --sort-bam option sorts and indexes the files for each barcode, and --recursive reads any sub-folders of the pod5skip folder.
The hac model name selects the current HAC model, which may be newer than the model used during the run; for an exact match, use the full model name shown in the run report.
Analysis
Use *_analysis.bed, not *_enrichment.bed, to calculate coverage over your targets and to restrict variant calling to them. If coverage is calculated over the enrichment BED file, the result includes the buffer and no longer describes your targets. Use *_exons.bed where coverage is reported per exon.
For custom germline panels, variant calling, phasing and methylation analysis can be performed per sample with the wf-human-variation workflow in the EPI2ME Desktop App or on the command line, supplying the BAM files for one barcode, the reference FASTA file and the analysis BED file.
For the Hereditary Cancer Panel, use the wf-hereditary-cancer workflow.
14. Flow cell reuse and returns
We do not recommend washing and reusing your flow cells for this method.
Due to the extended sequencing time, we do not recommend reusing the flow cells used in this method.
Reusing these flow cells for subsequent sequencing experiments may result in insufficient data generation for analysis.
Flow cells should be returned to ensure that Oxford Nanopore can responsibly recycle the finite materials within them.
Package the flow cells using the blister pack they arrived in to avoid damage during transit
Instructions for returning the flow cells can be found here.
If you encounter issues or have questions about your sequencing experiment, please refer to the Troubleshooting guide below.
15. Issues during DNA extraction and library preparation
Below is a list of the most commonly encountered issues, with some suggested causes and solutions.
We also have an FAQ section available on the Nanopore Community Support section.
If you have tried our suggested solutions and the issue persists, please contact Technical Support via email (support@nanoporetech.com) or via LiveChat in the Nanopore Community.
Low sample quality
| Observation | Possible cause | Comments and actions |
|---|---|---|
| Low DNA purity (Nanodrop reading for DNA OD 260/280 is <1.8 and OD 260/230 is <2.0–2.2) | The DNA extraction method does not provide the required purity | The effects of contaminants are shown in the Contaminants know-how piece. Please try an alternative extraction method that does not result in contaminant carryover. Consider performing an additional AMPure bead clean-up step. |
Low DNA recovery after AMPure bead clean-up
| Observation | Possible cause | Comments and actions |
|---|---|---|
| Low recovery | DNA loss due to a lower than intended AMPure beads-to-sample ratio | 1. AMPure beads settle quickly, so ensure they are well resuspended before adding them to the sample. 2. When the AMPure beads-to-sample ratio is lower than 0.4:1, DNA fragments of any size will be lost during the clean-up. |
| Low recovery | DNA fragments are shorter than expected | The lower the AMPure beads-to-sample ratio, the more stringent the selection against short fragments. Please always determine the input DNA length on an agarose gel (or other gel electrophoresis methods) and then calculate the appropriate amount of AMPure beads to use. ![]() |
| Low recovery after end-prep | The wash step used ethanol <70% | DNA will be eluted from the beads when using ethanol <70%. Make sure to use the correct percentage. |
16. Issues during the sequencing run
Below is a list of the most commonly encountered issues, with some suggested causes and solutions.
We also have an FAQ section available on the Nanopore Community Support section.
If you have tried our suggested solutions and the issue still persists, please contact Technical Support via email (support@nanoporetech.com) or via LiveChat in the Nanopore Community.
Fewer pores at the start of sequencing than after Flow Cell Check
| Observation | Possible cause | Comments and actions |
|---|---|---|
| MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check | An air bubble was introduced into the nanopore array | After the Flow Cell Check it is essential to remove any air bubbles near the priming port before priming the flow cell. If not removed, the air bubble can travel to the nanopore array and irreversibly damage the nanopores that have been exposed to air. The best practice to prevent this from happening is demonstrated in this video how to load a PromethION Flow Cell. |
| MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check | The flow cell is not correctly inserted into the device | Stop the sequencing run, remove the flow cell from the sequencing device and insert it again, checking that the flow cell is firmly seated in the device and that it has reached the target temperature. If applicable, try a different position on the device. |
| MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check | Contaminations in the library damaged or blocked the pores | The pore count during the Flow Cell Check is performed using the QC DNA molecules present in the flow cell storage buffer. At the start of sequencing, the library itself is used to estimate the number of active pores. Because of this, variability of about 10% in the number of pores is expected. A significantly lower pore count reported at the start of sequencing can be due to contaminants in the library that have damaged the membranes or blocked the pores. Alternative DNA extraction or purification methods may be needed to improve the purity of the input material. The effects of contaminants are shown in the Contaminants know-how piece. Please try an alternative extraction method that does not result in contaminant carryover. |
MinKNOW script failed
| Observation | Possible cause | Comments and actions |
|---|---|---|
| MinKNOW shows "Script failed" | Restart the computer and then restart MinKNOW. If the issue persists, please collect the MinKNOW log files and contact Technical Support. If you do not have another sequencing device available, we recommend storing the flow cell and the loaded library at 4°C and contact Technical Support for further storage guidance. |
Pore occupancy below 40%
| Observation | Possible cause | Comments and actions |
|---|---|---|
| Pore occupancy <40% | Not enough library was loaded on the flow cell | For the human genome sequencing protocols, 250-300 ng of good quality library should be loaded on to an R10.4.1 flow cell to keep pore occupancy high. |
| Pore occupancy close to 0 | Ethanol was used instead of LFB or SFB at the wash step after sequencing adapter ligation | Ethanol can denature the motor protein on the sequencing adapters. Make sure the LFB or SFB buffer was used after ligation of sequencing adapters. |
| Pore occupancy close to 0 | No tether on the flow cell | Tethers are added during flow cell priming (FCT tube). Make sure FCT was added to FCF before priming. |
Shorter than expected read length
| Observation | Possible cause | Comments and actions |
|---|---|---|
| Shorter than expected read length | Unwanted fragmentation of DNA sample | Read length reflects input DNA fragment length. Input DNA can be fragmented during extraction and library prep. 1. Please review the extraction methods in the Nanopore Community for best practice for extraction. 2. Visualise the input DNA fragment length distribution on an agarose gel before proceeding to the library prep. In the image above, Sample 1 is of high molecular weight, whereas Sample 2 has been fragmented.3. During library prep, avoid pipetting and vortexing when mixing reagents. Flicking or inverting the tube is sufficient. |
Large proportion of unavailable pores
| Observation | Possible cause | Comments and actions |
|---|---|---|
Large proportion of unavailable pores (shown as blue in the channels panel and pore activity plot) The pore activity plot above shows an increasing proportion of "unavailable" pores over time. | Contaminants are present in the sample | Some contaminants can be cleared from the pores by the unblocking function built into MinKNOW. If this is successful, the pore status will change to "sequencing pore". If the portion of unavailable pores stays large or increases: A nuclease flush using the Flow Cell Wash Kit (EXP-WSH004) can be performed. |
Large proportion of inactive pores
| Observation | Possible cause | Comments and actions |
|---|---|---|
| Large proportion of inactive/unavailable pores (shown as light blue in the channels panel and pore activity plot. Pores or membranes are irreversibly damaged) | Air bubbles have been introduced into the flow cell | Air bubbles introduced through flow cell priming and library loading can irreversibly damage the pores. Watch the how to load a PromethION Flow Cell video for best practice. |
| Large proportion of inactive/unavailable pores | Certain compounds co-purified with DNA | Known compounds, include polysaccharides. Clean-up using the QIAGEN PowerClean Pro kit. |
| Large proportion of inactive/unavailable pores | Contaminants are present in the sample | The effects of contaminants are shown in the Contaminants know-how piece. Please try an alternative extraction method that does not result in contaminant carryover. |
Temperature fluctuation
| Observation | Possible cause | Comments and actions |
|---|---|---|
| Temperature fluctuation | The flow cell has lost contact with the device | Check that there is a heat pad covering the metal plate on the back of the flow cell. Re-insert the flow cell and press it down to make sure the connector pins are firmly in contact with the device. If the problem persists, please contact Technical Services. |
Failed to reach target temperature
| Observation | Possible cause | Comments and actions |
|---|---|---|
| MinKNOW shows "Failed to reach target temperature" | The instrument was placed in a location that is colder than normal room temperature, or a location with poor ventilation (which leads to the flow cells overheating) | MinKNOW has a default timeframe for the flow cell to reach the target temperature. Once the timeframe is exceeded, an error message will appear and the sequencing experiment will continue. However, sequencing at an incorrect temperature may lead to a decrease in throughput and lower q-scores. Please adjust the location of the sequencing device to ensure that it is placed at room temperature with good ventilation, then re-start the process in MinKNOW. |

In the image above, Sample 1 is of high molecular weight, whereas Sample 2 has been fragmented.
The pore activity plot above shows an increasing proportion of "unavailable" pores over time.