How can I determine the number of reads that are being rejected due to adaptive sampling?

How can I determine the number of reads that are being rejected due to adaptive sampling?

You can determine the number of reads that are being rejected due to adaptive sampling via two methods:

  1. During the run: in the Experiments tab in MinKNOW by opening the Read length histogram and ticking the box "Split by read end reason". If you are running in a mode where enough reads should be rejected to produce a peak of short reads, there should be a bimodal distribution with a peak at a length much lower than what you expect from the sample. The reads rejected by adaptive sampling will have the label "Adaptive sampling voltage reversal" and are shown in purple in the read length histogram. The image below zooms in on this peak of the rejected reads.
9 AG DeviceSpec AS histogram1
  1. Immediately after the sequencing run starts: please ensure that an adaptive_sampling folder has been generated inside your data output folder. This folder should contain two files: AS_decisions_x_x_x.csv file and AS_timings_x_x_x.csv file.
    The AS_decisions_x_x_x.csv file contains information about the final decision that was taken by adaptive sampling for each read. Filtering this file by “action” column provides not only the number of accepted/rejected reads but also the specific readIDs of each read accepted and rejected. Accepted reads are marked as “sequence” and rejected reads are marked as “unblock”.

Getting started

Buy a MinION starter pack Nanopore store Sequencing Service providers Channel Partners

Quick links

Intellectual property Cookie policy Corporate reporting Privacy policy Terms, conditions and policies Modern slavery policy Accessibility

About Oxford Nanopore

Contact us News Media resources & contacts Investor centre Careers BSI 27001 accreditationBSI 90001 accreditation
English flag