Flow Cell Wash Kit (EXP-WSH004 or EXP-WSH004-XL)


概览

For Research Use Only

Document version: WFC_9120_v1_revQ_08Dec2020

1. Overview of the protocol

Introduction to the Flow Cell Wash Kit

The Flow Cell Wash Kit allows sequential runs of multiple sequencing libraries on the same flow cell. It works by flushing out and digesting the first library, refreshing the system for a subsequent library to be loaded. This procedure provides the opportunity to utilise the same flow cell a number of times, maximising the available run time, particularly for cases where less data per library is required. Following the wash step, Storage Buffer can be introduced into the flow cell, allowing storage of the flow cell at 4-8°C. The Flow Cell Wash Kit is compatible with Q-Line, R9.4.1 and R10.4.1 flow cells.

The nuclease digests DNA library and for effective removal from the flow cell before loading a new library, we recommend removing the waste buffer from the waste ports after each flush step. This is to ensure there is no nuclease in the waste channel that may diffuse through the flow cell during sequencing. To further improve flow cell output after a wash, we recommend pipetting slowly during flushing steps by either twisting the pipette wheel down slowly or pressing the plunger down slowly.

Please note, although the wash procedure should remove 99.9% of the library, some residual DNA may remain on the flow cell. For this reason, we recommend to barcode your libraries when used in conjunction with the Flow Cell Wash Kit, to ensure that reads from different libraries can be separated from each other.

Successful deconvolution of DNA reads has been demonstrated in Oxford Nanopore's internal development:

WSH004-XL barcodes1 Figure 1. A sample with four barcodes was sequenced and washed using EXP-WSH004 before a sample with four different barcodes was loaded. This was repeated for a third sequencing run.

For users who wish to use barcoding to run multiple libraries at one time rather than washing the flow cells, please see the barcoding kits we have available in the chemistry technical document.

The Flow Cell Wash Kit is can also be used with our RNA flow cells to flush out RNA libraries, but will not remove RNA-related blocking and restore nanopores.

Nuclease activity of the Flow Cell Wash Kit

The Flow Cell Wash Kit contains DNase I, that is used to digest any remaining DNA library on a flow cell. Once the library is removed, the flow cell can be re-used immediately or stored for later use.

During sequencing, an accumulation of pores in the “unavailable" state (Figure 2) may be observed, causing the rate of data acquisition to decline as fewer pores are available to accept and sequence strands. We have demonstrated that in these circumstances, pores can be reverted to the “active pore” state by pausing sequencing and washing the flow cell with the DNase I in the Flow Cell Wash Kit. In Figure 2, the astrisks indicate where sequencing has been paused and the flow cell washed. Note: If the sequencing run is paused in MinKNOW for the flow cell wash, you will only see the restoration of sequencing pores after a new pore scan has been performed.

The wash step is recommended where sequencing channels are lost to the “unavailable” state (Figure 2). In circumstances where channels have been lost by other means, for example “saturated”, the wash step is not effective at reverting channels to the “active pore” state. If your sample is not known to block and cause saturation of the flow cell, we recommend loading a fresh library on a new flow cell. If you do not have excess library to load on a new flow cell, you can recover the library and reload on a new flow cell following method 1 of the Library recovery from flow cells protocol.

image (109) Figure 2. Pore states observed on a flow cell before and after wash steps are performed. A flow cell has been loaded with a sequencing library that has resulted in an accumulation of pores in the “unavailable” state, leading to a decrease in the rate of data acquisition. The red asterisks indicate when a wash step has been performed. A significant number of the pores that had been lost to the “unavailable” state have reverted to the “Pore available” state and are available for sequencing once again.

In experiments where output is limited by the increase in pores in the “unavailable” state, we have shown that output can be improved by performing several wash steps over the lifetime of a flow cell. Figure 3 shows the output obtained from a PromethION Flow Cell loaded with a library of DNA extracted from chicken - Gallus gallus, and a MinION Flow Cell loaded with a library of DNA extracted from a type of Japense ricefish - Oryzias latipes, where unavailable pores increased over the course of the experiment, and so flow cell washes were performed to unblock the pores (Figure 3). In each case, the use of multiple washes allowed for an improvement of the output from the flow cell, without any compromise in observed read length (Figure 4).

2018 10 08 Figure 3 v2 DS

Figure 3. Throughput observed from Gallus gallus and Oryzias latipes libraries run on a PromethION Flow Cell and a MinION Flow Cell, respectively. The arrows indicate the timing of each wash step: wash steps were performed at the point where the rate of data acquisition started to slow due to the accumulation of “recovering” pores. In each case, output is more than doubled from the point of the first wash.

2018 10 03 Figure 4 DS

Figure 4. Effective inactivation of the DNase I prevents read length deterioration in the experiments after a nuclease wash is performed. In this example, the read length of the Gallus gallus library was recorded before the first wash (load #1) and then again after the first and second washes (load #2 and #3, respectively). No decrease in read length is observed.

2. Equipment and consumables

材料
  • 测序芯片清洗剂盒(EXP-WSH004)或测序芯片清洗试剂盒 XL(EXP-WSH004-XL)
  • Flow cell priming reagents available in your sequencing kit or in the following kits:
  • Sequencing Auxiliary Vials V14 (EXP-AUX003)
  • Flow Cell Priming Kit (EXP-FLP004)
仪器
  • P1000 移液枪和枪头
  • P20 移液枪和枪头
  • 盛有冰的冰桶

Reloading a library

Additional buffers are required for reloading a library following the washing of a flow cell. These can be found in the one of the following expansion kits:

  • Sequencing Auxiliary Vials V14 (EXP-AUX003). This expansion contains vials of Sequencing Buffer (SB), Elution Buffer (EB), Library Solution (LIS), and Library Beads (LIB) with Kit 14 flow cell priming reagents: Flow Cell Flush (FCF) and Flow Cell Tether (FCT). This expansion is only compatible with our Kit 14 chemistry e.g. SQK-LSK114.

  • Flow Cell Priming Kit (EXP-FLP004). This expansion contains both Kit 14 flow cell priming reagents required: Flow Cell Flush (FCF) and Flow Cell Tether (FCT). This expansion is only compatible with Kit 14 chemistry.


For our previous chemistries:

  • Sequencing Auxiliary Vials expansion (EXP-AUX001). This expansion contains vials of Elution Buffer (EB), Sequencing Buffer (SQB), and Loading Beads (LB), additional to those found in DNA sequencing kits for our Kit 9 chemistry.

  • Sequencing Auxiliary Vials expansion (EXP-AUX002). This expansion contains vials of Sequencing Buffer II (SBII), Elution Buffer (EB), Loading Solution (LS), and Loading Beads II (LBII), additional to those found in Kit 10 and 11 chemistry, such as:

    • Kit 10 e.g. SQK-LSK110
    • Kit 11 e.g. SQK-PCS111

  • Flow Cell Priming Kit (EXP-FLO002). This expansion contains both flow cell priming reagents required: Flush Buffer (FB) and Flush Tether (FLT). This expansion is compatible with Kit 9, 10 and 11 chemistry.

Flow Cell Wash Kit (EXP-WSH004) contents

EXP-WSH004 kit contents v2

Contents Volume (µl) No. of tubes No. of uses
Wash Mix (WMX) 15 1 6
Wash Diluent (DIL) 1,300 2 6
Storage Buffer (S) 1,600 2 6
  • Wash Mix (WMX) contains DNase I.
  • Wash Diluent (DIL) contains the exonuclease buffer that maximises activity of the DNase I.
  • The Storage Buffer allows flow cells to be stored for extended periods of time.

Flow Cell Wash Kit XL (EXP-WSH004-XL) contents

EXP-WSH004-XL

Name Acronym Fill volume per vial (µl) Cap colour No. of vials No. of uses
Wash Mix WMX 150 Brown 1 48
Wash Diluent DIL 20,000 White cap, brown stripe 1 48
Storage Buffer S 25,000 White cap, yellow stripe 1 48
  • Wash Mix (WMX) contains DNase I.
  • Wash Diluent (DIL) contains the exonuclease buffer that maximises activity of the DNase I.
  • The Storage Buffer allows flow cells to be stored for extended periods of time.

Flow Cell Priming Kit (EXP-FLP004) contents

EXP-FLP004v2

Name Acronym Cap colour No. of vials Fill volume per vial (μl)
Flow Cell Flush FCF 6 Clear cap, light blue lable 8,000
Flow Cell Tether FCT 1 Purple 200

Sequencing Auxiliary Vials V14 (EXP-AUX003) contents

EXP-AUX003 bottles

Name Acronym Cap colour No. of vials Fill volume per vial (μl)
Elution Buffer EB Black 2 500
Sequencing Buffer SB Red 2 700
Library Solution LIS White cap, pink label 2 600
Library Beads LIB Pink 2 600
Flow Cell Flush FCF Light blue label 2 8,000
Flow Cell Tether FCT Purple 2 200

3. Flushing a MinION/GridION Flow Cell

材料
  • 测序芯片清洗剂盒(EXP-WSH004)或测序芯片清洗试剂盒 XL(EXP-WSH004-XL)
仪器
  • P1000 移液枪和枪头
  • P20 移液枪和枪头
  • 盛有冰的冰桶

Preparation to run the washing procedure

  • This protocol assumes that the flow cell has already had a DNA/RNA library run on it
  • The aim is to remove most of this initial library from the flow cell
  • The Wash Kit contains all solutions required for removal of the initial library
  • Data acquisition in MinKNOW should be stopped (if loading a new library or storing the flow cell), or paused (if loading more of the same library after the wash)
  • After the flow cell has been washed, a new library can be loaded or the flow cell can be stored at 4°C
提示

We recommend keeping the light shield on the flow cell during washing if a second library will be loaded straight away.

If the flow cell is to be washed and stored, the light shield can be removed.

重要

A P1000 pipette must be used for all flushing steps to create a seal with the flow cell ports.

Place the tube of Wash Mix (WMX) on ice. Do not vortex the tube.

Thaw one tube of Wash Diluent (DIL) at room temperature.

Mix the contents of Wash Diluent (DIL) thoroughly by vortexing, then spin down briefly and place on ice.

In a fresh 1.5 ml Eppendorf DNA LoBind tube, prepare the following Flow Cell Wash Mix:

Reagent Volume per flow cell
Wash Mix (WMX) 2 μl
Wash Diluent (DIL) 398 μl
Total 400 μl

Mix well by pipetting, and place on ice. Do not vortex the tube.

Stop or pause the sequencing experiment in MinKNOW, and leave the flow cell in the device.

重要

It is vital that the flow cell priming port and SpotON sample port are closed before removing the waste buffer to prevent air from being drawn across the sensor array area, which would lead to a significant loss of sequencing channels.

Remove the waste buffer, as follows:

  1. Close the priming port and SpotON sample port cover, as indicated in the figure below.
  2. Insert a P1000 pipette into waste port 1 and remove the waste buffer.

Note: As both the priming port and SpotON sample port are closed, no fluid should leave the sensor array area.

Flow cell ports

Slide the flow cell priming port cover clockwise to open.

Flow Cell Loading Diagrams Step 2 (3)

重要

从测序芯片中反旋排出缓冲液。请勿吸出超过20-30µl的缓冲液,并确保芯片上的纳米孔阵列一直有缓冲液覆盖。将气泡引入阵列会对纳米孔造成不可逆转地损害。

After opening the priming port, check for a small air bubble under the cover. Draw back a small volume to remove any bubbles:

  1. Set a P1000 pipette to 200 µl.
  2. Insert the tip into the flow cell priming port.
  3. Turn the wheel until the dial shows 220-230 µl, or until you can see a small volume of buffer/liquid entering the pipette tip.
  4. Visually check that there is continuous buffer from the flow cell priming port across the sensor array.

Flow Cell Loading Diagrams Step 03 V5

Slowly load 200 µl of the prepared flow cell wash mix into the priming port, as follows:

  1. Using a P1000 pipette, take 200 µl of the flow cell wash mix
  2. Insert the pipette tip into the priming port, ensuring there are no bubbles in the tip
  3. Slowly twist the pipette wheel down to load the flow cell (if possible with your pipette) or push down the plunger very slowly, leaving a small volume of buffer in the pipette tip.
  4. Set a timer for a 5 minute incubation.

Loading wash mix 200ul slow min grid

Once the 5 minute incubation is complete, carefully load the remaining 200 µl of the prepared flow cell wash mix into the priming port, as follows:

  1. Using a P1000 pipette, take the remaining 200 µl of the flow cell wash mix
  2. Insert the pipette tip into the priming port, ensuring there are no bubbles in the tip
  3. Slowly twist the pipette wheel down to load the flow cell (if possible with your pipette) or push down the plunger very slowly, leaving a small volume of buffer in the pipette tip.

Loading wash mix 200ul slow min grid

Close the priming port and wait for 1 hour.

Flow Cell Loading Diagrams Step 9

重要

It is vital that the flow cell priming port and SpotON sample port are closed before removing the waste buffer to prevent air from being drawn across the sensor array area, which would lead to a significant loss of sequencing channels.

Remove the waste buffer, as follows:

  1. Ensure the priming port and SpotON sample port covers are closed, as indicated in the figure below.
  2. Insert a P1000 pipette into waste port 1 and remove the waste buffer.

Note: As both the priming port and SpotON sample port are closed, no fluid should leave the sensor array area.

Flow cell ports

步骤结束

Follow one of the two options described in the next steps of the protocol.

  • Run a second library on the flow cell straight away
  • Store the flow cell for later use

4. To run a second library on a MinION/GridION Flow Cell straight away

材料
  • 测序芯片清洗剂盒(EXP-WSH004)或测序芯片清洗试剂盒 XL(EXP-WSH004-XL)
  • Flow cell priming reagents available in your sequencing kit or in the following kits:
  • Sequencing Auxiliary Vials V14 (EXP-AUX003)
  • Flow Cell Priming Kit (EXP-FLP004)
仪器
  • P1000 移液枪和枪头
  • P20 移液枪和枪头
  • 盛有冰的冰桶
重要

The sequencing reagents outlined in this method are for our most recent V14 chemisty.

If using a previous version of our chemistry or a kit with specific sequencing reagents, please ensure you are using the correct sequencing reagents for your flow cell.

提示

The buffers used in this process are incompatible with conducting a Flow Cell Check prior to loading a subsequent library. However, the first pore scan once a sequencing run has started will report the number of nanopores available.

重要

A P1000 pipette must be used for all flushing steps to create a seal with the flow cell ports.

于室温下解冻测序缓冲液(SB)、文库颗粒(LIB)或文库溶液(LIS)、测序芯片系绳(FCT)和一管测序芯片冲洗液(FCF)。完全解冻后,涡旋振荡混匀,然后瞬时离心并置于冰上。

重要

为在MinION及GridION R10.4.1测序芯片(FLO-MIN114)上获得最优的测序表现并提高测序产出,我们推荐您向测序芯片预处理液中加入终浓度为0.2 mg/ml的牛血清白蛋白(BSA)。

请注意: 我们不推荐使用其它类型的白蛋白(例如重组人血清白蛋白)。

按下表制备测序芯片的预处理液,室温下吹打混匀。 (1)

请注意: 我们正在将部分试剂的包装形式由单次管装改为瓶装。请根据您所用试剂盒的包装,请遵循与您所使用的试剂盒包装相对应的操作说明。

单次使用管装: 向一整管测序芯片冲洗液(FCF)中加入5µl 50mg/ml的牛血清白蛋白(BSA)及 30µl 测序芯片系绳(FCT)。

瓶装: 请另拿一支适当体积的离心管制备测序芯片预处理液。

试剂 体积(每张芯片)
测序芯片冲洗液 (FCF) 1,170 µl
50mg/ml的牛血清白蛋白 (BSA) 5 µl
测序芯片系绳 (FCT) 30 µl
总体积 1,205 µl

顺时针转动预处理孔孔盖,使预处理孔显露出来。

中文-测序芯片预处理上样2

重要

从测序芯片中反旋排出缓冲液。请勿吸出超过20-30µl的缓冲液,并确保芯片上的纳米孔阵列一直有缓冲液覆盖。将气泡引入阵列会对纳米孔造成不可逆转地损害。

将预处理孔打开后,检查孔周围是否有小气泡。请按照以下方法,从孔中排出少量液体以清除气泡:

  1. 将P1000移液枪转至200µl刻度。
  2. 将枪头垂直插入预处理孔中。
  3. 反向转动移液枪量程调节转纽,直至移液枪刻度在220-230 µl之间,或直至您看到有少量缓冲液进入移液枪枪头。
    __请注意:__ 肉眼检查,确保从预处理孔到传感器阵列的缓冲液连续且无气泡。

中文-测序芯片预处理上样3

通过预处理孔向芯片中加入800µl预处理液,避免引入气泡。等待5分钟。在此期间,请按照以下步骤准备用于上样的DNA文库。 (1)

中文-测序芯片预处理上样4

重要

It is vital to wait five minutes between the priming mix flushes to effectively remove the nuclease.

Close the priming port and wait five minutes.

During this time, prepare the library for loading by following the steps below.

将含有文库颗粒的LIB管用移液枪吹打混匀。

重要

LIB管内的文库颗粒分散于悬浮液中。由于颗粒沉降速度非常快,因此请在混匀颗粒后立即使用。

对于大多数测序实验,我们建议您使用文库颗粒(LIB)。但如文库较为粘稠,您可考虑使用文库溶液(LIS)。

In a new tube, prepare the library for loading according to the "Priming and loading the MinION and GridION Flow Cell" section of the suitable protocol to ensure you are using the correct reagents and volumes.

For Kit 14 chemistry:

Reagent Volume per flow cell
Sequencing Buffer (SB) 37.5 µl
Library Beads (LIB) mixed immediately before use, or Library Solution (LIS), if using 25.5 µl
Recovered DNA library 12 µl
Total 75 µl
重要

It is vital that the flow cell priming port and SpotON sample port are closed before removing the waste buffer to prevent air from being drawn across the sensor array area, which would lead to a significant loss of sequencing channels.

Remove the waste buffer, as follows:

  1. Ensure the priming port and SpotON sample port covers are closed, as indicated in the figure below.
  2. Insert a P1000 pipette into waste port 1 and remove the waste buffer.

Note: As both the priming port and SpotON sample port are closed, no fluid should leave the sensor array area.

Flow cell ports

Slide the flow cell priming port cover clockwise to open.

Flow Cell Loading Diagrams Step 2 (3)

重要

从测序芯片中反旋排出缓冲液。请勿吸出超过20-30µl的缓冲液,并确保芯片上的纳米孔阵列一直有缓冲液覆盖。将气泡引入阵列会对纳米孔造成不可逆转地损害。

将预处理孔打开后,检查孔周围是否有小气泡。请按照以下方法,从孔中排出少量液体以清除气泡:

  1. 将P1000移液枪转至200µl刻度。
  2. 将枪头垂直插入预处理孔中。
  3. 反向转动移液枪量程调节转纽,直至移液枪刻度在220-230 µl之间,或直至您看到有少量缓冲液进入移液枪枪头。
    __请注意:__ 肉眼检查,确保从预处理孔到传感器阵列的缓冲液连续且无气泡。

中文-测序芯片预处理上样3

Slowly load 200 µl of the priming mix into the flow cell priming port, as follows:

  1. Ensure the priming port is open and gently lift open the SpotON sample port.
  2. Using a P1000 pipette, take 200 µl of the priming mix
  3. Insert the pipette tip into the priming port, ensuring there are no bubbles in the tip
  4. Slowly twist the pipette wheel down to load the flow cell (if possible with your pipette) or push down the plunger very slowly, as illustrated in the video above, leaving a small volume of buffer in the pipette tip.
重要

It is vital that the flow cell priming port and SpotON sample port are closed before removing the waste buffer to prevent air from being drawn across the sensor array area, which would lead to a significant loss of sequencing channels.

Remove the waste buffer, as follows:

  1. Close the priming port and SpotON sample port cover, as indicated in the figure below.
  2. Insert a P1000 pipette into waste port 1 and remove the waste buffer.

Note: As both the priming port and SpotON sample port are closed, no fluid should leave the sensor array area.

Flow cell ports

Slide open the priming port cover and gently lift open the SpotON sample port cover.

临上样前,用移液枪轻轻吹打混匀制备好的文库。

通过SpotON加样孔向芯片中逐滴加入75µl样品。确保液滴流入孔内后,再加下一滴。

中文-测序芯片预处理上样7

轻轻合上SpotON加样孔孔盖,确保塞头塞入加样孔内。逆时针转动预处理孔孔盖,盖上预处理孔。

中文-测序芯片预处理上样8

中文-测序芯片预处理上样9

重要

为获得最佳测序产出,在文库样本上样后,请立即在测序芯片上安装遮光片。

我们建议在清洗芯片并重新上样时,将遮光片保留在测序芯片上。一旦文库从测序芯片中吸出,即可取下遮光片。

按下述步骤安装测序芯片遮光片:

  1. 小心将遮光片的前沿(平端)与金属固定夹的边沿对齐。 请注意: 请勿将遮光片强行压到固定夹下方。

  2. 将遮光片轻轻盖在测序芯片上。遮光片的SpotON加样孔孔盖缺口应与芯片上的SpotON加样孔孔盖接合,遮盖住整个测序芯片的前部。

MinION加装遮光片

注意

MinION测序芯片的遮光片并非固定在测序芯片上,因此当为芯片加装遮光片后,请小心操作。

步骤结束

Close the device lid and continue sequencing run on MinKNOW.

提示

Library storage recommendations

We recommend storing libraries in Eppendorf DNA LoBind tubes at 4°C for short term storage or repeated use, for example, reloading flow cells between washes. For single use and long-term storage of more than 3 months, we recommend storing libraries at -80°C in Eppendorf DNA LoBind tubes. For further information, please refer to the DNA library stability Know-How document.

5. To store the MinION/GridION Flow Cell for later use

材料
  • 测序芯片清洗剂盒(EXP-WSH004)或测序芯片清洗试剂盒 XL(EXP-WSH004-XL)
仪器
  • P20 移液枪和枪头
  • P1000移液枪和枪头

Storage Buffer (S) can be used to flush flow cells for storage for later use or to check number of available nanopores before loading another library.

Thaw one tube of Storage Buffer (S) at room temperature.

Mix contents thoroughly by pipetting and spin down briefly.

Slide the flow cell priming port cover clockwise to open.

Flow Cell Loading Diagrams Step 2 (3)

After opening the priming port, check for a small air bubble under the cover. Draw back a small volume to remove any bubbles:

  1. Set a P1000 pipette to 200 µl.
  2. Insert the tip into the flow cell priming port.
  3. Turn the wheel until the dial shows 220-230 µl, or until you can see a small volume of buffer/liquid entering the pipette tip.
  4. Visually check that there is continuous buffer from the flow cell priming port across the sensor array.

Flow Cell Loading Diagrams Step 03 V5

Slowly add 500 μl of Storage Buffer (S) through the flow cell priming port.

Close the priming port.

Flow Cell Loading Diagrams Step 9

Remove all fluid from the waste channel through waste port 1 using a P1000 pipette.

As both the flow cell priming port and SpotON sample port are closed, no fluid should leave the sensor array area.

Flow cell ports

重要

It is vital that the flow cell priming port and SpotON sample port are closed before removing the waste buffer to prevent air from being drawn across the sensor array area, which would lead to a significant loss of sequencing channels.

The flow cell can now be stored at 4-8°C.

步骤结束

When you wish to reuse the flow cell, remove the flow cell from storage, and allow it to warm to room temperature for ~5 minutes.

重要

After performing a flow cell wash or storing your flow cell, we recommend using running a 'Flow cell check' to check number of available nanopores.

Load your flow cell into the device with Storage Buffer (S) and start a Flow cell check to detect the number of active pores. For more information, please visit the Flow cell check section of our MinKNOW protocol.

After the Flow cell check, prime your flow cell and load the library before starting a new sequencing run.

Last updated: 4/25/2024

Document options

MinION