VolTRAX V2b


Descripción general

For Research Use Only

Document version: VTX_2004_v1_revV_08Feb2019

1. Introduction to VolTRAX V2b

IMPORTANTE

VolTRAX upgrade path

In early Q2 2022, the VolTRAX V2 devices in field began to be phased out and replaced by VolTRAX V2b. The VolTRAX V2b Starter Packs contain:

  • 1 VolTRAX V2b device
  • 1 VolTRAX Configuration Kit (VCK-V2003b), including two blue cartridges
  • 4 VolTRAX Cartridge Packs (VCT-V2002), each pack contains three blue cartridges
  • 4 VolTRAX Sequencing Kits (VSK-VSK004), each kit is sufficient for three reactions
  • 6 MinION Flow Cells
  • 1 USB3.1 / USBC cable

Also in 2022, five new products were introduced that can be purchased individually:

  • VolTRAX Sequencing Kit (VSK-VSK004)
  • VolTRAX PCR Tiling 1-12 COVID-19 Kit (VSK-PTC001)
  • VolTRAX RT-PCR Sequencing Kit 1-12 (VSK-VPS001)
  • VolTRAX Multiplex Kit (VSK-VMK004)
  • VolTRAX Cartridge Packs (VCT-V2002D), containing blue cartridges

Introduction to VolTRAX V2b

The VolTRAX V2b is a portable, multi-purpose device designed to automatically prepare DNA or RNA libraries ready for analysis on a nanopore sequencing platform. The device is compact and USB-powered, to allow portability of library preparation and sequencing in a laboratory or non-laboratory environment.

VolTRAX Right 45 Elevated Full Cable Transparency

The VolTRAX V2b with a disposable cartridge inserted.

Simple, automated, and portable library preparation of VolTRAX V2b

The VolTRAX V2b device along with the cartridge allows a user to choose a library preparation protocol, load their sample with supplied reagents and press go. The device prepares the library following a series of steps programmed by the software.

The device includes heating elements for incubations and PCR, and magnetic elements for bead-based separations. These features combine with the VolTRAX capabilities to precisely move, mix and incubate droplets to enable both basic and complex sample preparation methods to be carried out with a click of a button.

VolTRAX V2b is simple to use and needs only minutes of hands-on time to prepare a library.

2. Laboratory equipment and consumables

Material
  • VolTRAX Configuration Kit (VCK-V2003b)
  • VolTRAX Sequencing Kit (VSK-VSK004)
  • VolTRAX Multiplex Kit (VSK-VMK004)
  • VolTRAX PCR Tiling 1-12 COVID-19 Kit (VSK-PTC001)
  • VolTRAX RT-PCR Sequencing Kit 1-12 (VSK-VPS001)
  • VolTRAX Cartridge Pack (VCT-V2002B)

Consumibles
  • VolTRAX cartridge
  • MinION Flow Cell
  • Tubos de 1,5 ml Eppendorf DNA LoBind
  • 0.2 ml thin-walled PCR tubes

Instrumental
  • VolTRAX V2b
  • Pipeta y puntas P20
  • Pipeta y puntas P10
  • MinION
IMPORTANTE

VolTRAX V2b pipette tip compatibility

We have found that certain pipette tips give better results during the reagent loading and library extraction. Using the wrong tip type can lead to sample loss. The Mettler Toledo (Rainin™) UNV 10 µl tips have been verified as having the correct geometry for loading reagents on the VolTRAX cartridge. We recommend the Mettler Toledo (Rainin™) UNV 20 µl tips for extracting the prepared library at the end of the run.

Sample requirements to prepare high-quality libraries (see individual kit sections for details)

The VolTRAX Sequencing Kit (VSK-VSK004) and VolTRAX Multiplex Kit (VSK-VMK004) protocols for VolTRAX V2b require clean, extracted DNA that is of high molecular weight, accurately-quantified and free from contaminants and detergents. The DNA should be prepared in 10 mM Tris-HCl, pH 8.0.

The VolTRAX PCR Tiling 1-12 COVID-19 Kit (VSK-PTC001) and VolTRAX RT-PCR Sequencing Kit 1-12 protocol requires total RNA in 10 mM Tris-HCl, pH 8.0, extracted from samples that have been screened by a suitable qPCR assay. Samples must have a Ct value of 30 or less, and be free from contaminants and detergents.

Contaminants

Contaminants can impede the function of the library preparation chemistry for VolTRAX V2b. For information about the impact of contaminants on reaction efficiency, see the Contaminants Know-How piece.

3. Computer requirements for using VolTRAX V2b

Computer requirements for VolTRAX V2b

VolTRAX V2b is driven by the VolTRAX software, which is responsible for the control of the device and consumable cartridge array.

VolTRAX V2b computer requirement is less demanding than a MinION™. If you are currently running a MinION Mk1B on Windows, or you plan to use a MinION Mk1B with VolTRAX, the computer requirements for your MinION Mk1B should be sufficient to operate the VolTRAX.

If you plan to purchase a system dedicated to VolTRAX V2b, the minimum requirements are:

Component Minimum specification
Operating system Windows 10
macOS and Linux are in development.
Memory/RAM 2 GB RAM
CPU 64-bit dual core processor
Storage/hard drive > 128 GB SSD
User account privilege level Local administrator
Ports USB TYPE C, 5 V, 3 A †
Antivirus software Installation highly recommended
Internet connection Appropriate firewall settings are required for downloading the VolTRAX software.

† Only the USB TYPE C – Gen 3.1 (I or II) or higher – 5 V, 3 A can provide the full range of functionalities on VolTRAX V2b.

The VolTRAX status is displayed by LEDs

To ensure users can easily determine their VolTRAX status; a series of LEDs are included and described below.

Under-side LEDs:

The LEDs shown in the figure are indicators of the power supplied to the device.

LED labelled diagram - development use only 06 Sep 19 2.0

  1. Red - This indicates that the USB Type-C cable is correctly inserted into the port on the device. This is for Development use only.
  2. Red - This LED signals that 900 mA are being supplied to the device; this is the least powerful type of USB-Type C or 3.1 Type I/II port. The device will operate basic protocols, but will not be able to perform thermal cycling required for PCR-based library preparations.
  3. Blue - This LED signals that 1.5 A are being supplied to the device; this is a commonly occurring power supply of USB-Type C ports. The device will operate basic protocols, but will not be able to perform thermal cycling required for PCR-based library preparations.
  4. Green - This LED signals 3 A are being supplied to the device; this is the requirement for the full capabilities of the VolTRAX V2B device.
  5. Blue - This LED signals power to the device.
  6. Blue - This LED signals power to the device.
  7. These are for use by Oxford Nanopore Technologies.

It is particularly important that LED 4 (green) is illuminated, if the user requires the thermal cycler to prepare their library. If the LED is not on, 3 A are not being supplied to the device.

To check that the laptop to be used with the VolTRAX has a port that supplies 3 A and that the power rating is set to 3 A, contact your IT Department.

Internally at Oxford Nanopore Technologies, we have used USB-power meters as a way of measuring the current provided by a USB port when a laptop is running off charge or plugged in. Users may wish to look into these in advance of their VolTRAX devices arriving.

Configurar el cortafuegos en los dispositivos de Oxford Nanopore Technologies

El programa de Oxford Nanopore Technologies pedirá acceso a los rangos IP de AWS enumerados a continuación:

http://docs.aws.amazon.com/general/latest/gr/aws-ip-ranges.html

También se necesitará acceso a las siguientes direcciones IP:

              178.79.175.200  
               96.126.99.215  

Updates

VolTRAX software is under rapid development and we ask users to keep up-to-date with the latest software revisions.

Latest software versions will be available on our software downloads page.

Microsoft Windows updates

It is essential to check the settings for Windows updates. Windows Updates Automatic restarts need to be disabled and it is recommended that Windows Update installs are set to occur on shutdown only.

4. Installing the VolTRAX V2b software on Windows

ATENCIÓN

You must not plug the VolTRAX V2b into the designated laptop before the VolTRAX Client is installed, operational, and instructs you to.

To download the VolTRAX Client to run a VolTRAX V2b, follow the instructions below:

The most current version of VolTRAX Client is available from the Software Downloads page.

Note: You will need Administrative rights to install the VolTRAX Client and VolTRAX firmware. The VolTRAX Client is currently only supported on Windows 10 operating systems.

The executable (.exe) will be downloaded to your default downloads folder. Unless otherwise specified, this is the 'Downloads' folder on the C:\ drive.

Double-click on the VolTRAX installer executable to begin installation.

The installer offers two choices:

  1. "Anyone who uses this computer (all users)", which will install to C:\Program Files\OxfordNanopore\VolTRAX
  2. "Only for me", which will install to C:\Users\[username]\AppData\Local\OxfordNanopore\VolTRAX

Screenshot 2022-03-16 at 15.26.19

Once installation is complete, double-click the VolTRAX Client shortcut on the desktop.

The VolTRAX Client will load a window. Click "Start VolTRAX".

Start VolTRAX

CONSEJO

The Windows Firewall may interrupt installation.

The Windows Defender Firewall, or similar, may block the download of some of the features of the application.

If this is the case, contact your IT department to enable the full download of the VolTRAX Client.

firewall

The VolTRAX Client will load up the protocol selector screen, where the VolTRAX Client will lead you through the requirements to execute the selected method.

The Protocol Selector screen provides the supported protocol options that can be run with VolTRAX V2b.

0.21.0 Home Screen

IMPORTANTE

Default output folder

For both types of installation (whether the software is installed for "Anyone who uses this computer" or "Only for me"), by default the output data is logged to:

C:\Users\[username]\AppData\Local\VolTRAX\run\

5. Running a Cartridge Test for VolTRAX V2b

Material
  • VolTRAX Configuration Kit (VCK-V2003b)
Instrumental
  • VolTRAX V2b

The Cartridge Test may be used to pre-screen cartridges prior to use. The Cartridge Test is also performed as part of each kit workflow.

Once the software is running, plug in your VolTRAX device. Click the "Start" button.

You will be presented with the following options for protocols to run. Select "Cartridge Test" to start the protocol and follow the instructions in the software.

0.21.0 Home Screen Cartridge test

CONSEJO

Please refer to the videos displayed in the VolTRAX UI, which demonstrate how to perform each step in the protocol.

IMPORTANTE

If the Cartridge Test fails:

  1. Removing the cartridge as shown in the animation in the GUI, re-insert the cartridge and repeat the Cartridge Test.
  2. If the Cartridge Test fails a second time, contact Customer Support at support@nanoporetech.com

Picture2

CONSEJO

Cancelling a run

To cancel a run, select Cancel in the top right corner of the GUI.

The user is able to look through all the steps of the selected protocol before confirming cancellation.

The run will stop immediately and return the user to the "Start VolTRAX" page. Another run can be started immediately.

Note: The cartridge must be replaced if reagents have been input during a cancelled run. If no reagent has been inserted, the cartridge can be used again.

CONSEJO

Run history

The UI provides a run history option to view previously-run protocols, which can be accessed from the link in the top right corner.

Run history

Clicking Open on any of the protocols will navigate the user to the relevant data folder in Windows File Explorer:

Run history2

6. Running a VolTRAX Configuration Kit

Material
  • VolTRAX Configuration Kit (VCK-V2003b)
Instrumental
  • VolTRAX V2b
  • Pipeta y puntas P20
  • Pipeta y puntas P10

The VolTRAX Configuration Kit (VCK-V2003b) enables the practice of loading and extracting material from the VolTRAX V2b cartridge.

Below is an overview of the configuration experiment using a VolTRAX V2b with the Configuration Kit (VCK-V2003b).

An overview of VolTRAX V2b configuration

1. Loading a protocol

  • Use the protocol selector in the UI to load the VCK-V2003b protocol.
  • When prompted, connect your VolTRAX V2b device, press Continue and await connection.

2. System QC - Cartridge Test

  • When prompted, connect a cartridge to the VolTRAX device and press Continue to proceed to the Cartridge Test.

3. Running and loading VCS

  • Follow instructions and advice on the equipment required and pipetting technique
  • Prime the cartridge as instructed by the UI
  • Load the VolTRAX Configuration Solution (VCS) and the VX AMPure XP Config (VXC) into the ports as instructed by the UI
  • The solutions will be manipulated to indicate what the user should expect during a run

4. Extracting the VolTRAX Configuration Solution from the cartridge

  • Follow the extraction instructions, using UI as a visual guide, to slowly extract the VCS
  • Discard the extracted VCS

Start the VolTRAX software. You will be presented with the following options for protocols to run. Select "Configuration Kit" to start the protocol and follow the instructions in the software.

0.21.0 Home Screen Configuration test

Protocol preview

Once an option is selected from the menu, the UI will show a preview of the steps involved in library preparation. The example below is for the Configuration Kit.

Protocol Preview Dialog

When prompted by the UI, connect your VolTRAX device to the computer. When it is connected, you will then be prompted to insert a cartridge into the device.

Prime the cartridge as instructed in the UI by twisting the Priming Fluid container anti-clockwise and removing the foil tab to allow the fluid to enter the cartridge.

Select Continue when each step is completed.

Follow the steps as described in the UI, loading VolTRAX Configuration Solution and VX AMPure XP Config into the wells instead of the reagents stated.

Select Continue after each step. After all the steps have been completed, the library prep process will start automatically.

CONSEJO

Please refer to the videos displayed in the VolTRAX UI, which demonstrate how to perform each step in the protocol.

CONSEJO

Loading solutions into the VolTRAX V2b cartridge

The VolTRAX V2b device can measure solution volumes loaded into the cartridge. The GUI displays a coloured-pixel grid to show:

  • Blue: Volume of solution too low. You will need to load more solution to proceed with the protocol.
  • Green: Correct volume of solution.
  • Orange: slightly overfilled. You will still be able to proceed with the protocol.
  • Red: solution overfilled. You will need to cancel the experiment and start again.

Volume colours

FIN DEL PROCESO

At the end of the experiment, the VolTRAX Configuration Solution can be extracted from the E port of the VolTRAX V2b cartridge. This can be discarded.

CONSEJO

Cancelling a run

To cancel a run, select Cancel in the top right corner of the GUI.

The user is able to look through all the steps of the selected protocol before confirming cancellation.

The run will stop immediately and return the user to the "Start VolTRAX" page. Another run can be started immediately.

Note: The cartridge must be replaced if reagents have been input during a cancelled run. If no reagent has been inserted, the cartridge can be used again.

CONSEJO

Run history

The UI provides a run history option to view previously-run protocols, which can be accessed from the link in the top right corner.

Run history

Clicking Open on any of the protocols will navigate the user to the relevant data folder in Windows File Explorer:

Run history2

7. Overview of the VolTRAX Sequencing Kit

The VSK-VSK004 DNA library preparation kit

The VolTRAX V2b provides an automated process for preparing a DNA library. The chemistry of the library preparation is the same as for manual library preparation for a MinION or GridION sequencing experiment.

Below is an overview of the library preparation stages using a VolTRAX V2b, blue cartridge and the VolTRAX Sequencing Kit (VSK-VSK004).

Overview of the VolTRAX Sequencing Kit protocol

1. Loading a protocol

  • Use the protocol selector in the UI to load the VSK-VSK004 protocol.
  • When prompted, connect your VolTRAX V2b device, press Continue and await connection.

2. System QC - Cartridge Test

  • When prompted, connect a cartridge to the VolTRAX device and press Continue to proceed to the Cartridge Test.

3. Loading the Priming Fluid and reagents into the cartridge

  • Prime the cartridge by twisting the oil container and removing the foil tab.
  • The UI will recommend the quality, concentration, buffer and Absorbance (A) readings for your input DNA. These are stated at the stage of 'Reagent Loading' later in this protocol.
  • Follow the instructions on the equipment required and pipetting technique.
  • Load the appropriate reagents and extracted DNA into the ports as instructed by the UI.
  • The reagents will be manipulated by the VolTRAX V2b to prepare the library for nanopore sequencing.

4. Extracting the library from the cartridge

  • Follow the extraction instructions, using UI as a visual guide, to slowly extract the library from the cartridge, then transfer it into an Eppendorf tube.
  • Prepare the library for loading onto the flow cell.

VolTRAX Sequencing Kit (VSK-VSK004) contents

VSK-VSK004

Name Acronym Cap colour Number of vials Fill volume (μl)
VX Fragmentation Mix VFA Amber 1 25
EDTA EDTA Clear 1 20
VX AMPure XP Beads VXP Pink 1 45
VX Long Fragment Buffer VLF Orange 1 40
VX Rapid Adapter T VRA T Green 1 25
VX Elution Buffer VEB Black 1 65
VX Running Buffer VRB Red 1 250
Flush Buffer FB Blue 3 1,170
Flush Tether FLT White cap, purple label 1 200

VSK-VSK004 chemistry

The VSK-VSK004 library preparation protocol applies our rapid chemistry, with a bead-based purification step.

VSK-VSK002 - Work Flow - chemistry for V2

8. Run a VolTRAX Sequencing Kit

Material
  • 9.5 μl DNA in 10 mM Tris-HCl, pH 8.0 at a DNA concentration of 50 ng/μl (475 ng total)
  • VolTRAX Sequencing Kit (VSK-VSK004)
  • VolTRAX Cartridge Pack (VCT-V2002B)

Consumibles
  • Tubos de PCR de pared fina (0,2 ml)
  • Qubit dsDNA HS Assay Kit (ThermoFisher, Q32851)

Instrumental
  • VolTRAX V2b
  • Pipeta y puntas P20
  • Pipeta y puntas P10
Equipo opcional
  • Fluorímetro Qubit (o equivalente para el control de calidad)
IMPORTANTE

VolTRAX V2b pipette tip compatibility

We have found that certain pipette tips give better results during the reagent loading and library extraction. Using the wrong tip type can lead to sample loss. The Mettler Toledo (Rainin™) UNV 10 µl tips have been verified as having the correct geometry for loading reagents on the VolTRAX cartridge. We recommend the Mettler Toledo (Rainin™) UNV 20 µl tips for extracting the prepared library at the end of the run.

Start the VolTRAX software. You will be presented with the following options for protocols to run. Select "Sequencing Kit" to start the protocol and follow the instructions in the software.

0.21.0 Home Screen Sequencing Kit

Protocol preview

Once an option is selected from the menu, the UI will show a preview of the steps involved in library preparation. The example below is for the Configuration Kit.

Protocol Preview Dialog

When prompted by the UI, connect your VolTRAX device to the computer. When it is connected, you will then be prompted to insert a cartridge into the device.

Prime the cartridge as instructed in the UI by twisting the Priming Fluid container anti-clockwise and removing the foil tab to allow the fluid to enter the cartridge.

Select Continue when each step is completed.

CONSEJO

Please refer to the videos displayed in the VolTRAX UI, which demonstrate how to perform each step in the protocol.

CONSEJO

Loading solutions into the VolTRAX V2b cartridge

The VolTRAX V2b device can measure solution volumes loaded into the cartridge. The GUI displays a coloured-pixel grid to show:

  • Blue: Volume of solution too low. You will need to load more solution to proceed with the protocol.
  • Green: Correct volume of solution.
  • Orange: slightly overfilled. You will still be able to proceed with the protocol.
  • Red: solution overfilled. You will need to cancel the experiment and start again.

Volume colours

Wait for the library prep to finish.

A timer can be found on the right hand side of the UI, indicating the remaining time.

CONSEJO

Cancelling a run

To cancel a run, select Cancel in the top right corner of the GUI.

The user is able to look through all the steps of the selected protocol before confirming cancellation.

The run will stop immediately and return the user to the "Start VolTRAX" page. Another run can be started immediately.

Note: The cartridge must be replaced if reagents have been input during a cancelled run. If no reagent has been inserted, the cartridge can be used again.

At the end of the experiment, extract the prepared library from the E port of the VolTRAX cartridge. Dispense the library into a 0.2 ml PCR tube for preparation of the sequencing mix, as discussed in the following section.

CHECKPOINT

Quantify 1 µl of eluted sample using a Qubit fluorometer.

CONSEJO

Run history

The UI provides a run history option to view previously-run protocols, which can be accessed from the link in the top right corner.

Run history

Clicking Open on any of the protocols will navigate the user to the relevant data folder in Windows File Explorer:

Run history2

9. Priming and loading the SpotON flow cell

Material
  • Flush Buffer (FB)
  • Flush Tether (FLT)

Consumibles
  • MinION Flow Cell
  • Tubos de 1,5 ml Eppendorf DNA LoBind

Instrumental
  • Pipeta y puntas P1000
  • Pipeta y puntas P100

Open the MinION Mk1B lid and slide the flow cell under the clip.

Press down firmly on the flow cell to ensure correct thermal and electrical contact.

Flow Cell Loading Diagrams Step 1a

Flow Cell Loading Diagrams Step 1b

MEDIDA OPCIONAL

Antes de cargar la biblioteca, verifique la celda de flujo para determinar el número de poros disponible.

Si se ha verificado con anterioridad la cantidad de poros presentes en la celda de flujo, este paso se puede omitir.

Dispone de más información en las instrucciones de comprobación de la celda de flujo, del protocolo de MinKNOW.

Slide the priming port cover clockwise to open the priming port.

Flow Cell Loading Diagrams Step 2

IMPORTANTE

Tenga cuidado a la hora de extraer el tampón de la celda de flujo. No retire más de 20-30 μl y asegúrese de que el tampón cubra la matriz de poros en todo momento. La introducción de burbujas de aire en la matriz puede dañar los poros de manera irreversible.

After opening the priming port, check for a small air bubble under the cover. Draw back a small volume to remove any bubbles (a few µls).

  1. Set a P1000 pipette to 200 µl
  2. Insert the tip into the priming port
  3. Turn the wheel until the dial shows 220-230 ul, to draw back 20-30 ul, or until you can see a small volume of buffer entering the pipette tip

Note: Visually check that there is continuous buffer from the priming port across the sensor array.

Flow Cell Loading Diagrams Step 03 V5

Cargar 800 μl de solución en el puerto de cebado, evitando introducir burbujas de aire. Esperar 5 minutos. Durante este tiempo, preparar la biblioteca para cargar siguiendo los pasos a continuación.

Flow Cell Loading Diagrams Step 04 V5 SPANISH

To prepare the flow cell priming mix, add 30 µl of thawed and mixed Flush Tether (FLT) directly to the tube of thawed and mixed Flush Buffer (FB), and mix by vortexing at room temperature.

Thoroughly mix the contents of the VX Running Buffer (VRB) tube by pipetting up and down.

In a new tube, prepare the library for loading as follows:

Reagent Volume per flow cell
VX Running Buffer (VRB) 64 µl
DNA library 11 µl
Total 75 µl

Completar el cebado de la celda de flujo:

  1. Levantar suavemente la tapa del puerto de carga SpotON.
  2. Cargar 200 µl de solución en el puerto de cebado (no en el puerto de muestra SpotON), evitando introducir burbujas de aire.

Flow Cell Loading Diagrams Step 5

Flow Cell Loading Diagrams Step 06 V5 SPANISH 2

Mezclar la biblioteca pipeteando suavemente, justo antes de cargar.

Añadir, gota a gota, 75 μl de la biblioteca preparada en el puerto de muestra SpotON. Procurar que cada gota fluya hacia adentro del puerto antes de añadir la siguiente.

Flow Cell Loading Diagram Step 07 V5 SPANISH

Gently replace the SpotON sample port cover, making sure the bung enters the SpotON port, close the priming port and close the lid of the sequencing device.

Flow Cell Loading Diagrams Step 8

Flow Cell Loading Diagrams Step 9

FIN DEL PROCESO

You are ready to begin your sequencing experiment.

Please refer to the MinKNOW protocol for instructions on setting up your sequencing experiment.

10. An overview of the VolTRAX Multiplex Kit

The VSK-VMK004 DNA library preparation kit

The VolTRAX V2b provides an automated process for preparing a DNA library. The chemistry of the library preparation is the same as for manual library preparation for a MinION or GridION sequencing experiment.

Below is an overview of the library preparation stages using a VolTRAX V2b and the VolTRAX Multiplex Kit (VSK-VMK004).

VolTRAX Multiplex Kit (VSK-VMK004) contents

VSK-VMK004 tubes

Name Acronym Number of vials Cap colour Fill volume per vial (µl)
VolTRAX Buffer Solution VBS 1 Orange 100
EDTA EDTA 1 Clear 20
VX AMPure XP Beads VXP 1 Pink 45
VX S Fragment Buffer VSF 1 White 30
VX Rapid Adapter T VRA T 1 Green 25
VX Elution Buffer VEB 1 Black 65
VX Running Buffer VRB 1 Red 250
VX BC Frag Mix 01-10 VB01-10 10 Amber 20
Flush Buffer FB 3 Blue 1,170
Flush Tether FLT 1 Purple 200

VolTRAX Multiplex Kit barcode sequences

Component Sequence
VB01 AAGAAAGTTGTCGGTGTCTTTGTG
VB02 TCGATTCCGTTTGTAGTCGTCTGT
VB03 GAGTCTTGTGTCCCAGTTACCAGG
VB04 TTCGGATTCTATCGTGTTTCCCTA
VB05 CTTGTCCAGGGTTTGTGTAACCTT
VB06 TTCTCGCAAAGGCAGAAAGTAGTC
VB07 GTGTTACCGTGGGAATGAATCCTT
VB08 TTCAGGGAACAAACCAAGTTACGT
VB09 AACTAGGCACAGCGAGTCTTGGTT
VB10 AAGCGTTGAAACCTTTGTCCTCTC

VolTRAX Multiplex Kit chemistry

The VSK-VMK004 library preparation protocol uses our rapid chemistry, with a bead-based clean-up step. The kit contains ten barcodes for multiplexing up to ten samples in one library preparation.

2019-11-14 VMK002 workflow v1 DS

An overview of the VolTRAX Multiplex Kit protocol

1. Loading a protocol

  • Use the protocol selector in the UI to load the VSK-VMK004 protocol.
  • When prompted, connect your VolTRAX V2b device, press Continue and await connection.

2. System QC - Cartridge Test

  • When prompted, connect a cartridge to the VolTRAX device and press Continue to proceed to the Cartridge Test.

3. Loading the Priming Fluid and reagents into the cartridge

  • Prime the cartridge by twisting the oil container and removing the foil tab.
  • The UI will recommend the quality, concentration, buffer and Absorbance (A) readings for your input DNA. These are stated at the stage of 'Reagent Loading' later in this protocol.
  • Follow the instructions on the equipment required and pipetting technique.
  • Load the appropriate reagents and extracted DNA into the ports as instructed by the UI.
  • The reagents will be manipulated by the VolTRAX V2b to prepare the library for nanopore sequencing.

4. Extracting the library from the cartridge

  • Follow the extraction instructions, using UI as a visual guide, to slowly extract the library from the cartridge, then transfer it into an Eppendorf tube.
  • Prepare the library for loading onto the flow cell.

11. Run a VolTRAX Multiplex Kit

Material
  • VolTRAX Multiplex Kit (VSK-VMK004)
  • VolTRAX Cartridge Pack (VCT-V2002B)
  • 1.8 μl of extracted gDNA in 10 mM Tris-HCl, pH 8.0 at a DNA concentration of 50 ng/μl (90 ng total) for each sample to be barcoded

Consumibles
  • Tubos de PCR de pared fina (0,2 ml)
  • Qubit dsDNA HS Assay Kit (ThermoFisher, Q32851)

Instrumental
  • VolTRAX V2b
  • Pipeta y puntas P20
  • Pipeta y puntas P2
  • Fluorímetro Qubit (o equivalente para el control de calidad)
IMPORTANTE

VolTRAX V2b pipette tip compatibility

We have found that certain pipette tips give better results during the reagent loading and library extraction. Using the wrong tip type can lead to sample loss. The Mettler Toledo (Rainin™) UNV 10 µl tips have been verified as having the correct geometry for loading reagents on the VolTRAX cartridge. We recommend the Mettler Toledo (Rainin™) UNV 20 µl tips for extracting the prepared library at the end of the run.

Start the VolTRAX software. You will be presented with the following options for protocols to run. Select "VolTRAX Multiplex Kit" to start the protocol and follow the instructions in the software.

0.21.0 Home Screen Multiplex kit

Protocol preview

Once an option is selected from the menu, the UI will show a preview of the steps involved in library preparation. The example below is for the Configuration Kit.

Protocol Preview Dialog

When prompted by the UI, connect your VolTRAX device to the computer. When it is connected, you will then be prompted to insert a cartridge into the device.

Once the software is running, plug in your VolTRAX device. Click the "Start" button.

You will be presented with five options for test experiments to run. Select "Multiplexing Kit" to start the workflow.

VolTRAX 0.16 VMK002

Insert the cartridge firmly into the device until you hear a click. From here on, do not remove the cartridge until the end of the experiment.

Capture 4

Prime the cartridge as instructed in the UI by twisting the Priming Fluid container anti-clockwise and removing the foil tab to allow the fluid to enter the cartridge.

Select Continue when each step is completed.

CONSEJO

Please refer to the videos displayed in the VolTRAX UI, which demonstrate how to perform each step in the protocol.

CONSEJO

Loading solutions into the VolTRAX V2b cartridge

The VolTRAX V2b device can measure solution volumes loaded into the cartridge. The GUI displays a coloured-pixel grid to show:

  • Blue: Volume of solution too low. You will need to load more solution to proceed with the protocol.
  • Green: Correct volume of solution.
  • Orange: slightly overfilled. You will still be able to proceed with the protocol.
  • Red: solution overfilled. You will need to cancel the experiment and start again.

Volume colours

Select the number of samples you are preparing and assign sample aliases to barcodes.

VMK Sample Selection Screen

Once you have entered the required details, click Confirm. You will be taken to the Sample Label Check page where you can confirm that the details are correct.

VMK Confirm Selection

If you are running fewer than 10 samples, you can adjust the "Sample number". Add VBS (VolTRAX Buffer Solution) to wells that will be without a sample.

VMK 6 Sample Confirm Selection

Wait for the library prep to finish.

A timer can be found on the right hand side of the UI, indicating the remaining time.

CONSEJO

Cancelling a run

To cancel a run, select Cancel in the top right corner of the GUI.

The user is able to look through all the steps of the selected protocol before confirming cancellation.

The run will stop immediately and return the user to the "Start VolTRAX" page. Another run can be started immediately.

Note: The cartridge must be replaced if reagents have been input during a cancelled run. If no reagent has been inserted, the cartridge can be used again.

At the end of the experiment, extract the prepared library from the E port of the VolTRAX cartridge. Dispense the library into a 0.2 ml PCR tube for preparation of the sequencing mix, as discussed in the following section.

MEDIDA OPCIONAL

Quantify 1 µl of eluted sample using a Qubit fluorometer.

CONSEJO

Run history

The UI provides a run history option to view previously-run protocols, which can be accessed from the link in the top right corner.

Run history

Clicking Open on any of the protocols will navigate the user to the relevant data folder in Windows File Explorer:

Run history2

12. Priming and loading the SpotON flow cell

Material
  • Flush Buffer (FB)
  • Flush Tether (FLT)

Consumibles
  • Celda de flujo MinION/GridION
  • Tubos de 1,5 ml Eppendorf DNA LoBind

Instrumental
  • Pipeta y puntas P1000
  • Pipeta y puntas P100

Open the MinION Mk1B lid and slide the flow cell under the clip.

Press down firmly on the flow cell to ensure correct thermal and electrical contact.

image

Slide the priming port cover clockwise to open the priming port.

open-priming-port

How to prime and load the SpotON Flow Cell

Priming and loading: The steps for priming and loading the SpotON Flow Cell. Written instructions are given below. The library is loaded dropwise without putting the pipette tip firmly into the port.

Take care to avoid introducing any air during pipetting.

IMPORTANTE

Tenga cuidado a la hora de extraer el tampón de la celda de flujo. No retire más de 20-30 μl y asegúrese de que el tampón cubra la matriz de poros en todo momento. La introducción de burbujas de aire en la matriz puede dañar los poros de manera irreversible.

After opening the priming port, check for a small air bubble under the cover. Draw back a small volume to remove any bubbles (a few µls).

  1. Set a P1000 pipette to 200 µl
  2. Insert the tip into the priming port
  3. Turn the wheel until the dial shows 220-230 ul, to draw back 20-30 ul, or until you can see a small volume of buffer entering the pipette tip

Note: Visually check that there is continuous buffer from the priming port across the sensor array.

Flow Cell Loading Diagrams Step 03 V5

To prepare the flow cell priming mix, add 30 µl of thawed and mixed Flush Tether (FLT) directly to the tube of thawed and mixed Flush Buffer (FB), and mix by vortexing at room temperature.

Cargar 800 μl de solución en el puerto de cebado, evitando introducir burbujas de aire. Esperar 5 minutos. Durante este tiempo, preparar la biblioteca para cargar siguiendo los pasos a continuación.

Flow Cell Loading Diagrams Step 04 V5 SPANISH

Thoroughly mix the contents of the VX Running Buffer (VRB) tube by pipetting up and down.

In a new tube, prepare the library for loading as follows:

Reagent Volume per flow cell
VX Running Buffer (VRB) 64 µl
DNA library 11 µl
Total 75 µl

Completar el cebado de la celda de flujo:

  1. Levantar suavemente la tapa del puerto de carga SpotON.
  2. Cargar 200 µl de solución en el puerto de cebado (no en el puerto de muestra SpotON), evitando introducir burbujas de aire.

Flow Cell Loading Diagrams Step 5

Flow Cell Loading Diagrams Step 06 V5 SPANISH 2

Mezclar la biblioteca pipeteando suavemente, justo antes de cargar.

Añadir, gota a gota, 75 μl de la biblioteca preparada en el puerto de muestra SpotON. Procurar que cada gota fluya hacia adentro del puerto antes de añadir la siguiente.

Flow Cell Loading Diagram Step 07 V5 SPANISH

Gently replace the SpotON sample port cover, making sure the bung enters the SpotON port, close the priming port and close the lid of the sequencing device.

Flow Cell Loading Diagrams Step 8

Flow Cell Loading Diagrams Step 9

FIN DEL PROCESO

You are ready to begin your sequencing experiment.

Please refer to the MinKNOW protocol for instructions on setting up your sequencing experiment.

13. Overview of the VolTRAX PCR Tiling 1-12 COVID-19 Kit

The VSK-PTC001 RNA library preparation kit

The VolTRAX PCR Tiling 1-12 COVID-19 Kit generates tiled PCR amplicons from the SARS-CoV-2 viral RNA for whole genome sequencing using the VolTRAX V2b platform to improve consistency in library preparation due to the precise fluid handling of the device.

This kit provides all the reagents required to generate reverse-transcribed and PCR-amplified material from SARS-CoV-2 RNA samples using VolTRAX automation to prepare the library, without the need to purchase further components. The PCR tiling generated overlapping amplicons across the whole SARS-CoV-2 genome before barcoding and pooling multiple samples together for sequencing.

Below is an overview of the library preparation stages using a VolTRAX V2b, blue cartridges and the VolTRAX PCR Tiling 1-12 COVID-19 Kit (VSK-PTC001).

Overview of the VolTRAX PCR Tiling 1-12 COVID-19 protocol

1. Loading a protocol

  • Use the protocol selector in the UI to load the VSK-PTC001 protocol.
  • When prompted, connect your VolTRAX V2b device, press Continue and await connection.

2. System QC - Cartridge Test

  • When prompted, connect a cartridge to the VolTRAX device and press Continue to proceed to the Cartridge Test.

3. Loading the Priming Fluid and reagents into the cartridge

  • Follow the instructions on the equipment required and pipetting technique.
  • Prime the cartridge by twisting the oil container and removing the foil tab.
  • Load the appropriate reagents and extracted RNA into the ports as instructed by the UI.
  • The reagents will be manipulated by the VolTRAX V2b to prepare the library for nanopore sequencing.

4. Extracting the library from the cartridge

  • Follow the extraction instructions, using UI as a visual guide, to slowly extract the library from the cartridge, then transfer it into an Eppendorf tube.
  • Prepare the library for loading onto the flow cell.

VolTRAX PCR Tiling 1-12 COVID-19 Kit (VSK-PTC001) contents

VSK-PTC001 tubes

Name Acronym Number of vials Fill volume per vial (µl)
LunaScript RT SuperMix LS RT 1 10
Q5 HS Master Mix Q5 1 35
VX Midnight Primer A VMP A 1 15
VX Midnight Primer B VMP B 1 15
VX BC Frag Mix 01-12 VXB01-VXB12 12 10
VX AMPure XP Beads VXP 1 45
VX Rapid Adapter F VRA F 1 12
VX Elution Buffer VELB 1 50
VX Running Buffer VRB 1 250
Negative Control NEG 1 10
Flush Buffer FB 3 1170
Flush Tether FLT 1 200

VolTRAX PCR Tiling 1-12 COVID-19 Kit barcode sequences

Component Sequence
VXB01 AAGAAAGTTGTCGGTGTCTTTGTG
VXB02 TCGATTCCGTTTGTAGTCGTCTGT
VXB03 GAGTCTTGTGTCCCAGTTACCAGG
VXB04 TTCGGATTCTATCGTGTTTCCCTA
VXB05 CTTGTCCAGGGTTTGTGTAACCTT
VXB06 TTCTCGCAAAGGCAGAAAGTAGTC
VXB07 GTGTTACCGTGGGAATGAATCCTT
VXB08 TTCAGGGAACAAACCAAGTTACGT
VXB09 AACTAGGCACAGCGAGTCTTGGTT
VXB10 AAGCGTTGAAACCTTTGTCCTCTC
VXB11 GTTTCATCTATCGGAGGGAATGGA
VXB12 CAGGTAGAAAGAAGCAGAATCGGA

VSK-PTC001 chemistry

The VolTRAX PCR Tiling 1-12 COVID-19 Kit generates overlapping amplicons in a tiled fashion across the SARS-CoV-2 genome. First, the RNA samples are reverse-transcribed and then amplified by tiled PCR using separate primer pools. The primer pools are combined and SPRI purified before moving into the rapid barcoding stage which attaches Rapid Barcodes to the DNA ends for samples to be pooled.

VSK-PTC001 workflow v2

14. Run a VolTRAX PCR Tiling 1-12 COVID-19 Kit

Material
  • Input RNA in 10 mM Tris-HCl, pH 8.0
  • VolTRAX PCR Tiling 1-12 COVID-19 Kit (VSK-PTC001)
  • VolTRAX Cartridge Pack (VCT-V2002B)

Consumibles
  • Tubos de PCR de pared fina (0,2 ml)
  • Qubit dsDNA HS Assay Kit (ThermoFisher, Q32851)

Instrumental
  • VolTRAX V2b
  • Pipeta y puntas P20
  • Pipeta y puntas P10
Equipo opcional
  • Fluorímetro Qubit (o equivalente para el control de calidad)
IMPORTANTE

VolTRAX V2b pipette tip compatibility

We have found that certain pipette tips give better results during the reagent loading and library extraction. Using the wrong tip type can lead to sample loss. The Mettler Toledo (Rainin™) UNV 10 µl tips have been verified as having the correct geometry for loading reagents on the VolTRAX cartridge. We recommend the Mettler Toledo (Rainin™) UNV 20 µl tips for extracting the prepared library at the end of the run.

Board test

The Board test should be carried out on your VolTRAX device before proceeding with PCR experiments. The most recent models of the VolTRAX V2b feature improved temperature calibration for PCR, and the Board test checks whether you have a version of the device that is PCR-compatible.

Start the VolTRAX software. You will be presented with the following options for protocols to run. Select "Board Test" to start the protocol and follow the instructions in the software.

0.21.0 Home Screen Board test

Board test results

If the board test has been successful, you can proceed with PCR experiments.

Board Test Success

If the board test has failed, please contact Support.

Board Test Failed

Click "Continue", which will take you back to the protocol selection screen. Select "VolTRAX PCR Tiling 1-12 COVID-19" to start the protocol and follow the instructions in the software.

0.21.0 Home Screen COVID

Protocol preview

Once an option is selected from the menu, the UI will show a preview of the steps involved in library preparation. The example below is for the Configuration Kit.

Protocol Preview Dialog

When prompted by the UI, connect your VolTRAX device to the computer. When it is connected, you will then be prompted to insert a cartridge into the device.

Prime the cartridge as instructed in the UI by twisting the Priming Fluid container anti-clockwise and removing the foil tab to allow the fluid to enter the cartridge.

Select Continue when each step is completed.

CONSEJO

Please refer to the videos displayed in the VolTRAX UI, which demonstrate how to perform each step in the protocol.

CONSEJO

Loading solutions into the VolTRAX V2b cartridge

The VolTRAX V2b device can measure solution volumes loaded into the cartridge. The GUI displays a coloured-pixel grid to show:

  • Blue: Volume of solution too low. You will need to load more solution to proceed with the protocol.
  • Green: Correct volume of solution.
  • Orange: slightly overfilled. You will still be able to proceed with the protocol.
  • Red: solution overfilled. You will need to cancel the experiment and start again.

Volume colours

Wait for the library prep to finish.

A timer can be found on the right hand side of the UI, indicating the remaining time.

CONSEJO

Cancelling a run

To cancel a run, select Cancel in the top right corner of the GUI.

The user is able to look through all the steps of the selected protocol before confirming cancellation.

The run will stop immediately and return the user to the "Start VolTRAX" page. Another run can be started immediately.

Note: The cartridge must be replaced if reagents have been input during a cancelled run. If no reagent has been inserted, the cartridge can be used again.

At the end of the experiment, extract the prepared library from the E port of the VolTRAX cartridge. Dispense the library into a 0.2 ml PCR tube for preparation of the sequencing mix, as discussed in the following section.

MEDIDA OPCIONAL

Quantify 1 µl of eluted sample using a Qubit fluorometer.

CONSEJO

Run history

The UI provides a run history option to view previously-run protocols, which can be accessed from the link in the top right corner.

Run history

Clicking Open on any of the protocols will navigate the user to the relevant data folder in Windows File Explorer:

Run history2

15. Priming and loading the SpotON flow cell

Material
  • Flush Buffer (FB)
  • Flush Tether (FLT)

Consumibles
  • MinION Flow Cell
  • Tubos de 1,5 ml Eppendorf DNA LoBind

Instrumental
  • Pipeta y puntas P1000
  • Pipeta y puntas P100

Open the MinION Mk1B lid and slide the flow cell under the clip.

Press down firmly on the flow cell to ensure correct thermal and electrical contact.

Flow Cell Loading Diagrams Step 1a

Flow Cell Loading Diagrams Step 1b

MEDIDA OPCIONAL

Antes de cargar la biblioteca, verifique la celda de flujo para determinar el número de poros disponible.

Si se ha verificado con anterioridad la cantidad de poros presentes en la celda de flujo, este paso se puede omitir.

Dispone de más información en las instrucciones de comprobación de la celda de flujo, del protocolo de MinKNOW.

Slide the priming port cover clockwise to open the priming port.

Flow Cell Loading Diagrams Step 2

IMPORTANTE

Tenga cuidado a la hora de extraer el tampón de la celda de flujo. No retire más de 20-30 μl y asegúrese de que el tampón cubra la matriz de poros en todo momento. La introducción de burbujas de aire en la matriz puede dañar los poros de manera irreversible.

After opening the priming port, check for a small air bubble under the cover. Draw back a small volume to remove any bubbles (a few µls).

  1. Set a P1000 pipette to 200 µl
  2. Insert the tip into the priming port
  3. Turn the wheel until the dial shows 220-230 ul, to draw back 20-30 ul, or until you can see a small volume of buffer entering the pipette tip

Note: Visually check that there is continuous buffer from the priming port across the sensor array.

Flow Cell Loading Diagrams Step 03 V5

To prepare the flow cell priming mix, add 30 µl of thawed and mixed Flush Tether (FLT) directly to the tube of thawed and mixed Flush Buffer (FB), and mix by vortexing at room temperature.

Cargar 800 μl de solución en el puerto de cebado, evitando introducir burbujas de aire. Esperar 5 minutos. Durante este tiempo, preparar la biblioteca para cargar siguiendo los pasos a continuación.

Flow Cell Loading Diagrams Step 04 V5 SPANISH

Thoroughly mix the contents of the VX Running Buffer (VRB) tube by pipetting up and down.

In a new tube, prepare the library for loading as follows:

Reagent Volume per flow cell
VX Running Buffer (VRB) 64 µl
DNA library 11 µl
Total 75 µl

Completar el cebado de la celda de flujo:

  1. Levantar suavemente la tapa del puerto de carga SpotON.
  2. Cargar 200 µl de solución en el puerto de cebado (no en el puerto de muestra SpotON), evitando introducir burbujas de aire.

Flow Cell Loading Diagrams Step 5

Flow Cell Loading Diagrams Step 06 V5 SPANISH 2

Mezclar la biblioteca pipeteando suavemente, justo antes de cargar.

Añadir, gota a gota, 75 μl de la biblioteca preparada en el puerto de muestra SpotON. Procurar que cada gota fluya hacia adentro del puerto antes de añadir la siguiente.

Flow Cell Loading Diagram Step 07 V5 SPANISH

Gently replace the SpotON sample port cover, making sure the bung enters the SpotON port, close the priming port and close the lid of the sequencing device.

Flow Cell Loading Diagrams Step 8

Flow Cell Loading Diagrams Step 9

FIN DEL PROCESO

You are ready to begin your sequencing experiment.

Please refer to the MinKNOW protocol for instructions on setting up your sequencing experiment.

16. Overview of the VolTRAX RT-PCR Sequencing Kit 1-12 (VSK-VPS001)

The VSK-VPS001 RNA library preparation kit

The VolTRAX RT-PCR Sequencing Kit 1-12 generates PCR amplicons from the user's input RNA using the VolTRAX V2b platform to improve consistency in library preparation due to the precise fluid handling of the device.

This kit provides the reagents required to generate reverse-transcribed and PCR-amplified material from RNA samples using VolTRAX automation to prepare the library. The user provides their own PCR primers to generate custom amplicons, before barcoding and pooling multiple samples together for sequencing. We recommend using 10 mM Tris-HCl pH 8.0 as a buffer for the primers, and using the VolTRAX device to optimise the PCR conditions. When loading PCR primers onto the VolTRAX, we recommend loading at 3X the desired reaction concentration.

Below is an overview of the library preparation stages using a VolTRAX V2b, blue cartridges and the VolTRAX RT-PCR Sequencing Kit 1-12 (VSK-VPS001).

Overview of the VolTRAX RT-PCR Sequencing Kit 1-12 protocol

1. Loading a protocol

  • Use the protocol selector in the UI to load the VolTRAX RT-PCR Sequencing Kit 1-12 protocol.
  • When prompted, connect your VolTRAX V2b device, press Continue and await connection.

2. System QC - Cartridge Test

  • When prompted, connect a cartridge to the VolTRAX device and press Continue to proceed to the Cartridge Test.

3. Loading the Priming Fluid and reagents into the cartridge

  • Follow the instructions on the equipment required and pipetting technique.
  • Prime the cartridge by twisting the oil container and removing the foil tab.
  • Load the appropriate reagents, custom primers and extracted RNA into the ports as instructed by the UI.
  • You will be given the option to adjust the PCR timings and temperatures to allow you to use custom primers.
  • The reagents will be manipulated by the VolTRAX V2b to prepare the library for nanopore sequencing.

4. Extracting the library from the cartridge

  • Follow the extraction instructions, using UI as a visual guide, to slowly extract the library from the cartridge, then transfer it into an Eppendorf tube.
  • Prepare the library for loading onto the flow cell.

VolTRAX RT-PCR Sequencing Kit 1-12 (VSK-VPS001) contents

VSK-VPS001 tubes 1

Name Acronym Cap colour Number of vials Fill volume per vial (µl)
LunaScript RT SuperMix LS RT Blue 1 10
Q5 HS Master Mix Q5 Orange 1 35
VX BC Frag Mix VXB01-12 Amber 12 10
VX AMPure XP Beads VXP Pink 1 45
VX Rapid Adapter F VRA F Green 1 12
VX Elution Buffer VELB Black 1 50
VX Running Buffer VRB Red 1 250
Negative Control NEG Clear 1 10
Flush Buffer FB Blue 3 1170
Flush Tether FLT White cap, purple label 1 200

VolTRAX RT-PCR Sequencing Kit 1-12 barcode sequences

Component Sequence
VXB01 AAGAAAGTTGTCGGTGTCTTTGTG
VXB02 TCGATTCCGTTTGTAGTCGTCTGT
VXB03 GAGTCTTGTGTCCCAGTTACCAGG
VXB04 TTCGGATTCTATCGTGTTTCCCTA
VXB05 CTTGTCCAGGGTTTGTGTAACCTT
VXB06 TTCTCGCAAAGGCAGAAAGTAGTC
VXB07 GTGTTACCGTGGGAATGAATCCTT
VXB08 TTCAGGGAACAAACCAAGTTACGT
VXB09 AACTAGGCACAGCGAGTCTTGGTT
VXB10 AAGCGTTGAAACCTTTGTCCTCTC
VXB11 GTTTCATCTATCGGAGGGAATGGA
VXB12 CAGGTAGAAAGAAGCAGAATCGGA

VSK-VPS001 chemistry

The VolTRAX RT-PCR Sequencing Kit 1-12 generates amplicons across target regions or organisms of interest within the extracted RNA sample. First, the RNA samples are reverse-transcribed and then amplified by PCR using the user-defined primers. The samples are purified with AMPure XP beads before tagmenting the DNA amplicons with the Rapid Barcodes, adding the sequencing adapter and pooling for sequencing.

VSK-VPS001 workflow

17. Run a VolTRAX RT-PCR Sequencing Kit 1-12 (VSK-VPS001)

Material
  • Input RNA in 10 mM Tris-HCl, pH 8.0
  • VolTRAX RT-PCR Sequencing Kit 1-12 (VSK-VPS001)
  • PCR primers in 10 mM Tris-HCl, pH 8.0
  • VolTRAX Cartridge Pack (VCT-V2002B)

Consumibles
  • Tubos de PCR de pared fina (0,2 ml)
  • Qubit dsDNA HS Assay Kit (ThermoFisher, Q32851)

Instrumental
  • VolTRAX V2b
  • Pipeta y puntas P20
  • Pipeta y puntas P10
Equipo opcional
  • Fluorímetro Qubit (o equivalente para el control de calidad)
IMPORTANTE

VolTRAX V2b pipette tip compatibility

We have found that certain pipette tips give better results during the reagent loading and library extraction. Using the wrong tip type can lead to sample loss. The Mettler Toledo (Rainin™) UNV 10 µl tips have been verified as having the correct geometry for loading reagents on the VolTRAX cartridge. We recommend the Mettler Toledo (Rainin™) UNV 20 µl tips for extracting the prepared library at the end of the run.

Board test

The Board test should be carried out on your VolTRAX device before proceeding with PCR experiments. The most recent models of the VolTRAX V2b feature improved temperature calibration for PCR, and the Board test checks whether you have a version of the device that is PCR-compatible.

Start the VolTRAX software. You will be presented with the following options for protocols to run. Select "Board Test" to start the protocol and follow the instructions in the software.

0.21.0 Home Screen Board test

Board test results

If the board test has been successful, you can proceed with PCR experiments.

Board Test Success

If the board test has failed, please contact Support.

Board Test Failed

Click "Continue", which will take you back to the protocol selection screen. Select "VolTRAX RT-PCR Sequencing Kit" to start the protocol and follow the instructions in the software.

0.21.0 Home Screen RT-PCR

Protocol preview

Once an option is selected from the menu, the UI will show a preview of the steps involved in library preparation. The example below is for the Configuration Kit.

Protocol Preview Dialog

When prompted by the UI, connect your VolTRAX device to the computer. When it is connected, you will then be prompted to insert a cartridge into the device.

Prime the cartridge as instructed in the UI by twisting the Priming Fluid container anti-clockwise and removing the foil tab to allow the fluid to enter the cartridge.

Select Continue when each step is completed.

CONSEJO

Please refer to the videos displayed in the VolTRAX UI, which demonstrate how to perform each step in the protocol.

CONSEJO

Loading solutions into the VolTRAX V2b cartridge

The VolTRAX V2b device can measure solution volumes loaded into the cartridge. The GUI displays a coloured-pixel grid to show:

  • Blue: Volume of solution too low. You will need to load more solution to proceed with the protocol.
  • Green: Correct volume of solution.
  • Orange: slightly overfilled. You will still be able to proceed with the protocol.
  • Red: solution overfilled. You will need to cancel the experiment and start again.

Volume colours

When loading PCR primers onto the VolTRAX, we recommend loading at 3X the desired reaction concentration.

Adjust the PCR parameters

The PCR parameters form is always populated with the default parameters, however you can adjust the times and temperatures if you are using custom primers. This form enables 3-step PCR by default, however the parameters may be modified to allow 2-step PCR as described below.

Screenshot 2022-09-14 at 16.16.57

Cycle Count

This section allows you to adjust the number of PCR cycles for your application and incorporate touchdown PCR cycles. Touchdown PCR is an approach that can increase the specificity of a PCR by reducing the amount of off-target priming. The annealing step begins at a temperature several degrees higher than the Tm of the primers. Over the following cycles, the annealing temperature is gradually reduced until it is several degrees below the Tm of the primers.

  • Number of Touchdown Cycles: allows control over the number of cycles where touchdown is applied. These cycles are separate from the "Number of Cycles" field below.
  • Number of Cycles (Post-Touchdown): total number of PCR cycles to complete after the touchdown PCR cycles.

Total Number of Cycles = Number of Touchdown Cycles + Number of Cycles

Denaturation

  • Time (seconds): allows control over the duration of the denaturation step of the PCR.
  • First Cycle Extra Time (seconds): this parameter gives you additional control over the initial denaturation step. If required, extend the time for the first denaturation step. If an extended first denaturation step is not required, leave the value at 0.
  • Temperature (°C): allows you to set the temperature (°C) for the denaturation step of the PCR.
  • Minimum Temperature (°C): this is the temperature at which the countdown for the denaturation step of the cycle will start. It is recommended that this is set to be 0.5°C less than the denaturation temperature.

Annealing

  • Time (seconds): allows you to control the duration of the annealing step of the PCR.
  • Temperature (°C): allows you to set the temperature (°C) for the annealing step of the PCR. This temperature is the annealing temperature at the start of the touchdown process.
  • Maximum Temperature (°C): this is the temperature at which the countdown for the extension step of the cycle will start. It is recommended that this is set to be 0.5°C more than the annealing temperature.
  • Total Temperature Change Over Touchdown Cycles (°C): allows to modify the extension temperature over the touchdown PCR cycles (e.g. a value of -5 with 5 touchdown cycles would reduce the annealing temperature of the reaction by 1°C on each touchdown PCR cycle. The subsequent cycles would use the Temperature -5).

Extension:

  • Time (seconds): allows you to control the duration of the extension step of the PCR.
  • Final Cycle Extra Time (seconds): this parameter gives additional control over the final extension step of the PCR. If required, extend the time for the final extension step. If an extended final extension step is not required, leave the value at 0.
  • Temperature (°C): allows you to set the temperature (°C) for the extension steps of the PCR. This temperature is the annealing and extension temperature at the start of the touchdown process.
  • Minimum Temperature (°C): this is the temperature at which the countdown for the extension step of the cycle will start. It is recommended that this is set to be 0.5°C less than the extension temperature.

To set up 2-step PCR: Set all Extension step fields to 0. Please note that this will result in no extra time being applied to the final cycle's last step. If extra time is required in the final cycle's last step, make the following adjustments:

  1. Fill in the desired values for the Cycle Count, Denaturation, and Annealing fields (the annealing step in this case is both annealing and extension).
  2. Set the Extension Time field to 0 (the extension will complete immediately once the minimum temperature is reached).
  3. Set the Final Cycle Extra Time field to the desired additional time on the final Annealing step.
  4. Set the Temperature field to the temperature of the final annealing step, taking the touchdown gain into account.
  5. Set the Minimum Temperature field to 0°C (this will immediately trigger the step timer).

Wait for the library prep to finish.

A timer can be found on the right hand side of the UI, indicating the remaining time.

CONSEJO

Cancelling a run

To cancel a run, select Cancel in the top right corner of the GUI.

The user is able to look through all the steps of the selected protocol before confirming cancellation.

The run will stop immediately and return the user to the "Start VolTRAX" page. Another run can be started immediately.

Note: The cartridge must be replaced if reagents have been input during a cancelled run. If no reagent has been inserted, the cartridge can be used again.

At the end of the experiment, extract the prepared library from the E port of the VolTRAX cartridge. Dispense the library into a 0.2 ml PCR tube for preparation of the sequencing mix, as discussed in the following section.

MEDIDA OPCIONAL

Quantify 1 µl of eluted sample using a Qubit fluorometer.

CONSEJO

Run history

The UI provides a run history option to view previously-run protocols, which can be accessed from the link in the top right corner.

Run history

Clicking Open on any of the protocols will navigate the user to the relevant data folder in Windows File Explorer:

Run history2

15. Priming and loading the SpotON flow cell

Material
  • Flush Buffer (FB)
  • Flush Tether (FLT)

Consumibles
  • MinION Flow Cell
  • Tubos de 1,5 ml Eppendorf DNA LoBind

Instrumental
  • Pipeta y puntas P1000
  • Pipeta y puntas P100

Open the MinION Mk1B lid and slide the flow cell under the clip.

Press down firmly on the flow cell to ensure correct thermal and electrical contact.

Flow Cell Loading Diagrams Step 1a

Flow Cell Loading Diagrams Step 1b

MEDIDA OPCIONAL

Antes de cargar la biblioteca, verifique la celda de flujo para determinar el número de poros disponible.

Si se ha verificado con anterioridad la cantidad de poros presentes en la celda de flujo, este paso se puede omitir.

Dispone de más información en las instrucciones de comprobación de la celda de flujo, del protocolo de MinKNOW.

Slide the priming port cover clockwise to open the priming port.

Flow Cell Loading Diagrams Step 2

IMPORTANTE

Tenga cuidado a la hora de extraer el tampón de la celda de flujo. No retire más de 20-30 μl y asegúrese de que el tampón cubra la matriz de poros en todo momento. La introducción de burbujas de aire en la matriz puede dañar los poros de manera irreversible.

After opening the priming port, check for a small air bubble under the cover. Draw back a small volume to remove any bubbles (a few µls).

  1. Set a P1000 pipette to 200 µl
  2. Insert the tip into the priming port
  3. Turn the wheel until the dial shows 220-230 ul, to draw back 20-30 ul, or until you can see a small volume of buffer entering the pipette tip

Note: Visually check that there is continuous buffer from the priming port across the sensor array.

Flow Cell Loading Diagrams Step 03 V5

To prepare the flow cell priming mix, add 30 µl of thawed and mixed Flush Tether (FLT) directly to the tube of thawed and mixed Flush Buffer (FB), and mix by vortexing at room temperature.

Cargar 800 μl de solución en el puerto de cebado, evitando introducir burbujas de aire. Esperar 5 minutos. Durante este tiempo, preparar la biblioteca para cargar siguiendo los pasos a continuación.

Flow Cell Loading Diagrams Step 04 V5 SPANISH

Thoroughly mix the contents of the VX Running Buffer (VRB) tube by pipetting up and down.

In a new tube, prepare the library for loading as follows:

Reagent Volume per flow cell
VX Running Buffer (VRB) 64 µl
DNA library 11 µl
Total 75 µl

Completar el cebado de la celda de flujo:

  1. Levantar suavemente la tapa del puerto de carga SpotON.
  2. Cargar 200 µl de solución en el puerto de cebado (no en el puerto de muestra SpotON), evitando introducir burbujas de aire.

Flow Cell Loading Diagrams Step 5

Flow Cell Loading Diagrams Step 06 V5 SPANISH 2

Mezclar la biblioteca pipeteando suavemente, justo antes de cargar.

Añadir, gota a gota, 75 μl de la biblioteca preparada en el puerto de muestra SpotON. Procurar que cada gota fluya hacia adentro del puerto antes de añadir la siguiente.

Flow Cell Loading Diagram Step 07 V5 SPANISH

Gently replace the SpotON sample port cover, making sure the bung enters the SpotON port, close the priming port and close the lid of the sequencing device.

Flow Cell Loading Diagrams Step 8

Flow Cell Loading Diagrams Step 9

FIN DEL PROCESO

You are ready to begin your sequencing experiment.

Please refer to the MinKNOW protocol for instructions on setting up your sequencing experiment.

Last updated: 3/1/2024

Document options