Singleplex Adaptive Sampling (SAS_9239_v114_revA_11Nov2026)


Overview

This protocol provides details for library preparation, sequencing, and analysis overview.

The protocol:

  • Uses genomic DNA (i.e. extracted from fresh or frozen tumour tissue, whole blood, or bone marrow)
  • Requires no PCR
  • Is compatible with R10.4.1 flow cells
  • Is for sequencing one sample per PromethION Flow Cell

For Research Use Only

Document version: SAS_9239_v114_revA_11Nov2026

1. Overview of the protocol

Introduction to the Singleplex Adaptive Sampling protocol

This protocol describes the process for library preparation and sequencing of one single sample per PromethION Flow Cell, using the Native Barcoding Kit and the Adaptive Sampling target enrichment function in MinKNOW. All Oxford Nanopore Technologies supplied reagents required for this protocol are included in the Singleplex Adaptive Sampling bundles available in 6-sample or 18-sample formats. These bundles include the Native Barcoding Kit, the auxiliary kits and all nanopore sequencing components, and are the standard configuration for implementing this workflow.

You can use an Oxford Nanopore ready-to-use, research-use-only digital panel as your BED-file-defined panel or create a custom digital panel using the Adaptive Sampling, Digital Panel Designer. For information on our ready-to-use panels or to create a custom panel, refer to the Adaptive Sampling technical guide - Digital Panel Designer.

We recommend sequencing for 72 hours to maximise depth across target regions. The resulting on-target coverage enables confident detection of somatic single nucleotide variants (SNVs), copy number variations (CNVs), indels, and structural variants (SVs), alongside whole-genome methylation profiling from the same dataset. The Somatic analysis bioinformatics guide provides an overview of methods for downstream analysis.

This protocol is recommended for tumour-only somatic oncology applications and other use cases requiring both methylation data and high on-target coverage (>75x).

2. Leveraging ready-to-use digital panels

Oxford Nanopore digital panel specification


Ready-to-use digital panels:

Hereditary Cancer Pan-Caner Pan-Haematological Cancer
Enriched genes 258 inherited cancer risk genes 723 cancer associated genes 144 haematological malignancies associated genes
Validated Adaptive Sampling protocols Multiplex Adaptive Sampling Ultra-rapid singleplex Adaptive Sampling

Singleplex Adaptive Sampling
Singleplex Adaptive Sampling
Input amount† 1 µg Ultra-rapid singleplex: 1 µg

Singleplex: 500 ng to 1 µg*
1 µg
Input sample type Blood Solid tumour tissue Blood
Bone marrow
Recommended extraction kits Genomic DNA extraction from human blood using the Qiagen Puregene kit‡ 1. QIAamp Fast DNA Tissue Kit (Qiagen, 51404)
Recommended for Ultra-rapid singleplex Adaptive Sampling

2. Puregene Tissue Kit (Qiagen, 158063)

3. MagAttract HMW DNA Kit (Qiagen, 67563)
1. Genomic DNA extraction from human blood using the Qiagen Puregene kit‡


2. QIAamp DSP DNA Mini Kit (Qiagen, 61304)

3. GenFind V3 Genomic DNA Extraction Kit (Beckman Coulter, C34880)
QC info (Femto Pulse) Extracted gDNA samples may be assessed on the Femto Pulse System to visualise the unsheared/sheared gDNA profiles. For example profiles refer to the Input DNA section. Extracted gDNA samples may be assessed on the Femto Pulse System to visualise the unsheared/sheared gDNA profiles. For example profiles refer to the DNA QC section.

For the Ultra-rapid singleplex AS protocol, perform gDNA quality control assessment after completion of the sequencing run using aliquots reserved during the relevant sample preparation steps.
Extracted gDNA samples may be assessed on the Femto Pulse System to visualise the unsheared/sheared gDNA profiles. For example profiles refer to the DNA QC section.
Shearing conditions (speed and time) Three centrifugation steps at ~3,381 RCF for 1 minute each.
Refer to the corresponding protocol step for further details.
Two centrifugation steps at ~3,381 RCF for 30 seconds each.
Refer to the corresponding protocol step for further details.
Two centrifugation steps at ~4,868 RCF for 1 minute each.
Refer to the corresponding protocol step for further details.
Reference genomes GRCh38_GIABv3 GRCh38_GIABv3 GCA_000001405.15_GRCh38_no_alt_analysis_set.fna
Enrichment BED file (links to the BED files can be found on the Adaptive Sampling page) ONT_AS_HCP_GRCh38_enrichment_v1.bed ONT_AS_PanCancer_GRCh38_enrichment_v1.bed ONT_AS_PanHaem_GRCh38_enrichment_v1.bed

*A lower input limit of 500 ng can be used; however 1 µg is recommended whenever possible to improve pore occupancy.

†A maximum of 2 µg input can be used to maximise pore occupancy.

‡This Oxford Nanopore adapted extraction method uses the Puregene Blood Kit (Qiagen, 158023). It has gentler mixing conditions and resuspends the final DNA at 50°C rather than 65°C.

3. Overview of steps in the workflow

Prepare for your experiment

You will need to:

  • Extract DNA from your input sample (fresh or frozen tumour tissue, whole blood or bone marrow).
  • Ensure you have your sequencing kit, the correct equipment, and third-party reagents.
  • Download the software for acquiring and analysing your data.
  • Ensure your analysis workstation or compute environment meets the hardware and software requirements of your chosen third-party analysis tools.
  • Check your flow cell to ensure it has sufficient pores for a good sequencing run.

Sample preparation, library preparation, and sequencing

Step in workflow Process Time Stop option
Sample preparation Extract gDNA from sample and quantify. Fragment the gDNA and quantify.

Check the length, quantity and purity of your extracted material. The quality checks performed during the protocol are recommended to ensure experimental success.
Dependent on extraction kit used

~15 minutes for shearing
At this stage, the extracted gDNA or fragmented gDNA can be stored at –20°C for later use.
DNA repair and end-prep Repair the gDNA, and prepare the DNA ends for barcode and adapter attachment. 60 minutes 4°C overnight
Native barcode and adapter ligation Ligate the native barcodes and sequencing adapters to the DNA ends. 90 minutes 4°C overnight
Priming and loading the flow cell Prime the flow cell, and load your DNA library into the flow cell. 10 minutes
Sequencing Sequence your library using your sequencing device. 72 hours
Data analysis Refer to section 11, Downstream analysis and to the Somatic analysis bioinformatics guide.

Compatibility of this protocol

This protocol should only be used in combination with:

Note: DNA libraries prepared from blood and bone marrow samples have been sequenced using the PromethION 2 Integrated only.

4. Equipment and consumables

Materials
  • 1 μg input DNA per sample
  • Singleplex Adaptive Sampling bundles (containing the materials listed below)
  • Native Barcoding Kit 24 V14 (SQK-NBD114.24)
  • LFB Expansion (EXP-LFB001)
  • Sequencing Auxiliary Vials V14 (EXP-AUX003) and Native Barcoding Auxiliary Kit V14 (EXP-NBA114) if using the 18 samples bundle

Consumables
  • PromethION Flow Cell (FLO-PRO114M)
  • g-TUBE (Covaris, 520079)
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Absolute ethanol (e.g. Sigma-Aldrich, 459836-100ML)
  • NEB Blunt/TA Ligase Master Mix (NEB, M0367)
  • NEBNext® Companion Module v2 for Oxford Nanopore Technologies® Ligation Sequencing (NEB, E7672S or E7672L)
  • Qubit™ dsDNA HS Assay Kit (Invitrogen, Q32851)
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • 0.2 ml thin-walled PCR tubes
  • Pipette tips (nuclease-free, filtered recommended)
  • Genomic DNA 165 kb Analysis Kit (Agilent, FP-1002-0275) (optional)

Equipment
  • PromethION device
  • Eppendorf® 5424 centrifuge (or equivalent capable of ~3,380 RCF) for tumour tissue or
  • Eppendorf® 5425 R centrifuge (or equivalent capable of ~4,868 RCF) for blood/ bone marrow
  • Microcentrifuge
  • Rotator mixer (e.g. Hula mixer)
  • Magnetic separation rack, suitable for 1.5 ml Eppendorf tubes
  • Vortex mixer
  • Thermal cycler with heated lid
  • Heat block (suitable for 1.5 ml tubes)
  • Qubit™ fluorometer (or equivalent for QC check)
  • Ice bucket with ice
  • Timer
  • P1000 pipette
  • P200 pipette
  • P100 pipette
  • P20 pipette
  • P10 pipette
  • P2 pipette
Optional equipment
  • Femto Pulse System (Agilent Technologies) or equivalent

Oxford Nanopore Technologies Singleplex Adaptive Sampling bundles

All Oxford Nanopore Technologies reagents required for this protocol are included in the Singleplex Adaptive Sampling bundles, available in the Nanopore Store in 6-sample and 18-sample formats.

These bundles include the Native Barcoding Kit, the auxiliary kits and all nanopore sequencing components, and are the standard configuration for implementing this workflow.

For more information on specific kit components found within these bundles, please refer to the relevant products found in the Nanopore Store page:

The above list of materials, consumables, and equipment is for the shearing of the DNA samples, as well as library preparation.

DNA extraction and input

This protocol supports DNA extraction from tumour tissue, blood and bone marrow using the recommended kits listed in section 2.

Applicable sample input requirements are also provided for each sample type.

Note: You will need to refer to the relevant kit handbook and protocol to determine which equipment and additional reagents are required for the extraction.

For this protocol, the following input is required for library preparation:

  • 1 μg of sheared gDNA

Input DNA

How to QC your input DNA

It is important that the input DNA meets the quantity and quality requirements. Using too little or too much DNA, or DNA of poor quality (e.g. highly fragmented or containing RNA or chemical contaminants) can affect your library preparation.

For instructions on how to perform quality control of your DNA sample, please read the Input DNA/RNA QC protocol.

Chemical contaminants

Depending on how the DNA is extracted from the raw sample, certain chemical contaminants may remain in the purified DNA, which can affect library preparation efficiency and sequencing quality. Read more about contaminants on the Contaminants page of the Community.

Third-party reagents

We have validated and recommend the use of all the third-party reagents used in this protocol. Alternatives have not been tested by Oxford Nanopore.

For all third-party reagents, we recommend following the manufacturer's instructions to prepare the reagents for use, unless specified in this protocol.

Check your flow cell

We highly recommend that you check the number of pores in your flow cell prior to starting a sequencing experiment, within the warranty period. Flow cells should be removed from refrigeration and their packaging and allowed to come to room temperature for 30 minutes before they are used.

To perform the flow cell check, please follow the instructions in the Flow Cell Check document. As part of your flow cell check, you will see the number of pores in your new flow cell. The warranty on your flow cell covers a minimum pore count of 5,000 pores for a PromethION Flow Cell. You can find more information about the warranty period, stability and storage conditions of flow cells on the Oxford Nanopore Store. If you have any concerns or questions, please contact Oxford Nanopore.

Oxford Nanopore will replace any unused PromethION Flow Cell with fewer than 5,000 pores, when the result is reported within two days of performing the flow cell check, and when the storage recommendations have been followed.

The Native Adapter (NA) used in this protocol is not interchangeable with other sequencing adapters.

5. Genomic DNA extraction

Extract genomic DNA from your sample using a recommended kit listed in section 2, following the manufacturer's instructions.

Following extraction, quantify 1 μl of extracted gDNA using the Qubit dsDNA HS Assay Kit following the manufacturer's instructions.

Optional: If interested in fragment size distribution analysis, reserve an additional 1 μl aliquot of the unsheared DNA for analysis on the Femto Pulse System. This should be run after sequencing to avoid increasing library preparation times.

Optional DNA QC

Tumour tissue

Following gDNA extraction, samples may be assessed on the Femto Pulse System to visualise the unsheared gDNA profiles. The comparison with representative traces from samples that did or did not subsequently yield acceptable unsheared fragment distributions can be used to guide QC (Figure 1).

Fragment length profiles for unsheared gDNA

Figure 1: Unsheared genomic DNA fragment length profiles from seven commercial tumour tissue samples extracted using the QIAamp Fast DNA Tissue Kit. Samples were assessed using the Agilent Femto Pulse System with the Genomic DNA 165 kb Kit.
Sample 1 displayed reduced fragment lengths consistent with higher levels of necrosis, while Samples 2–7 demonstrated acceptable fragment size distributions.

Take forward your extracted gDNA into the next step, DNA shearing.

Alternatively, your sample(s) can be stored at 4°C overnight or at –20°C for later use.

6. DNA shearing using g-TUBE

Materials
  • 1 μg extracted gDNA per sample

Consumables
  • g-TUBE (Covaris, 520079)
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Qubit™ dsDNA HS Assay Kit (Invitrogen, Q32851)
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • Pipette tips (nuclease-free, filtered recommended)
  • Genomic DNA 165 kb Analysis Kit (Agilent, FP-1002-0275) (optional)

Equipment
  • Eppendorf® 5424 centrifuge (or equivalent capable of ~3,380 RCF) for tumour tissue or
  • Eppendorf® 5425 R centrifuge (or equivalent capable of ~4,868 RCF) for blood/ bone marrow
  • Microcentrifuge
  • Qubit™ fluorometer (or equivalent for QC check)
  • Ice bucket with ice
  • P1000 pipette
  • P200 pipette
  • P100 pipette
  • P20 pipette
  • P10 pipette
  • P2 pipette
Optional equipment
  • Femto Pulse System (Agilent Technologies) or equivalent

To fragment gDNA, dilute 1 μg of sample DNA in nuclease-free water to a total volume of 52 μl.

Load the diluted sample into a g-TUBE according to the manufacturer’s instructions.

Centrifuge the g-TUBE using the recommended conditions below to achieve an ideal target fragment length of 8–10 kb. Balance the centrifuge with an additional tube if necessary.

The following conditions are recommended to achieve optimal target fragment length (8–10 kb) for different sample inputs:

Tumour tissue (fresh or frozen) Whole blood or bone marrow
Centrifugation conditions with g-TUBE ~3,381 RCF (6,000 RPM) for 30 seconds ~4,868 RCF (7,200 RPM) for 1 minute

Visually inspect the g-TUBE to confirm that the sample has moved completely from the upper to the lower chamber. If it has, invert the g-TUBE and centrifuge at the specified speed and time for your sample type.

If the liquid has not passed into the lower chamber completely, do not invert the tube and centrifuge again at the specified speed and time for your sample type to ensure that all material has passed through the gTUBE constriction. Once completed, invert the g-TUBE and continue.

Repeat steps 3 and 4 to perform a second pass of shearing.

With the sample volume in the lower chamber, unscrew the g-TUBE body, leaving the screw-cap containing the sample volume.

Retrieve the sample from the g-TUBE screw-cap and transfer it into a clean 1.5 ml Eppendorf tube.

Quantify 1 μl of sheared DNA using the Qubit dsDNA HS Assay Kit following the manufacturer's instructions.

The minimum concentration should be ~16 ng/µl.

Optional: If interested in fragment size distribution analysis, reserve an additional 1 μl aliquot of the sheared DNA for analysis on the Femto Pulse System. This should be run after sequencing to avoid increasing library preparation times.

Optional DNA QC

Tumour tissue

Following shearing, samples may be assessed on the Femto Pulse System to visualise the sheared gDNA profiles. The comparison with representative traces from samples that did or did not subsequently yield acceptable sheared fragment distributions can be used to guide QC (Figure 2).

Fragment length profiles for sheared DNA QC

Figure 2: Sheared genomic DNA fragment length profiles from seven commercial tumour tissue samples extracted using the QIAamp Fast DNA Tissue Kit followed by g-TUBE shearing. Samples were assessed using the Agilent Femto Pulse System with the Genomic DNA 165 kb Kit.
Sample 1 exhibited reduced fragment lengths after shearing, consistent with increased necrosis, whereas Samples 2–7 showed acceptable fragment size distributions. Post-shearing showed a modal fragment size of approximately 10 kb.

Take forward 1 μg of sheared DNA per sample into the library preparation section.

7. DNA repair and end-prep

Materials
  • 1 µg sheared input DNA per sample
  • AMPure XP Beads (AXP from the Native Barcoding Kit 24 V14)

Consumables
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Absolute ethanol (e.g. Sigma-Aldrich, 459836-100ML)
  • NEBNext® Companion Module v2 for Oxford Nanopore Technologies® Ligation Sequencing (NEB, E7672S or E7672L)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • 0.2 ml thin-walled PCR tubes
  • Pipette tips (nuclease-free, filtered recommended)
  • Qubit™ dsDNA HS Assay Kit (Invitrogen, Q32851)
  • Qubit™ Assay Tubes (Invitrogen, Q32856)

Equipment
  • Thermal cycler with heated lid
  • Microcentrifuge
  • Vortex mixer
  • Rotator mixer (e.g. Hula mixer)
  • Magnetic separation rack, suitable for 1.5 ml Eppendorf tubes
  • Heat block (suitable for 1.5 ml tubes) set to 37°C
  • Qubit™ fluorometer (or equivalent for QC check)
  • Timer
  • Ice bucket with ice
  • P1000 pipette
  • P200 pipette
  • P100 pipette
  • P20 pipette
  • P10 pipette
  • P2 pipette

Prepare the required reagents from the NEBNext Companion Module v2 in accordance with the manufacturer’s instructions.

To a 0.2 ml PCR tube, add the following reagents:

If multiple samples are being prepped, a mastermix of the NEBNext reagents can be made.

Reagent Volume
1 μg sheared input DNA (from previous step) 48 µl
NEBNext® FFPE DNA Repair Buffer v2 7 µl
NEBNext® FFPE DNA Repair Mix 2 µl
NEBNext® Ultra™ II End Prep Enzyme Mix 3 µl
Total 60 µl

Mix the components thoroughly by pipetting and spin down briefly.

Incubate the reactions in a thermal cycler at 20°C for 5 minutes, then 65°C for 5 minutes.

Cool the reactions briefly on ice.

Note: This is to ensure that the reactions are at room temperature before the addition of the AMPure XP beads.

Spin down and transfer each sample to a clean 1.5 ml Eppendorf tube and add 60 μl (1x ratio) of resuspended AMPure XP beads.

Pipette mix and incubate the samples on a rotator mixer (e.g. Hula mixer) at room temperature for 5 minutes.

Freshly prepare sufficient 80% ethanol in nuclease-free water for the number of samples being prepped.

Spin down each sample and pellet on a magnetic rack for at least 1–3 minutes, until the eluate is clear and colourless. Keep the tubes on the magnetic rack and pipette off the supernatants.

Keep the tubes on the magnetic rack and wash the beads with 200 µl of freshly prepared 80% ethanol, without disturbing the pellets. Remove the supernatants using a pipette and discard.

Repeat the previous step.

Spin down and place the samples back on the magnetic rack. Pipette off any residual supernatants. Allow the pellets to dry for ~30 seconds, but do not dry the pellets to the point of cracking.

Remove the samples from the magnetic rack and resuspend each pellet in 20 μl of nuclease-free water.

Incubate the samples on a heat block set to 37°C for 2 minutes with intermittent flicking to aid DNA elution from the beads.

Spin down the samples and pellet on the magnetic rack for at least 1–3 minutes, until the eluates are clear and colourless.

Collect and retain each eluate in separate, clean 1.5 ml Eppendorf tubes.

Quantify the eluates using the Qubit dsDNA HS Assay Kit following the manufacturer's instructions.

The expected minimum concentration for the eluate is ~16 ng/μl.

Take forward your end-prepped DNA samples into the native barcode and adapter ligation step.

8. Native barcode and adapter ligation

Materials
  • Reagents from the Native Barcoding Kit 24 V14 (SQK-NBD114.24):
  • Native Barcodes (NB01-24)
  • AMPure XP Beads (AXP)
  • Native Adapter (NA)
  • L Fragment Buffer (LFB)
  • Elution Buffer (EB)

Consumables
  • NEBNext® Companion Module v2 for Oxford Nanopore Technologies® Ligation Sequencing (NEB, E7672S or E7672L)
  • NEB Blunt/TA Ligase Master Mix (NEB, M0367)
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Qubit™ dsDNA HS Assay Kit (Invitrogen, Q32851)
  • Qubit™ Assay Tubes (Invitrogen, Q32856)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • Pipette tips (nuclease-free, filtered recommended)

Equipment
  • Microcentrifuge
  • Vortex mixer
  • Rotator mixer (e.g. Hula mixer)
  • Magnetic separation rack, suitable for 1.5 ml Eppendorf tubes
  • Heat block (suitable for 1.5 ml tubes) set to 37°C
  • Qubit™ fluorometer (or equivalent for QC check)
  • Timer
  • Ice bucket with ice
  • P1000 pipette
  • P200 pipette
  • P100 pipette
  • P20 pipette
  • P10 pipette
  • P2 pipette

Check your flow cell

We recommend performing a flow cell check before starting native barcode and adapter ligation to ensure you have a flow cell with sufficient pores for a good sequencing run.

Flow cells should be removed from refrigeration and their packaging and allowed to come to room temperature for 30 minutes prior to use.

Condensation can form on the flow cells in humid environments. Inspect the gold connector pins on the top and underside of the flow cell for condensation and wipe off with a lint-free wipe if any is observed. Ensure the heat pad (dark grey) is present on the underside of the flow cell.

Refer to the flow cell check document for more information.

Prepare the required reagents from NEB in accordance with the manufacturer’s instructions.

Thaw the Native Barcodes (NB01-24) at room temperature . Individually mix the barcodes by pipetting, spin down, and place them on ice.

The wells of the barcoding plate are intended for single use only. Please ensure your barcode well is sealed before use. Do not reuse the barcode wells.

In a 1.5 ml Eppendorf tube, add the reagents in the following order for the barcoding reaction:

Note: Only use one unique barcode per sample, if prepping multiple samples.

Reagent Volume
End-prepped DNA 18 µl
Native Barcode (NB01-24) 2 µl
Blunt/TA Ligase Master Mix 20 µl
Total 40 µl

Mix well by pipetting and spin down the reactions. Incubate at room temperature for 20 minutes.

Add 2 µl of Native Adapter (NA) to each barcoding reaction.

The Native Adapter (NA) used in this protocol is not interchangeable with other sequencing adapters.

Add 5 µl Salt-T4® DNA Ligase to each sample.

Mix well by pipetting and incubate the samples at room temperature for 20 minutes.

Add 23.5 μl (0.5x) of resuspended AMPure XP beads to each sample and incubate on a rotator mixer at room temperature for 10 minutes.

Spin down and place the samples on a magnetic rack. Allow the beads to pellet for 3 minutes. Keep the tubes on the magnetic rack and pipette off the supernatants.

Wash the beads by adding 250 μl of L Fragment Buffer (LFB) to each sample and fully resuspend the pellet by flicking the tube. Pellet the beads on the magnetic rack for at least 3–5 minutes, until the supernatants are clear. Remove the supernatants using a pipette and discard.

Note: Take care when removing the supernatant as the viscosity of the buffer can contribute to loss of beads from the pellet.

Repeat the previous step.

Spin down and place the samples back on the magnetic rack. Pipette off any residual supernatants. Allow the pellets to dry for ~30 seconds, but do not dry the pellets to the point of cracking.

Remove the samples from the magnetic rack and resuspend each pellet in 34 μl of Elution Buffer (EB).

Spin down the samples and incubate in a heat block set at 37°C for 20 minutes. Every 2 minutes, agitate the samples by gently flicking for 10 seconds to aid DNA elution.

Pellet the beads on the magnetic rack for at least 1–3 minutes, until the eluates are clear and colourless.

Remove and retain 33 μl of each eluate containing the DNA library into a clean 1.5 ml Eppendorf DNA LoBind tube.

Quantify the eluates using the Qubit dsDNA HS Assay Kit following the manufacturer's instructions.

The expected minimum concentration is ~4 ng/µl.

The prepared library is used for loading onto the flow cell. Store the library on ice or at 4°C until ready to load.

9. Priming and loading the PromethION Flow Cell

Materials
  • Reagents from the Native Barcoding Kit 24 V14 (SQK-NBD114.24):
  • Flow Cell Flush (FCF)
  • Flow Cell Tether (FCT)
  • Library Beads (LIB)
  • Sequencing Buffer (SB)

Consumables
  • PromethION Flow Cell (FLO-PRO114M)
  • 1.5 ml Eppendorf DNA LoBind tubes
  • Pipette tips (nuclease-free, filtered recommended)

Equipment
  • PromethION device
  • PromethION Flow Cell Light Shield
  • Microcentrifuge
  • Vortex mixer
  • Timer
  • Ice bucket with ice
  • P1000 pipette
  • P200 pipette
  • P100 pipette
  • P20 pipette

This kit is only compatible with R10.4.1 flow cells (FLO-PRO114M).

We recommend removing your flow cells from refrigeration and their packaging, and allowing them to come to room temperature for 30 minutes before they are inserted into the device.

Condensation can form on the flow cell in humid environments. Inspect the gold connector pins on the top and underside of the flow cell for condensation and wipe off with a lint-free wipe if any is observed. Ensure the heat pad (dark grey) is present on the underside of the flow cell.

Thaw the Sequencing Buffer (SB), Library Beads (LIB), Flow Cell Tether (FCT) and Flow Cell Flush (FCF) at room temperature before mixing by vortexing. Then spin down and store on ice.

Prepare the flow cell priming mix in a suitable tube for the number of flow cells required. Once combined, mix well by briefly vortexing.

Reagent Volume per flow cell
Flow Cell Tether (FCT) 30 µl
Flow Cell Flush (FCF) 1,170 µl
Total volume 1,200 µl

Load the flow cell(s) into the device docking ports.

Note: DNA libraries prepared from blood and bone marrow samples have been sequenced using the PromethION 2 Integrated only.

Make sure that the dark grey heat pad on the underside of the flow cell is not torn or deformed.


For the PromethION 24:

  1. Line up the flow cell with the connector horizontally and vertically before smoothly inserting into position.
  2. Press down firmly onto the flow cell and ensure the latch engages and clicks into place.

Flow cell insertion on P24

Step 1B


For the PromethION 2 Integrated:

  1. Place the flow cell flat on the metal plate.

  2. Slide the flow cell into the docking port until the gold pins cannot be seen.

  3. Press the flow cell firmly into the loader until a click is heard.

Flow cell insertion on P2

Insertion of the flow cells at the wrong angle can cause damage to the pins on the PromethION and affect your sequencing results. If you find the pins on a PromethION position are damaged, please contact support@nanoporetech.com for assistance.

Pin damage on PromethION Flow Cell

Complete a flow cell check to assess the number of pores available before loading the library.

This step can be omitted if the flow cell has been checked previously.

Refer to the flow cell check document for more information.

Slide the inlet port cover clockwise to open.

Prom loading 2

Take care when drawing back buffer from the flow cell. Do not remove more than 20–30 µl, and make sure that the array of pores are covered by buffer at all times. Introducing air bubbles into the array can irreversibly damage pores.

After opening the inlet port, draw back a small volume to remove any air bubbles:

  1. Set a P1000 pipette tip to 200 µl.
  2. Insert the tip into the inlet port.
  3. Turn the wheel until the dial shows 220–230 µl, or until you see a small volume of buffer entering the pipette tip.

Step 3 v1

Load 500 µl of the priming mix into the flow cell via the inlet port, avoiding the introduction of air bubbles. Wait five minutes. During this time, prepare the library for loading using the next steps in the protocol.

Step 4 v1

Thoroughly mix the contents of the Library Beads (LIB) by pipetting.

The Library Beads (LIB) tube contains a suspension of beads. These beads settle very quickly. It is vital that they are mixed immediately before use.

In a new 1.5 ml Eppendorf DNA LoBind tube, prepare the library for loading as follows:

Reagent Volume per flow cell
Sequencing Buffer (SB) 100 µl
Library Beads (LIB) thoroughly mixed before use 68 µl
DNA library 32 µl
Total 200 µl

Complete the flow cell priming by slowly loading 500 µl of the priming mix into the inlet port.

Step 5 v1

Mix the prepared library gently by pipetting up and down just prior to loading.

Load 200 µl of library into the inlet port using a P1000 pipette.

Step 6 v1

Close the valve to seal the inlet port.

Step 7 V2

For optimal sequencing output, install the light shield on your flow cell as soon as the library has been loaded.

We recommend leaving the light shield on the flow cell when library is loaded, including during any washing and reloading steps. The shield can be removed when the library has been removed from the flow cell.

If the light shield has been removed from the flow cell, install the light shield as follows:

  1. Align the inlet port cut out of the light shield with the inlet port cover on the flow cell. The leading edge of the light shield should sit above the flow cell ID.
  2. Firmly press the light shield around the inlet port cover. The inlet port clip will click into place underneath the inlet port cover.

J2264 - Light shield animation PromethION Flow Cell 8a FAW

J2264 - Light shield animation PromethION Flow Cell 8b FAW

Close the PromethION lid when ready to start a sequencing run on MinKNOW.

Wait a minimum of 10 minutes after loading the library and before initiating any experiments. This will help to increase the sequencing output.

10. Data acquisition and basecalling

How to start sequencing

The sequencing device control, data acquisition and real-time basecalling are carried out by the MinKNOW software.

We recommend basecalling with the super accurate (SUP) basecaller in real-time with BAM selected as output type using the PromethION 24 (P24) or PromethION 2 Integrated (P2i) device.

Refer to the links below containing the detailed instructions for setting up the device and sequencing run:

Note: DNA libraries prepared from blood and bone marrow samples have been sequenced using the PromethION 2 Integrated only.

If using a custom digital panel, refer to the Adaptive Sampling technical guide for information on setting up your sequencing run in MinKNOW and additional details and guidance.

MinKNOW settings for Adaptive Sampling on PromethION

We strongly recommend using MinKNOW version 26.01.21 LTS or newer for optimal results.

Make sure you have the correct Adaptive Sampling files for MinKNOW (reference genome FASTA file and target region BED file) available on your device for data acquisition and analysis.

For ready-to-use digital panels, refer to section 2.

If you require assistance ensuring you are on the correct version of MinKNOW, please contact your Oxford Nanopore Technologies representative.

Below are the recommended sequencing parameters for MinKNOW.

Setting up your sequencing run in MinKNOW

Open the MinKNOW software UI via your desktop application.

In the MinKNOW UI, choose Start sequencing.

Follow the sequencing run set up on-screen using the settings defined below.

  • Positions

    • Flow cell position: [user defined]
    • Experiment name: [user defined]
    • Flow cell type: FLO-PRO114M
    • Sample ID: [user defined]
  • Kit selection

    • Kit selection: Native Barcoding Sequencing Kit (SQK-NBD114.24)
  • Run configuration

    • Sequencing and analysis
    • Basecalling: On [default]
    • Modified bases: On (5hmC and 5mC in CG contexts)
    • Model: Super accurate basecalling: (SUP)
    • Barcoding: On [default]

Note: If running multiple flow cells simultaneously, there may be a basecalling lag, so account for the extra time needed for catch-up.

  • Alignment: On

    • Reference for alignment: see recommendations below
    • Optional file enumerating regions: add BED file of choice
  • Adaptive sampling: On

    • Enrich selected
    • Basecall on-target reads only: Off [default]
    • Reference for alignment: see recommendations below
    • File enumerating regions: add BED file of choice

The following ready-to-use panels have been tested with this protocol:

  • Pan-Cancer
  • Pan-Haematological Cancer (available soon)
Ready-to-use panel Reference file for alignment
Pan-Cancer GRCh38_GIABv3
Pan-Haematological Cancer GCA_000001405.15_GRCh38_no_alt_analysis_set.fna

This protocol is compatible with BED files exported from the Digital Panel Designer or any other ready-to-use Oxford Nanopore digital panel.

  • Data targets

    • Run limit: Stop when sequencing reaches 72 h [default]
  • Output Settings

    • Saved output file location: remote location for third-party methylation classification analysis
  • Output format

    • POD5: Off
      Note: Enable POD5 output if downstream re-basecalling is anticipated
    • FASTQ: Off
    • BAM: On [default]
    • Based on: Time elapsed
    • Frequency: Every 1 hour [default]

When performing methylation-based classification, consult the requirements of your chosen third-party classifier regarding recommended BAM file size.

MinKNOW software can generate BAM files based on either a specified time interval or a predefined read count, providing flexibility to accommodate different analysis workflows. As methylation classifiers may have different input requirements, ensure the BAM file generation settings are configured accordingly.

11. Downstream analysis

Methylation calling

Methylation calls are contained within the BAM files generated by MinKNOW, which are saved in the run output directory:

/data/{experiment_group}/{sample_id}/{start_time}_{device_id}_{flow_cell_id}_{short_run_id}/bam_pass/

Example:

/data/TumourMethylationProfiling/VQL/20260514_1430_P2S-01234-_PAA12345_e05460c6/bam_pass/

Methylation based classification using third-party tools can be performed as early as 1 hour after sequencing, once approximately 0.75 Gb of pass data has been generated.

Variant calling

If proceeding to variant calling, we recommend this is performed after 72 hours of sequencing, to achieve maximum coverage.


Refer to the Somatic analysis bioinformatics guide for an overview of possible methods for the analysis.

12. Flow cell reuse and returns

We do not recommend washing and reusing your flow cells for this method.

Due to the extended sequencing time, we do not recommend reusing the flow cells used in this method.

Reusing these flow cells for subsequent sequencing experiments may result in insufficient data generation for analysis.

Flow cells should be returned to ensure that Oxford Nanopore can responsibly recycle the finite materials within them.

Package the flow cells using the blister pack they arrived in to avoid damage during transit.

Instructions for returning the flow cells can be found here.

If you encounter issues or have questions about your sequencing experiment, please refer to the Troubleshooting guide below.

13. Issues during DNA extraction and library preparation

Below is a list of the most commonly encountered issues, with some suggested causes and solutions.

We also have an FAQ section available on the Nanopore Community Support section.

If you have tried our suggested solutions and the issue still persists, please contact Technical Support via email (support@nanoporetech.com) or via LiveChat in the Nanopore Community.


Low sample quality

Observation Possible cause Comments and actions
Low DNA purity (Nanodrop reading for DNA OD 260/280 is <1.8 and OD 260/230 is <2.0–2.2) The DNA extraction method does not provide the required purity The effects of contaminants are shown in the Contaminants know-how piece. Please try an alternative extraction method that does not result in contaminant carryover.

Consider performing an additional AMPure bead clean-up step.

Low DNA recovery after AMPure bead clean-up

Observation Possible cause Comments and actions
Low recovery DNA loss due to a lower than intended AMPure beads-to-sample ratio 1. AMPure beads settle quickly, so ensure they are well resuspended before adding them to the sample.

2. When the AMPure beads-to-sample ratio is lower than 0.4:1, DNA fragments of any size will be lost during the clean-up.
Low recovery DNA fragments are shorter than expected The lower the AMPure beads-to-sample ratio, the more stringent the selection against short fragments. Please always determine the input DNA length on an agarose gel (or other gel electrophoresis methods) and then calculate the appropriate amount of AMPure beads to use. SPRI cleanup
Low recovery after end-prep The wash step used ethanol <70% DNA will be eluted from the beads when using ethanol <70%. Make sure to use the correct percentage.

14. Issues during the sequencing run

Below is a list of the most commonly encountered issues, with some suggested causes and solutions.

We also have an FAQ section available on the Nanopore Community Support section.

If you have tried our suggested solutions and the issue still persists, please contact Technical Support via email (support@nanoporetech.com) or via LiveChat in the Nanopore Community.


Fewer pores at the start of sequencing than after Flow Cell Check

Observation Possible cause Comments and actions
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check An air bubble was introduced into the nanopore array After the Flow Cell Check it is essential to remove any air bubbles near the priming port before priming the flow cell. If not removed, the air bubble can travel to the nanopore array and irreversibly damage the nanopores that have been exposed to air. The best practice to prevent this from happening is demonstrated in this video how to load a PromethION Flow Cell.
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check The flow cell is not correctly inserted into the device Stop the sequencing run, remove the flow cell from the sequencing device and insert it again, checking that the flow cell is firmly seated in the device and that it has reached the target temperature. If applicable, try a different position on the device.
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check Contaminations in the library damaged or blocked the pores The pore count during the Flow Cell Check is performed using the QC DNA molecules present in the flow cell storage buffer. At the start of sequencing, the library itself is used to estimate the number of active pores. Because of this, variability of about 10% in the number of pores is expected. A significantly lower pore count reported at the start of sequencing can be due to contaminants in the library that have damaged the membranes or blocked the pores. Alternative DNA extraction or purification methods may be needed to improve the purity of the input material. The effects of contaminants are shown in the Contaminants know-how piece. Please try an alternative extraction method that does not result in contaminant carryover.

MinKNOW script failed

Observation Possible cause Comments and actions
MinKNOW shows "Script failed"
Restart the computer and then restart MinKNOW. If the issue persists, please collect the MinKNOW log files and contact Technical Support. If you do not have another sequencing device available, we recommend storing the flow cell and the loaded library at 4°C and contact Technical Support for further storage guidance.

Pore occupancy below 40%

Observation Possible cause Comments and actions
Pore occupancy <40% Not enough library was loaded on the flow cell For this protocol, 50 to 65 fmol of good quality library should be loaded on to an R10.4.1 flow cell to keep pore occupancy high.
Pore occupancy close to 0 Ethanol was used instead of LFB or SFB at the wash step after sequencing adapter ligation Ethanol can denature the motor protein on the sequencing adapters. Make sure the LFB buffer was used after ligation of sequencing adapters.
Pore occupancy close to 0 No tether on the flow cell Tethers are added during flow cell priming (FCT tube). Make sure FCT was added to FCF before priming.

Shorter than expected read length

Observation Possible cause Comments and actions
Shorter than expected read length Unwanted fragmentation of DNA sample Read length reflects input DNA fragment length. Input DNA can be fragmented during extraction and library prep.

1. Please review the extraction methods in the Nanopore Community for best practice for extraction.

2. Visualise the input DNA fragment length distribution on an agarose gel before proceeding to the library prep. DNA gel2 In the image above, Sample 1 is of high molecular weight, whereas Sample 2 has been fragmented.

3. During library prep, avoid pipetting and vortexing when mixing reagents. Flicking or inverting the tube is sufficient.

Large proportion of unavailable pores

Observation Possible cause Comments and actions
Large proportion of unavailable pores (shown as blue in the channels panel and pore activity plot)

image2022-3-25 10-43-25 The pore activity plot above shows an increasing proportion of "unavailable" pores over time.
Contaminants are present in the sample Some contaminants can be cleared from the pores by the unblocking function built into MinKNOW. If this is successful, the pore status will change to "sequencing pore". If the portion of unavailable pores stays large or increases:

A nuclease flush using the Flow Cell Wash Kit (EXP-WSH004) can be performed.

Large proportion of inactive pores

Observation Possible cause Comments and actions
Large proportion of inactive/unavailable pores (shown as light blue in the channels panel and pore activity plot. Pores or membranes are irreversibly damaged) Air bubbles have been introduced into the flow cell Air bubbles introduced through flow cell priming and library loading can irreversibly damage the pores. Watch the how to load a PromethION Flow Cell video for best practice.
Large proportion of inactive/unavailable pores Certain compounds co-purified with DNA Known compounds, include polysaccharides.

Clean-up using the QIAGEN PowerClean Pro kit.
Large proportion of inactive/unavailable pores Contaminants are present in the sample The effects of contaminants are shown in the Contaminants know-how piece. Please try an alternative extraction method that does not result in contaminant carryover.

Temperature fluctuation

Observation Possible cause Comments and actions
Temperature fluctuation The flow cell has lost contact with the device Check that there is a heat pad covering the metal plate on the back of the flow cell. Re-insert the flow cell and press it down to make sure the connector pins are firmly in contact with the device. If the problem persists, please contact Technical Services.

Failed to reach target temperature

Observation Possible cause Comments and actions
MinKNOW shows "Failed to reach target temperature" The instrument was placed in a location that is colder than normal room temperature, or a location with poor ventilation (which leads to the flow cells overheating) MinKNOW has a default timeframe for the flow cell to reach the target temperature. Once the timeframe is exceeded, an error message will appear and the sequencing experiment will continue. However, sequencing at an incorrect temperature may lead to a decrease in throughput and lower q-scores. Please adjust the location of the sequencing device to ensure that it is placed at room temperature with good ventilation, then re-start the process in MinKNOW.
Oxford Nanopore Technologies, the Wheel icon, AmPORE-TB, EPI2ME, GridION, MinION, MinKNOW, PromethION, P2 Solo, and P2 are registered trademarks or the subject of trademark applications of Oxford Nanopore Technologies plc in various countries. Information contained herein may be protected by copyright, patents or patents pending of Oxford Nanopore Technologies plc. All other brands and names contained are the property of their respective owners. Oxford Nanopore Technologies products are RUO. Products labelled/branded as Oxford Nanopore Diagnostics may be RUO or may be regulated as in‐vitro diagnostic devices in some jurisdictions, please check individual product labelling. ONT plc is a member of the producer compliance scheme run by ERP UK Ltd, who manage the submission of documentation in support of WEEE compliance for ONT plc’s manufacture and supply of Electrical and Electronic equipment in the UK. ONT’s WEEE PRN is WEE/MM3828AA.

Last updated: 11/11/2026

Document options

PromethION
Language:

Getting started

Buy a MinION starter pack Nanopore store Sequencing Service providers Channel Partners

Quick links

Intellectual property Cookie policy Corporate reporting Privacy policy Terms, conditions and policies Modern slavery policy Accessibility

About Oxford Nanopore

Contact us News Media resources & contacts Investor centre Careers BSI 27001 accreditationBSI 90001 accreditation
English flag