Oxford Nanopore Support | Oxford Nanopore Technologies
Community support
open menu
![a member of the Nanopore support team aiding a customer at a conference](https://a.storyblok.com/f/196663/328x143/0198140ca8/support-prebody.webp/m/filters:format(webp))
Contact support
You can contact our Customer Services and Technical Services teams round the clock using the contact support function.
Why is a clean-up step required after the DNA repair and end-prep for NBD114.24, but not for NBD114.96?
Why is a clean-up step required after the DNA repair and end-prep for NBD114.24, but not for NBD114.96?
This is due to the different volume of end-prep DNA taken forward between the two different protocols (7.5 µL vs 0.75 µL for NBD114.24 and NBD114.96 respectively). A clean-up is required for NBD114.24 to limit carry over of enzymes from the UltraII and FFPE repair kits.