ElysION user guide (ELY_9212_v1_revE_10Oct2025)


1. Device Information

ElysION™

The ElysION™ is a fully automated sample-to-answer sequencing solution with an integrated MinION™ and onboard computing. It is designed to bring streamlined genomics into the lab by offering a range of end-to-end sequencing workflows for a broad field of research applications.

The ElysION is a combined liquid handling and sequencing device that is intended for use in a laboratory setting for the preparation and sequencing of nanopore libraries from a range of input materials.


This device is for research use only.


Current supported Oxford Nanopore workflows:

  • Rapid Barcoding Kit 96 V14 (SQK-RBK114.96)
  • Native Barcoding Kit 96 V14 (SQK-NBD114.96)
  • Nanopore-only Microbial Isolate Sequencing Solution (NO-MISS)
  • Rapid sequencing V14 - Plasmid sequencing

Current supported Oxford Nanopore kits:

  • Rapid Barcoding Kit 96 V14 (SQK-RBK114.96)
  • Native Barcoding Kit 96 V14 (SQK-NBD114.96)
  • Flow Cell Wash Kit XL (EXP-WSH004-XL)

Current supported flow cells:

  • MinION & GridION Flow Cells R10.4.1 (FLO-MIN114)

2. What's in the box

Package includes:

  • 1x ElysION Sample Preparation Device
  • 1x ElysION Integration Kit - MinION
  • 1x ElysION Control Unit - MinION
  • 3x MinION Mk1D sequencing devices
  • 3x MinION & GridION Flow Cells R10.4.1 (FLO-MIN114)
  • 1x Rapid Barcoding Kit 96 V14 (SQK-RBK114.96)
  • 1x Software Licence and Device Warranty (12 months)

Also included:

  • Installation of instrument and software
  • Installation qualification
  • On-site training

3. Power on the ElysION device

Switch on the ElysION by pressing the power button. The central light above the ElysION device door should switch on.

Power on ElysION

The touch screen monitor attached to the ElysION will automatically power on.

  • If the touch screen monitor does not automatically switch on, press the monitor power button on the right-hand side of the underside of the monitor.

  • The Background Services (BotQL) will automatically start during power on.

4. Power on the compute control unit

Turn on the compute control unit by pressing the power button.

Power on Compute Control Unit

5. Power off the devices

Power off the ElysION.

You can turn off the ElysION using either of the following options:

  • Press the Start icon at the bottom of the Windows menu on the touch screen monitor, then select Shut down.

Power off ElysION

or

  • Press and hold the ElysION power button for five seconds.

Power off the compute control unit.

Turn off the compute control unit by selecting the power button in the top right corner and then selecting Power off.

6. Reference: Correct cable plan of the ElysION

Attach the cables and peripherals to the ports shown in the diagram. The USB-C port can be used to connect an adapter to plug in USB Type A cables.

ElysION connections & ports

Plug the ElysION power cable into a wall socket, then plug the power cable into the ElysION.

Switch on the ElysION by pressing down the power button.

The touch screen monitor attached to the ElysION will automatically power on.

If the ElysION monitor does not automatically switch on, press the monitor power button on the right-hand side of the underside of the monitor.

To connect a USB type A keyboard/mouse or hard drive, connect a USB-C to USB-A multi adapter into the USB-C port.

We recommend doing this for the accurate editing of sample sheets.

7. Reference: Correct cable plan of the compute control unit

Attach the cables and peripherals to the ports shown in the image.

Compute control unit

Connect a keyboard and mouse to the USB type A ports.

Connect the computer to the internet with an ethernet cable connection to ethernet port 1.

Connect the computer to the ElysION with an ethernet cable connection to ethernet port 2.

Connect a display port cable between the computer and a monitor screen.

Connect the power cable to the computer, then plug it into a wall power socket.

Turn on the computer by pressing the power button.

8. Start the user interface and initialise

Before initialising the instrument, perform all daily maintenance steps and ensure that the worktable has been cleared of any labware or reagents and there are no obstructions.

If you are unsure if the instrument is ready to initialise, please contact Customer Support.

Select the 'ElysION' application from the desktop of the ElysION touch screen monitor.

The window shown below will open.

Select ElysION application

Wait for the Liquid handler and MinION to connect to the compute control unit software.

Once connected, the Liquid handler section will show an Initialise option and the MinION section will show the details of the MinION Mk1D connected, as shown in the image below.

Liquid handler and MinION connecting to Compute Unit

Ensure the worktable of the ElysION is cleared. Close the device door and select the 'Initialise' option through the ElysION UI.

Initialising Liquid handler

Once initialisation has successfully completed, you will be able to run an automated method.

9. Check data storage capacity ahead of starting a run

Select 'System', and 'Device information'.

Ensure there is sufficient storage space for your planned assay run.

Check for data storage capacity

10. Inserting a flow cell onto the ElysION device

Remove the MinION Mk1D flow cell cover.

MinION Mk1D flow cell cover in place

Flow cell cover removed

Insert a MinION Flow Cell with all ports closed into the MinION Mk1D.

Placing flow cell under the clip

Inserting flow cell into position

Press the flow cell down firmly into position.

Flow cell firmly in position

Replace the MinION Mk1D flow cell cover.

Flow cell cover replaced

11. Choose a method to run

Through the ElysION UI select the 'Run Method' tab on the left side of the window.

To begin a method, select from the list of available methods. Then select 'Next' at the bottom right corner of the screen.

Refer to the method specific user manual for further instructions on preparing your workflow.

Selecting run method

12. Guide to setting the run variables

Types of run variables

During the preparation of the device and ahead of starting a workflow, there are three optional variables that can be set.

(A) Sequencing time limit

This variable gives you the option to set the maximum length of time (in hours) the instrument will sequence for or the option to run until the flow cell pores are depleted (5% of total pore count). When either the time limit has been reached or the pores have been depleted, the instrument will stop sequencing and proceed to the next step in the workflow.

Sequencing time limit

(B) Data target

This variable allows you to set a maximum sequencing data output (in Gbases). Once this data target has been achieved, the instrument will stop sequencing and proceed to the next step in the workflow. Alternatively, where no data target is set, sequencing will stop according to the set 'sequencing time limit' variable.

Note: Sequencing will stop when either the set sequencing time limit or set data target has been met. For example, if the sequencing time limit is set to 72 hours and the data target is set to 20 Gbases, and sequencing produces 20 Gbases in 48 hours, sequencing will stop before the 72 hours time limit.

Data target and flow cell flush/wash

(C) Flow cell flush/wash

This variable allows you to decide what happens to the flow cell once sequencing has completed. Flushing prepares the flow cell for return to Oxford Nanopore Technologies for recycling. Washing cleans the flow cell to allow multiple use of the same flow cell.

Note: Flush/Wash choices will require different reagents to be used. Washing will require an additional Oxford Nanopore Technologies Flow Cell Wash Kit (EXP-WSH004).

13. Guide to writing a sample sheet

Writing a sample sheet to perform a workflow

All ElysION workflows require a sample sheet in order to perform a workflow. The sample sheet performs several functions:

  • Instructs the instrument on how many samples to process.
  • Instructs the instrument on which barcodes to attach to samples and to include in sequencing and analysis.

Sample sheets contain unique parameter values required for the specific workflow being run. You can alter the parameters to ensure correct preparation of samples. If sample sheet parameters do not match the workflow assay or are in the incorrect format, the sample sheet will be rejected.

To avoid issues when writing a new sample sheet, we recommend you always design a new sample sheet from the template provided by Oxford Nanopore Technologies.


Values which must not be altered on the sample sheet (highlighted in pink): Values which must not be altered


Values which can be altered on the sample sheet (highlighted in green): Values which can be altered


Rules of editing values on the sample sheet:

  • library_id: Any text to identify one library from another in a human readable format.
  • created_by: Any text to identify the user who wrote the sample sheet in a human readable format.
  • created_at: The time stamp that the sample sheet file was created. This must be in the ISO 8601 format.

ISO 8601 formatting:

  • 2024-06-21: The date (Year-Month-Day).
  • T: The separator between the date and time (standard in ISO 8601).
  • 14:07:14: The time in 24-hour format (Hours:Minutes:Seconds).
  • +01:00: The time zone offset from UTC (in this case, it is 1 hour ahead of UTC).

sample_count: The exact number of samples being processed by the instrument given as an integer value. This value must be within the range possible for the workflow being prepared (see workflow specific documentation).


well_id: The id of the wells that the samples start in. This must always start at A1 and continue in a column wise list, with no gaps. The total number of well ids listed in the sample sheet must be equal to the sample count value also provided. Examples of correct and incorrect well_id configurations are shown in the Table below:

Correct well_id configuration Incorrect well_id configuration
A1, B1, C1, D1, E1, F1, G1, H1, A2 and sample count = 9 B1, C1, D1, E1, F1, G1, H1, A2, B2 (A1 missing)
A1, B1, C1, D1, F1, G1, H1, A2, B2 (E1 missing)
A1, A2, A3, A4, A5, A6, A7, A8 (Row format)
A1, B1, C1, D1, E1, F1, G1, H1 and sample count = 9 (only 8 samples listed)

barcode: This list of barcode ids specifies which barcode from Oxford Nanopore Technologies barcode plate is delivered to each sample (as specified in the well_id column). The first barcode can be taken from any position in the barcode plate however, all subsequent barcodes must follow in a sequential order. The total number of barcode ids must match the total number of samples provided. Each barcode can only be used once and must be formatted as a two-digit number. Examples of correct and incorrect barcode configurations are shown in the Table below:

Correct barcode configuration Incorrect barcode configuration
barcode01, barcode02, barcode03, barcode04, barcode05, barcode06 and sample count = 6 barcode03, barcode04, barcode05, barcode06, barcode01, barcode02 (not sequential)
barcode06, barcode07, barcode08, barcode09, barcode10, barcode11, and sample count = 6 barcode01, barcode02, barcode03, barcode04, barcode05, barcode08 (not sequential)
barcode08, barcode09, barcode010, barcode11, barcode12, barcode13 (barcode010 has 3 digits)
barcode01, barcode02, barcode03, barcode04, barcode05, barcode06 and sample count = 8 (only 6 samples listed)

sample_type: The sample type flag which is used during post-basecalling analysis in EPI2ME. Defaults as 'sample' if post-basecalling analysis is not required. If you plan to do analysis, see the workflow specific documentation for further instructions.


sample_id: The sample id which is used during post-basecalling analysis in EPI2ME. Defaults as a unique integer if post-basecalling analysis is not required. If you plan to do analysis, see the workflow specific documentation for further instructions.

14. Guide to running automated sample normalisation

To include automated sample normalisation, an additional column named sample_volume must be added to the sample sheet as follows.

sample_volume: The sample volume flag indicates that automated sample normalisation is to be implemented and should be added to the right of the sample_id flag. Rows underneath must contain the volume of sample to be transferred to each well of the PCR plate (in µl). This volume will be automatically normalised to the required input volume using nuclease-free water (NFW). If the sample_volume column is absent, no automated sample normalisation will be implemented.

In the example below, 5 µl of sample will be transferred from the sample plate to the PCR plate and normalised to the required sample volume taken forward with NFW (e.g. 15 µl for RBK).

sample volume for normalisation

15. Guide to configuring a custom method

Under settings, go to ‘Method configurator’ on bottom left of screen.

Method configurator

Select ‘Create variable’ for the relevant protocol that you wish to alter.

Go through each variable and enter your desired parameters.

Enter required parameter for each variable

Once the changes have been saved, the custom method will be assigned a version number (e.g. NBD_1-48 v.1) which can be selected from the main method page.

Version number for custom method

16. Accessing the sequencing data

Log in on the compute control unit monitor (separate from the ElysION touch screen monitor).

  • Using the username: nanopore

  • Using the password: nanopore

On the left-hand side of the screen, open the 'File manager' application.

Navigate to the 'data' folder.

Open 'output', then 'sequences', then 'no_group', then 'no_sample'.

  • This folder will display all the available sequencing run data within subfolders with the naming convention: 'YYYYMMDD_HHMM_sequencing device ID_Flow Cell ID_XXXXXXXX'.

We recommend that you copy all sequencing data to a long-term data storage system regularly to prevent the compute unit storage from becoming full.

17. Accessing the EPI2ME™ output data

Log in on the compute control unit monitor (separate from the ElysION touch screen monitor).

  • Using the username: nanopore

  • Using the password: nanopore

On the left-hand side of the screen, open the 'File manager' application.

Navigate to the 'data' folder.

Open 'reception', then 'instances'.

  • This folder will display all the available EPI2ME pipeline data within subfolders, e.g. 'wf-bacterial-genomes_01J6HWYJM7QSJDWV1SHVZEHSMZ'.

  • To identify the pipeline output which corresponds to an automated run, refer to the date edited on the right-hand side of the file manager screen.

Note: the date that a folder is edited/written as shown in the file manager will not be the same date as the start of the automated library prep. EPI2ME analysis will start once sequencing has finished. Therefore, the date-time written will depend on the date the automated prep was started and the length of time the sequencing was performed.

We recommend that you copy all EPI2ME data to a long-term data storage system regularly to prevent the compute unit storage from becoming full.

18. Labware sourcing and usage

The Table below describes the list of labware compatible with the ElysION device.

Labware Recommended supplier Product code
Hard-Shell® 96-Well PCR Plates, low profile, thin wall, skirted Bio-Rad™ HSP9611
Nunc™ 96 Well Polypropylene 2 ml DeepWell Plate Thermo Scientific 95040452
ElysION Disposable Tips 1000 µl Oxford Nanopore Technologies ELY-TIP1000
ElysION Disposable Tips 200 µl Oxford Nanopore Technologies ELY-TIP0200
ElysION Disposable Tips 50 µl Oxford Nanopore Technologies ELY-TIP0050
ElysION Disposable Tips Box Small Oxford Nanopore Technologies ELY-TBS01
ElysION Disposable Tips Box Large Oxford Nanopore Technologies ELY-TBL01
Reservoir Plate Porvair 390015
Azenta PCR Plate Lid Azenta 4ti-0291
Screw cap micro tube 2 ml, sterile Sarstedt 72.694.006
Screw cap micro tube 0.5 ml, sterile Sarstedt 72.730.106
ElysION Waste Bin Oxford Nanopore Technologies ELY-WB01

Reuse of labware:

  • The small and large tip boxes can be re-used 1000 times. Through normal use, tip boxes do not become contaminated, so do not require washing.

  • It is possible to re-use the Azenta PCR plate lids. However, the lids should not be used more than five times and should not be washed (i.e. washed with 80% ethanol or IPA).

19. System configuration

Changing the system password

Note: The instructions below will apply for your username.

  1. Open command line on the Linux compute.

  2. Type: sudo passwd username

  • This will prompt you to enter your current password on the first attempt.
  1. Enter a new UNIX password: Type new password and press enter.

  2. Re-enter the new UNIX password: Type new password again and press enter.

  3. passwd: Password successfully changed


Changing the system username

  • We do not recommend the changing of usernames.

20. Cleaning the device

Maintenance and decontamination procedures

For full details on the maintenance and decontamination procedures, refer to the Set-up and operating manual: Early access ElysION provided with your ElysION device.

Using chemical agents

The device should be cleaned with any of the listed agents and dried with a lint-free cloth. Use of any other chemical agents than those listed below or listed in the full User manual may cause damage to the instrument or risk injury to you.

Cleaning agent Specification
DI water Distilled or deionized water
Alcohol 70% ethanol or 100% isopropanol (2‑propanol)
Weak detergent Liqui-Nox
Disinfectant Bacillol plus, Bomix, Lysetol FF
System liquid Distilled or deionized water
Base 0.025 to 0.25 mol/l sodium hydroxide
Bleach 6% sodium hypochlorite
RNase AwayTM decontamination reagent Alkali hydroxide <0.1%

21. UV light disinfection

Select 'System' in the bottom left corner of the ElysION software, on the touch screen monitor, then 'Maintenance'.

UV light disinfection

Select 'UV light' from the list of maintenance methods.

Selecting UV light from maintenance methods

Define the length of time for the UV disinfection and press 'Start'.

Selecting UV light duration

A cautionary message that the UV disinfection is running will appear with a time remaining report.

UV time remaining report

22. Maintenance schedule

ElysION maintenance schedule

Please refer to the User manual included with your ElysION device for details of the maintenance schedule.

Daily maintenance

  • Empty both waste containers.
  • Either decontaminate or replace with new waste containers.

Weekly maintenance

Clean the following with 70% ethanol:

  • Waste drawer
  • DiTi drop station
  • ElysION robot touch screen monitor

Monthly, quarterly, bi-annual, quinquennial maintenance

Please refer to the User manual for more information.

23. How to collect logging data for troubleshooting support

Windows

  1. On the touch screen monitor, click the Export Logs application from the desktop.

Support application


  1. Copy the generated logging snapshot to a USB drive.

Linux

  1. Open Show Applications in the bottom left corner.

Show Applications


  1. Select the Elysion Logs application.

ElysION Logs

  1. Enter the compute password.

  2. Copy the generated logging snapshot to a USB drive.

Serial numbers

  • The serial number for the ElysION liquid handler is found on the rear panel of the device.

  • The serial number for the compute control unit can be found on the rear side of the unit, as shown in the image below:

Compute Unit serial number

Reporting issues

Please report issues to Nanopore Support with the serial numbers for both the liquid handler and compute unit and the logging snapshots from both the Windows and Linux computes attached.

24. The ElysION device does not switch on

Check that the ElysION touch screen monitor is switched on by pressing the monitor power button on the bottom right-hand side of the underside of the monitor.

Check that all power cables are attached correctly and are connected to a mains supply.

Press and hold the ElysION power button for one second.

25. The compute control unit does not switch on

  • Contact your customer support representative and include the device serial numbers in the description of your issue.

26. Data offload from the device to long-term storage takes too long

  • Contact your customer support representative.

  • Oxford Nanopore Technologies can provide support to connect your instrument to an internal network to allow automated data offload from the ElysION device to your long-term storage.

27. The automated method has not completed the library preparation

  • Contact your customer support representative and include the device serial numbers in the description of your issue and attach copies of your logging data.

28. The automated method has not completed the anticipated sequencing

  • Contact your customer support representative and include the device serial numbers in the description of your issue and attach copies of your logging data.

Last updated: 10/10/2025

Opciones de documento

Idioma:

Getting started

Buy a MinION starter pack Nanopore store Sequencing service providers Channel partners

Quick links

Intellectual property Cookie policy Corporate reporting Privacy policy Terms & conditions Accessibility

About Oxford Nanopore

Contact us News Media resources & contacts Investor centre Careers BSI 27001 accreditationBSI 90001 accreditationBSI mark of trust
Spanish flag