UNPUBLISHED Ligation sequencing gDNA - automated Hamilton NGS STAR 96 (SQK-LSK110-XL)

概要

  • This protocol uses genomic DNA
  • Automation of library preparation
  • Increased reproducibility and speed
  • Reduces human error
  • High sequencing output
  • Library preparation time ~1.5 hours hands-on-time and ~3-4 hours automation time
  • Fragmentation is optional
  • No PCR required
  • Suitable for processing multiple samples simultaneously
  • Compatible with R10.3 flow cells

For Research Use Only

Document version: auto_gDNA_rev2

1. Overview of the protocol

注意

This protocol is currently UNPUBLISHED

UNPUBLISHED protocols are for internal development use ONLY, and should NEVER be shared with external users and customers.

In case of doubt, please contact Product Management.

Ligation Sequencing Kit XL features

This kit is recommended for users who:

  • want to optimise their sequencing experiment for throughput
  • would like to utilise upstream processes such as size selection, whole genome amplification, or enrich for long reads
重要

Optional fragmentation and size selection

By default, the protocol contains no DNA fragmentation step, however in some cases it may be advantageous to fragment your sample. For example, when working with lower amounts of input gDNA (100 ng – 500 ng), fragmentation will increase the number of DNA molecules and therefore increase throughput. Instructions are available in the DNA Fragmentation section of Extraction methods.

Additionally, we offer several options for size-selecting your DNA sample to enrich for long fragments - instructions are available in the Size Selection section of Extraction methods.

Introduction to the automated Ligation Sequencing protocol for gDNA

This protocol describes how to carry out sequencing of a DNA sample using the Ligation Sequencing Kit XL (SQK-LSK110-XL). It is highly recommended that a Lambda control experiment is completed first to become familiar with the technology.

Steps in the sequencing workflow:

Prepare for your experiment You will need to:

  • Extract your DNA and check its length, quantity and purity. The quality checks performed during the protocol are essential in ensuring experimental success.
  • Ensure you have your sequencing kit, the correct equipment, primed liquid-handling robot and third-party reagents
  • Download the software for acquiring and analysing your data
  • Check your flow cells to ensure they have enough pores for a good sequencing run

__Library preparation__ You will need to:
  • Repair the DNA and prepare the DNA ends for adapter attachment
  • Attach sequencing adapters supplied in the kit to the DNA ends
  • Prime the flow cell and load your DNA library into the flow cell

2020 10 09 LSK109 hamilton workflow v1 DS

Sequencing and analysis You will need to:

  • Start a sequencing run using the MinKNOW software which will collect raw data from the device and convert it into basecalled reads
  • Start the EPI2ME software and select a workflow to further analysis (this step is optional)
重要

Compatibility of this protocol

This protocol should only be used in combination with:

  • Ligation Sequencing Kit XL (SQK-LSK110-XL)
  • Control Expansion (EXP-CTL001)
  • Flow Cell Wash Kit (EXP-WSH004)
  • Sequencing Auxiliary Vials (EXP-AUX002)
  • FLO-PRO002 (R9.4.1) flow cells
  • FLO-PRO111 (R10.3) flow cells

2. Equipment and consumables

材料
  • 1 µg (or 100-200 fmol) gDNA
  • 1.5-3 µg (or 150-300 fmol) high molecular weight genomic DNA for R10.3 flow cells
  • Ligation Sequencing Kit XL (SQK-LSK110-XL)

消耗品
  • Agencourt AMPure XP beads (Beckman Coulter, A63881)
  • NEBNext® Companion Module for Oxford Nanopore Technologies® Ligation Sequencing (NEB, E7180S or E7180L). Alternatively, you can use the NEBNext® products below:
  • NEBNext FFPE Repair Mix (NEB, M6630)
  • NEBNext Ultra II End repair/dA-tailing Module (NEB, E7546)
  • NEBNext Quick Ligation Module (NEB, E6056)
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • Freshly prepared 80% ethanol in nuclease-free water
  • Hamilton 50 µl CO-RE tips with filter (Cat# 235948)
  • Hamilton 300 µl CO-RE tips with filter (Cat# 235903)
  • Hamilton 1000 µl CO-RE tips with filter (Cat# 235905)
  • Hamilton 60 ml Reagent Reservoir, Self-Standing with Lid (Cat# 56694-01)
  • Hamilton PCR ComfortLid (Cat# 814300)
  • Bio-Rad Hard-Shell® 96-Well PCR Plates (Cat# HSP9601)
  • Roche Diagnotics MagNA Pure LC Medium Reagent Tubs 20 (Cat# 03004058001)
  • Sarstedt Inc Screw Cap Micro tube 2 ml, PP 1000/case (e.g. FisherScientific, Cat# NC0418367)
  • Thermo Scientific™ Abgene™ 96 Well 0.8 ml Polypropylene Deepwell Storage Plate (Thermo Scientific™, cat # AB0859)
  • Qubit dsDNA HS Assay Kit (Invitrogen, Q32851)
  • Qubit™ Assay Tubes (Invitrogen, Q32856)

装置
  • Ice bucket with ice
  • Vortex mixer
  • Microplate centrifuge, e.g. Fisherbrand™ Mini Plate Spinner Centrifuge (Fisher Scientific, 11766427)
  • Hamilton NGS STAR 96 (NGS STAR with Multi-Probe Head 96)
  • Hamilton On-Deck Thermal Cycler (ODTC)
オプション装置
  • Agilent Bioanalyzer (or equivalent)
  • Qubit fluorometer plate reader (or equivalent for QC check)

For this protocol, you will need 1 µg (or 100-200 fmol) high molecular weight genomic DNA for every sample to be barcoded if using R9.4.1 flow cells or 1.5-3 µg (or 150-300 fmol) for R10.3 flow cells.

Although 1 µg (or 100-200 fmol) gDNA is recommended, users can start with lower input quantities (down to 100 ng) if performing DNA fragmentation to increase the number of DNA molecules in the sample, or if amplifying the sample by PCR.

Input DNA

How to QC your input DNA

It is important to use a plate reader to ensure the input DNA meets the quantity and quality requirements. Using too little or too much DNA, or DNA of poor quality (e.g. highly fragmented or containing RNA or chemical contaminants) can affect your library preparation.

For instructions on how to perform quality control of your DNA sample, please read the Input DNA/RNA QC protocol.

Input file worklist

A worklist input Excel file is required prior to running the protocol on the Hamilton NGS STAR 96. These contain information regarding the appropriate number of samples and well identifiers.

Example:

Source_SampleID Source_Well Target_Well
Sample_01 A1 A1
Sample_02 B1 B2
Sample_03 C1 C1

Hamilton NGS STAR 96

This method has been tested and validated using the Hamilton NGS STAR 96 (with 8 channels and MPH96) including an on deck thermal cycler (ODTC). An option to not use the ODTC is available in the method. The protocol may require some fine tuning for the NGS STAR 96 setup and the temperature/humidity of the customer laboratory.

Please contact your Hamilton representative for further details.

Deck layout Ligation XL deck layout

  • ODTC: On-Deck Thermal Cycler Module
  • MIDI plates: Abgene™ 96 Well 0.8mL Polypropylene Deepwell Storage Plate
  • Troughs: Hamilton 20 ml Reagent Reservoirs
  • HSP plates: Bio-Rad Hard-Shell® 96-Well PCR Plates
  • Hamilton ComfortLid: Hamilton PCR ComfortLid

NEBNext® Companion Module for Oxford Nanopore Technologies® Ligation Sequencing

For customers new to nanopore sequencing, we recommend buying the NEBNext® Companion Module for Oxford Nanopore Technologies® Ligation Sequencing (catalogue number E7180S or E7180L), which contains all the NEB reagents needed for use with the Ligation Sequencing Kit.

Please note, for our amplicon protocols, NEBNext FFPE DNA Repair Mix and NEBNext FFPE DNA Repair Buffer are not required.

Consumables quantities:

Consumables X24 samples X48 samples X96 samples
Hamilton 50 µl CO-RE tips with filter 216 432 768
Hamilton 300 µl CO-RE tips with filter 451 694 1176
Hamilton 1000 µl CO-RE tips with filter 64 64 64
Hamilton 60 ml Reagent Reservoir, Self-Standing with Lid 3 3 3
Hamilton PCR ComfortLid 1 1 1
Bio-Rad Hard-Shell® 96-Well PCR Plate 2 2 2
Roche Diagnostics MagNA Pure LC Medium Reagent Tubs 20 2 2 2
Sarstedt Inc Screw Cap Micro Tube 2ml 2 4 5
Abgene™ 96 Well 0.8mL Polypropylene Deepwell Storage Plate 2 2 2
Kits X24 samples X48 samples X96 samples
Ligation Sequencing Kit XL (SQK-LSK110-XL) 1 kit 1 kit 2 kits
NEBNext Companion Module for Oxford Nanopore Technologies Ligation Sequencing (Cat# E7180S) 2 kits 3 kits 5 kits
Alternatively: - - -
NEBNext FFPE DNA Repair Mix (Cat# M6630L) 1 kit 1 kit 2 kits
NEBNext Ultra II End Repair/dA-Tailing Module (Cat# E7546L) 1 kit 2 kits 3 kits
NEBNext Quick Ligation Module (Cat# E6056L) 1 kit 2 kits 3 kits

Note: These are the number of kits required for one run through for the selected number of samples.

Ligation Sequencing Kit XL (SQK-LSK110-XL) contents

SQK-LSK110-XL

Name Acronym Cap colour Number of vials Fill volume per vial (µl)
Adapter Mix F AMX F Green 1 320
Sequencing Buffer SBII 10 ml bottle 1 5,000
Ligation Buffer LNB White 1 1,500
S Fragment Buffer SFB 10 ml bottle 4 7,500
L Fragment Buffer LFB 10 ml bottle 4 7,500
DNA CS DCS Yellow 1 100
Elution Buffer EB 10 ml bottle 1 10,000
Loading beads II LBII Pink 2 1,500
Loading solution LS White cap, pink sticker 2 1,500
Flush buffer XL FB 20 ml bottle 4 15,500
Flush tether FLT White cap, purple sticker 1 1,600

3. Computer requirements and software

PromethION 24/48 IT requirements

The PromethION device contains all the hardware required to control up to 24 (for the P24 model) or 48 (for the P48 model) sequencing experiments and acquire the data. The device is further enhanced with high performance GPU technology for real-time basecalling. Read more in the PromethION IT Requirements document.

PromethION 2 Solo IT requirements

The PromethION 2 (P2) Solo is a device which directly connects into a GridION Mk1 or a stand-alone computer that meets the miminum specifications for real-time data streaming and analysis. Up to two PromethION flow cells can be can be run and each is independently addressable, meaning experiments can be run concurrently or individually. For information on the computer IT requirements, please see the PromethION 2 Solo IT requirements document.

Software for nanopore sequencing

MinKNOW

The MinKNOW software controls the nanopore sequencing device, collects sequencing data and basecalls in real time. You will be using MinKNOW for every sequencing experiment to sequence, basecall and demultiplex if your samples were barcoded.

For instructions on how to run the MinKNOW software, please refer to the MinKNOW protocol.

EPI2ME (optional)

The EPI2ME cloud-based platform performs further analysis of basecalled data, for example alignment to the Lambda genome, barcoding, or taxonomic classification. You will use the EPI2ME platform only if you would like further analysis of your data post-basecalling.

For instructions on how to create an EPI2ME account and install the EPI2ME Desktop Agent, please refer to the EPI2ME Platform protocol.

Check your flow cell

We highly recommend that you check the number of pores in your flow cell prior to starting a sequencing experiment. This should be done within three months of purchasing for MinION/GridION/PromethION or within four weeks of purchasing Flongle Flow Cells. Oxford Nanopore Technologies will replace any flow cell with fewer than the number of pores in the table below, when the result is reported within two days of performing the flow cell check, and when the storage recommendations have been followed. To do the flow cell check, please follow the instructions in the Flow Cell Check document.

Flow cell Minimum number of active pores covered by warranty
Flongle Flow Cell 50
MinION/GridION Flow Cell 800
PromethION Flow Cell 5000

4. DNA repair and end-prep

材料
  • 1.5-3 µg (or 150-300 fmol) high molecular weight genomic DNA for R10.3 flow cells

消耗品
  • Nuclease-free water (e.g. ThermoFisher, AM9937)
  • NEBNext FFPE DNA Repair Mix (NEB, M6630)
  • NEBNext® Ultra II End Repair / dA-tailing Module (NEB, E7546)
  • Agencourt AMPure XP beads (Beckman Coulter™, A63881)
  • Freshly prepared 80% ethanol in nuclease-free water
  • Thermo Scientific™ Abgene™ 96 Well 0.8 ml Polypropylene Deepwell Storage Plate (Thermo Scientific™, cat # AB0859)
  • Sarstedt Inc Screw Cap Micro tube 2 ml, PP 1000/case (e.g. FisherScientific, Cat# NC0418367)
  • Roche Diagnotics MagNA Pure LC Medium Reagent Tubs 20 (Cat# 03004058001)
  • Bio-Rad Hard-Shell® 96-Well PCR Plates (Cat# HSP9601)
  • Hamilton PCR ComfortLid (Cat# 814300)
  • Hamilton 60 ml Reagent Reservoir, Self-Standing with Lid (Cat# 56694-01)
  • Hamilton 1000 µl CO-RE tips with filter (Cat# 235905)
  • Hamilton 300 µl CO-RE tips with filter (Cat# 235903)
  • Hamilton 50 µl CO-RE tips with filter (Cat# 235948)
  • Qubit dsDNA HS Assay Kit (Invitrogen, Q32851)
  • Qubit™ Assay Tubes (Invitrogen, Q32856)

装置
  • Ice bucket with ice
  • P1000 pipette and tips
  • P200 pipette and tips
  • P100 pipette and tips
  • P10 pipette and tips
  • Microplate centrifuge, e.g. Fisherbrand™ Mini Plate Spinner Centrifuge (Fisher Scientific, 11766427)
  • Vortex mixer
  • Qubit fluorometer (or equivalent)

Consumables and equipment quantities:

Consumable/equipment X24 samples X48 samples X96 samples
Hamilton 50 µl CO-RE tips with filter 96 192 384
Hamilton 300 µl CO-RE tips with filter 201 298 490
Hamilton 1000 µl CO-RE tips with filter 32 32 32
Hamilton 60 ml Reagent Reservoir, Self-Standing with Lid 2 (1 EtOH & H20) 2 (1 EtOH & H20) 2 (1 EtOH & H20)
Hamilton PCR ComfortLid 1 1 1
Bio-Rad Hard-Shell® 96-Well PCR Plate 1 (1 input sample & 1 end prepped sample) 1 (1 input sample & 1 end prepped sample) 1 (1 input sample & 1 end prepped sample)
Hamilton 20 ml Reagent Reservoirs 1 1 1
Sarstedt Inc Screw Cap Micro Tube 2 ml 1 2 2
Abgene™ 96 Well 0.8 ml Polypropylene Deepwell Storage Plate 1 1 1

Reagents quantities:

Reagents X24 samples X48 samples X96 samples
80% ethanol 16.5 ml 28 ml 51 ml
AMPure XP Beads 3.7 ml 5.4 ml 8.9 ml
NEBNext Companion Module for Oxford Nanopore Technologies Ligation Sequencing (Cat# E7180S)
regarding the reagents below
2 tubes 3 tubes 5 tubes
Alternatively: - - -
NEBNext FFPE DNA Repair Buffer 1 tube 1 tube 1 tube
NEBNext FFPE DNA Repair Mix 1 tube 1 tube 2 tubes
Ultra II End-prep Reaction Buffer 1 tube 2 tubes 3 tubes
Ultra II End-prep Enzyme Mix 1 tube 1 tube 2 tubes

Note: Dead volumes are included.

重要

Optional fragmentation and size selection

By default, the protocol contains no DNA fragmentation step, however in some cases it may be advantageous to fragment your sample. For example, when working with lower amounts of input gDNA (100 ng – 500 ng), fragmentation will increase the number of DNA molecules and therefore increase throughput. Instructions are available in the DNA Fragmentation section of Extraction methods.

Additionally, we offer several options for size-selecting your DNA sample to enrich for long fragments - instructions are available in the Size Selection section of Extraction methods.

Prepare the NEBNext FFPE DNA Repair Mix and NEBNext Ultra II End Repair / dA-tailing Module reagents in accordance with manufacturer’s instructions, and place on ice.

For optimal performance, NEB recommend the following:

  1. Thaw all reagents on ice.

  2. Flick and/or invert the reagent tubes to ensure they are well mixed.
    Note: Do not vortex the FFPE DNA Repair Mix or Ultra II End Prep Enzyme Mix.

  3. Always spin down tubes before opening for the first time each day.

  4. The Ultra II End Prep Buffer and FFPE DNA Repair Buffer may have a little precipitate. Allow the mixture to come to room temperature and pipette the buffer up and down several times to break up the precipitate, followed by vortexing the tube for 30 seconds to solubilise any precipitate.
    Note: It is important the buffers are mixed well by vortexing.

  5. The FFPE DNA Repair Buffer may have a yellow tinge and is fine to use if yellow.

Prepare each DNA sample per well with nuclease-free water in the input plate.

  • Per sample, transfer 1 μg (or 100-200 fmol) of genomic DNA into a well of the input plate
  • Adjust the volume to 48 μl with nuclease-free water
  • Mix thoroughly by pipetting
  • Spin down briefly in a microfuge

Quantify 1 µl of each eluted sample using a Qubit fluorometer plate reader off deck.

Switch on the Hamilton NGS STAR 96 robot and open 'Hamilton Run Control' on the computer by clicking the icon:

Hamilton icon

Click 'File' and 'Open' to choose the method to run on the liquid handling robot.

Click 'Process01: DNA repair and end-prep' to start.

Capture1

Click 'Process02: DNA repair and end-prep clean-up' to stop the automated library preparation and quantify the samples before the adapter ligation step.

Capture2

重要

It is mandatory for users to have an MPH module installed and we recommend the use of an ODTC module.

Select whether an ODTC module is available to use in the run and select Yes to use the MPH (96 Head) module.

Capture3

Click 'Browse' to choose the Input File Worklist for the specific number of samples in the run and click 'OK'.

Capture4

An input file worklist for the number of samples in the run must be generated before the run. Example of an input file worklist:

Source_SampleID Source_Well Target_Well
Sample_01 A1 A1
Sample_02 B1 B1
Sample_03 C1 C1

Prepare the End Prep Mastermix with the following reagents according to the Hamilton user interface. Click either 'Yes' or 'No' to continue.

Note: It is user preference whether to print and save the instructions.

Reagent volumes for all sample numbers:

Reagent Volume X24 samples Volume X48 samples Volume X96 samples
NEBNext FFPE DNA Repair Buffer 106.6 µl 213.3 µl 414.8 µl
NEBNext FFPE DNA Repair Mix 60.9 µl 121.8 µl 237 µl
Ultra II End-prep Reaction Buffer 106.6 µl 213.3 µl 414.8 µl
Ultra II End-prep Enzyme Mix 91.4 µl 182.8 µl 355.6 µl
Total 365.6 µl 731.2 µl 1422.2 µl

Capture5

Insert the ComfortLid position as displayed on screen. Click 'Ok' to continue.

Capture6

Insert plates to their corresponding positions. Click 'Ok' to continue.

Capture7

Load a full deck of 50 µl tips into the positions on screen. Click 'Ok' to continue.

Capture8

Highlight the 50 µl tips available to use on the 'Edit Tip Count' window and click 'Ok' to continue.

Capture9

Load a full deck of 300 µl tips in the positions on screen. Click 'Ok' to continue.

Capture10

Highlight the 300 µl tips available to use on the 'Edit Tip Count' window and click 'Ok' to continue.

Capture11

Freshly prepare 80% ethanol in nuclease-free water in a trough.

Reagents Volume X24 samples Volume X48 samples Volume X96 samples
80% ethanol 16.5 ml 28 ml 51 ml

Insert the trough of 80% ethanol in the position on screen and click 'Ok' to continue.

Capture12

ヒント

If the consumables used for troughs are not barcoded, click 'Exclude' on all the selected troughs inserted in the robot and click 'Execute' to continue.

Capture13

Prepare the AMPure XP beads by vortexing and load the 20 ml trough with the volume required:

Reagents Volume X24 samples Volume X48 samples Volume X96 samples
Beads 3.7 ml 5.4 ml 8.9 ml
重要

Ensure the AMPure XP beads are well mixed before use by vortexing.

Insert the trough of AMPure XP beads and nuclease-free water in their positions on screen. Click 'Ok' to continue.

Capture14

ヒント

If the consumables used for troughs are not barcoded, click 'Exclude' on all the selected troughs inserted in the robot and click 'Execute' to continue.

Capture13

Load 1000 µl tips and insert the input plate of DNA samples into the position on screen. Click 'Ok' to continue.

Capture16

Highlight the 1000 µl tips available to use on the 'Edit Tip Count' window and click 'Ok' to continue.

Capture17

注意

Ensure the mastermix is well mixed and homogenous before loading the 2 ml Sarstedt tubes. Mixing in the robot is not effective.

Mix and insert the prepared End Prep Mastermix into the positions on screen.

Capture18

Click 'Ok' to start the DNA repair and end-prep automation process.

Once the automation process has finished, there will be an on screen prompt to unload the plate. Click 'Ok' to continue.

DNA repair and end prep

Quantify 1 µl of each eluted sample using a Qubit fluorometer plate reader off deck.

最終ステップ

Take forward the repaired and end repaired DNA into the adapter ligation and clean-up step.

5. Adapter ligation and clean-up

材料
  • Adapter Mix F (AMX F)
  • Ligation Buffer (LNB)
  • Long Fragment Buffer (LFB)
  • Short Fragment Buffer (SFB)
  • Elution Buffer (EB)

消耗品
  • Agencourt AMPure XP beads (Beckman Coulter™, A63881)
  • NEBNext Quick Ligation Module (NEB, E6056)
  • Thermo Scientific™ Abgene™ 96 Well 0.8 ml Polypropylene Deepwell Storage Plate (Thermo Scientific™, cat # AB0859)
  • Sarstedt Inc Screw Cap Micro tube 2 ml, PP 1000/case (e.g. FisherScientific, Cat# NC0418367)
  • Roche Diagnotics MagNA Pure LC Medium Reagent Tubs 20 (Cat# 03004058001)
  • Bio-Rad Hard-Shell® 96-Well PCR Plates (Cat# HSP9601)
  • Hamilton 60 ml Reagent Reservoir, Self-Standing with Lid (Cat# 56694-01)
  • Hamilton 1000 µl CO-RE tips with filter (Cat# 235905)
  • Hamilton 300 µl CO-RE tips with filter (Cat# 235903)
  • Hamilton 50 µl CO-RE tips with filter (Cat# 235948)

装置
  • P1000 pipette and tips
  • P200 pipette and tips
  • P100 pipette and tips
  • P10 pipette and tips
  • Vortex mixer
  • Microplate centrifuge, e.g. Fisherbrand™ Mini Plate Spinner Centrifuge (Fisher Scientific, 11766427)

Consumables and equipment quantities:

Consumable/equipment X24 samples X48 samples X96 samples
Hamilton 50 µl CO-RE tips with filter 120 240 384
Hamilton 300 µl CO-RE tips with filter 250 396 686
Hamilton 1000 µl CO-RE tips with filter 32 32 32
Hamilton 60 ml Reagent Reservoir, Self-Standing with Lid 2 (1 L/SFB & 1 EB) 2 (1 L/SFB & 1 EB) 2 (1 L/SFB & 1 EB)
Bio-Rad Hard-Shell® 96-Well PCR Plate 1 1 1
Hamilton 20 ml Reagent Reservoirs 1 1 1
Sarstedt Inc Screw Cap Micro Tube 2 ml 1 2 3
Abgene™ 96 Well 0.8 ml Polypropylene Deepwell Storage Plate 1 1 1

Reagents quantities:

Reagents X24 samples X48 samples X96 samples
Adapter Mix F (AMX-F) 0.5 tube 1 tube 2 tubes
Ligation Buffer (LNB) 0.5 tube 1 tube 2 tubes
Elution Buffer (EB) 0.5 bottle 1 bottle 2 bottles
Long Fragment Buffer (LFB) 2 bottles 4 bottles 8 bottles
Short Fragment Buffer (SFB) 2 bottles 4 bottles 8 bottles
AMPure XP Beads 3.1 ml 4.3 ml 6.6 ml
NEBNext Companion Module for Oxford Nanopore Technologies Ligation Sequencing (Cat# E7180S)
regarding the reagent below:
2 tubes 3 tubes 5 tubes
Alternatively: - - -
Quick T4 DNA Ligase 1 tube 2 tubes 3 tubes

Note: Dead volumes are included.

重要

Although the recommended 3rd party ligase is supplied with its own buffer, the ligation efficiency of Adapter Mix F (AMX-F) is higher when using Ligation Buffer supplied within the Ligation Sequencing Kit.

Spin down and store the Quick T4 Ligase on ice until use.

Spin down and combine all the required tubes of Adapter Mix F (AMX-F) required, and place on ice.

Thaw the Ligation Buffer (LNB) at room temperature, spin down and combine all the required tubes. Place on ice immediately after thawing and mixing.

Thaw a bottle of Elution Buffer (EB) at room temperature, mix by vortexing and place on ice.

重要

Depending on the wash buffer (LFB or SFB) used, the clean-up step after adapter ligation is designed to either enrich for DNA fragments of >3 kb, or purify all fragments equally.

  • To enrich for DNA fragments of 3 kb or longer, use Long Fragment Buffer (LFB)
  • To retain DNA fragments of all sizes, use Short Fragment Buffer (SFB)

To enrich for DNA fragments of 3 kb or longer, thaw the Long Fragment Buffer (LFB) at room temperature, mix by vortexing and combine all the required bottles before storing on ice.

To retain DNA fragments of all sizes, thaw the Short Fragment Buffer (SFB) at room temperature, mix by vortexing and combine all the required bottles before storing on ice.

Click 'Process03: Adapter ligation' to start.

Capture19

Click 'Process04: Adapter ligation and clean-up' to stop the automated library preparation and quantify the samples before sequencing.

Capture20

重要

It is mandatory for the MPH module to be installed on the liquid handling robot. Select 'Yes' to use the MPH (96 Head) module.

Capture21

Click 'Browse' to choose the Input File Worklist used during DNA repair and end-prep.

Capture4

Prepare the Adapter Ligation Mastermix with the following reagents in 2 ml Sarstedt tubes according to the Hamilton user interface. Click either 'Yes' or 'No' to continue.

Note: It is user preference whether to print and save the instructions.

Reagent volumes for all sample numbers:

Reagent Volume X24 samples Volume X48 samples Volume X96 samples
Adapter Mix F (AMX-F) 140.5 µl 281 µl 559.5 µl
Ligation Buffer (LNB) 702.5 µl 1405 µl 2797.5 µl
Quick T4 DNA Ligase 281 µl 562 µl 1119 µl

image (14)

注意

Ensure the mastermix is well mixed and homogenous before loading the 2 ml Sarstedt tubes. Mixing in the robot is not effective.

Insert plates to their corresponding positions on screen. Click 'Ok' to continue.

Capture24

Load a full deck of 50 µl tips into the positions on screen. Click 'Ok' to continue.

Capture25

Highlight the 50 µl tips available to use on the 'Edit Tip Count' window. Click 'Ok' to continue.

Capture26

Load a full deck of 300 µl tips in the positions on screen. Click 'Ok' to continue.

Capture27

Highlight the 300 µl tips available to use on the 'Edit Tip Count' window. Click 'Ok' to continue.

Capture28

Prepare the AMPure XP beads by vortexing and load the 20 ml trough with the volume required:

Reagents Volume X24 samples Volume X48 samples Volume X96 samples
Beads 3.1 ml 4.3 ml 6.6 ml
重要

Ensure the AMPure XP beads are well mixed before use by vortexing.

Prepare troughs of L/SFB and EB in troughs and insert in the positions on screen with the trough of AMPure Beads. Click 'Ok' to continue.

Reagent Volume X24 samples Volume X48 samples Volume X96 samples
Long/Short Fragment Buffer 2 bottles 4 bottles 8 bottles
Elution Buffer 0.5 bottle 1 bottle 2 bottles

Capture29

ヒント

If the troughs are not barcoded, click 'Exclude' on all the selected troughs inserted in the robot and click 'Execute' to continue.

Capture13

Insert 1000 µl tips and the Clean End Prep Plate to the correct positions on screen. Click 'Ok' to continue.

Capture31

Highlight the 1000 µl tips available to use on the 'Edit Tip Count' window. Click 'Ok' to continue.

Capture32

Insert the prepared Adapter Ligation Mastermix into the positions on screen. Click 'Ok' to continue.

Capture33

注意

Ensure the mastermix is well mixed and homogenous before loading the 2 ml Sarstedt tubes. Mixing in the robot is not effective.

Once the automation process has finished, there will be an on screen prompt to unload the plate. Click 'Ok' to continue.

Adapter ligation and clean up2

Quantify 1 µl of each eluted sample using a Qubit fluorometer plate reader off deck.

最終ステップ

Seal the plate once the library is prepared and store on ice until ready to load onto the flow cell.

We do not recommend running the liquid handling robot overnight as the plate must be sealed and stored on ice as soon as library preparation is finished.

ヒント

Library storage recommendations

We recommend storing libraries in Eppendorf DNA LoBind tubes at 4°C for short-term storage or repeated use, for example, re-loading flow cells between washes. For single use and long-term storage of more than 3 months, we recommend storing libraries at -80°C in Eppendorf DNA LoBind tubes.

重要

We recommend loading 5-50 fmol of the final prepared library onto a flow cell.

Loading more than the maximal recommended amount of DNA can have a detrimental effect on output as higher quantities of DNA results in a larger number of ligated DNA ends with loaded motor protein. This depletes fuel in the Sequencing Buffer, regardless of whether or not the DNA fragments are being sequenced. This leads to fuel depletion and speed drop-off early in the sequencing run. Dilute the libraries in Elution Buffer if required.

If you are using the Flongle for sample prep development, we recommend loading 3-20 fmol instead.

オプショナルステップ

If quantities allow, the libraries may be diluted in Elution Buffer (EB) for splitting across multiple flow cells.

Additional buffer for doing this can be found in the Sequencing Auxiliary Vials expansion (EXP-AUX002), available to purchase separately. This expansion also contains additional vials of Sequencing Buffer (SBII) and Loading Beads (LBII), required for loading the libraries onto flow cells.

6. Priming and loading multiple flow cells on a PromethION

材料
  • Flush Buffer (FB)
  • Flush Tether (FLT)
  • Sequencing Buffer II (SBII)
  • Loading Beads II (LBII)
  • Loading Solution (LS)

消耗品
  • PromethION Flow Cell
  • 1.5 ml Eppendorf DNA LoBind tubes
  • 2 ml Eppendorf DNA LoBind tubes

装置
  • PromethION 2 Solo device
  • PromethION sequencing device
  • P1000 pipette and tips
  • P200 pipette and tips
  • P20 pipette and tips

Using the Loading Solution

We recommend using the Loading Beads II (LBII) for loading your library onto the flow cell for most sequencing experiments. However, if you have previously used water to load your library, you must use Loading Solution (LS) instead of water. Note: some customers have noticed that viscous libraries can be loaded more easily when not using Loading Beads II.

Thaw the Sequencing Buffer II (SBII), Loading Beads II (LBII), Flush Tether (FLT) and on tube of Flush Buffer (FB) at room temperature before mixing the reagents by vortexing and spin down at room temperature.

重要

Scale up reagent volumes as needed.

Ensure to prepare enough reagents for the total number of flow cells being processed and to take into account extra volume required for pipetting errors.

ヒント

Each vial provides enough reagent for the preparation of 12 samples. Thaw the appropriate number of vials of each reagent.

Prepare the flow cell priming mix in a suitable vial for the number of flow cells to flush. Once combined, mix well by briefly vortexing.

Reagent Volume per flow cell
Flush Tether (FLT) 30 µl
Flush Buffer (FB) 1,170 µl
重要

After taking flow cells out of the fridge, wait 20 minutes before inserting the flow cell into the PromethION for the flow cell to come to room temperature. Condensation can form on the flow cell in humid environments. Inspect the gold connector pins on the top and underside of the flow cell for condensation and wipe off with a lint-free wipe if any is observed. Ensure the heat pad (black pad) is present on the underside of the flow cell.

For PromethION 2 Solo, load the flow cell(s) as follows:

  1. Place the flow cell flat on the metal plate.

  2. Slide the flow cell into the docking port until the gold pins or green board cannot be seen.

J2068 FC-into-P2-animation V5

For the PromethION 24/48, load the flow cell(s) into the docking ports:

  1. Line up the flow cell with the connector horizontally and vertically before smoothly inserting into position.
  2. Press down firmly onto the flow cell and ensure the latch engages and clicks into place.

Step 1a V3

Step 1B

重要

Insertion of the flow cells at the wrong angle can cause damage to the pins on the PromethION and affect your sequencing results. If you find the pins on a PromethION position are damaged, please contact support@nanoporetech.com for assistance.

Screenshot 2021-04-08 at 12.08.37

If not already completed, perform a flow cell check on all flow cells.

Please refer to the Flow Cell Check protocol for further information.

Slide the inlet port cover clockwise to open.

Prom loading 2

重要

Take care when drawing back buffer from the flow cell. Do not remove more than 20-30 µl, and make sure that the array of pores are covered by buffer at all times. Introducing air bubbles into the array can irreversibly damage pores.

After opening the inlet port, draw back a small volume to remove any air bubbles:

  1. Set a P1000 pipette tip to 200 µl.
  2. Insert the tip into the inlet port.
  3. Turn the wheel until the dial shows 220-230 µl, or until you see a small volume of buffer entering the pipette tip.

Step 3 v1

Load 500 µl of the priming mix into the flow cell via the inlet port, avoiding the introduction of air bubbles. Wait five minutes. During this time, prepare the library for loading using the next steps in the protocol.

Step 4 v1

Thoroughly mix the contents of the Sequencing buffer II (SBII) and Loading Beads II (LBII) tubes by vortexing.

重要

The Loading Beads II (LBII) tube contains a suspension of beads. These beads settle very quickly. It is vital that they are mixed immediately before use.

In a separate tube for each library, prepare for loading by adding the following reagents:

Reagent Volume
SBII 75 µl
LB 51 µl
DNA library 24 µl
Total 150 µl
オプショナルステップ

The Ligation Sequencing Kit XL kit is designed for users running multiple samples/flow cells. When handling multiple DNA preparations, the Sequencing Buffer (SBII) and Loading Beads (LBII) can be combined in a master mix:

Mix the Sequencing Buffer (SBII) and Loading Beads (LBII) as described above, scaling up the final volume for the appropriate number of samples and adding up to 20% excess of each reagent. Mix the master mix by pipetting immediately before adding to the DNA samples. Pipette 126 µl of the master mix into each DNA sample-containing tube. Mix the sample by pipetting.

Complete the flow cell priming by slowly loading 500 µl of the priming mix into the inlet port.

Step 5 v1

Mix the prepared library gently by pipetting up and down just prior to loading.

Using a P1000, insert the pipette tip into the inlet port and add 150 µl of library.

Step 6 v2

Close the valve to seal the inlet port and close the PromethION lid when ready.

Wait a minimum of 10 minutes after loading the flow cells onto the PromethION before initiating any experiments. This will help to increase the sequencing output.

Step 7 V2

For multiple flow cell washing, use the same experiment name and identifying sample IDs for all runs to enable all flow cells to be paused simultaneously.

Screenshot 2023-02-14 114901

7. Data acquisition and basecalling

How to start sequencing

Once you have loaded your flow cell, the sequencing run can be started on MinKNOW, our sequencing software that controls the device, data acquisition and real-time basecalling. For more detailed information on setting up and using MinKNOW, please see the MinKNOW protocol.

MinKNOW can be used and set up to sequence in multiple ways:

  • On a computer either direcly or remotely connected to a sequencing device.
  • Directly on a GridION, MinION Mk1C or PromethION 24/48 sequencing device.

For more information on using MinKNOW on a sequencing device, please see the device user manuals:


To start a sequencing run on MinKNOW:

1. Navigate to the start page and click Start sequencing.

2. Fill in your experiment details, such as name and flow cell position and sample ID.

3. Select the sequencing kit used in the library preparation on the Kit page.

4. Configure the sequencing and output parameters for your sequencing run or keep to the default settings on the Run configuration tab.

Note: If basecalling was turned off when a sequencing run was set up, basecalling can be performed post-run on MinKNOW. For more information, please see the MinKNOW protocol.

5. Click Start on the Review page to start the sequencing run.

Data analysis after sequencing

After sequencing has completed on MinKNOW, the flow cell can be reused or returned, as outlined in the Flow cell reuse and returns section.

After sequencing and basecalling, the data can be analysed. For further information about options for basecalling and post-basecalling analysis, please refer to the Data Analysis document.

In the Downstream analysis section, we outline further options for analysing your data.

8. Flow cell reuse and returns

材料
  • Flow Cell Wash Kit (EXP-WSH004)

After your sequencing experiment is complete, if you would like to reuse the flow cell, please follow the Flow Cell Wash Kit protocol and store the washed flow cell at 2-8°C.

The Flow Cell Wash Kit protocol is available on the Nanopore Community.

ヒント

We recommend you to wash the flow cell as soon as possible after you stop the run. However, if this is not possible, leave the flow cell on the device and wash it the next day.

Alternatively, follow the returns procedure to flush out the flow cell ready to send back to Oxford Nanopore.

Instructions for returning flow cells can be found here.

Note: All flow cells must be flushed with deionised water before returning the product.

重要

If you encounter issues or have questions about your sequencing experiment, please refer to the Troubleshooting Guide that can be found in the online version of this protocol.

9. Downstream analysis

Post-basecalling analysis

There are several options for further analysing your basecalled data:

1. EPI2ME workflows

For in-depth data analysis, Oxford Nanopore Technologies offers a range of bioinformatics tutorials and workflows available in EPI2ME. The platform provides a vehicle where workflows deposited in GitHub by our Research and Applications teams can be showcased with descriptive texts, functional bioinformatics code and example data.

2. Research analysis tools

Oxford Nanopore Technologies' Research division has created a number of analysis tools, which are available in the Oxford Nanopore GitHub repository. The tools are aimed at advanced users, and contain instructions for how to install and run the software. They are provided as-is, with minimal support.

3. Community-developed analysis tools

If a data analysis method for your research question is not provided in any of the resources above, please refer to the Bioinformatics section of the Resource centre. Numerous members of the Nanopore Community have developed their own tools and pipelines for analysing nanopore sequencing data, most of which are available on GitHub. Please be aware that these tools are not supported by Oxford Nanopore Technologies, and are not guaranteed to be compatible with the latest chemistry/software configuration.

10. Issues during automation of library preparation

Please contact your automation vendor FAS and/or Nanopore FAS if you have any issues.

11. Issues during the sequencing run

Below is a list of the most commonly encountered issues, with some suggested causes and solutions.

We also have an FAQ section available on the Nanopore Community Support section.

If you have tried our suggested solutions and the issue still persists, please contact Technical Support via email (support@nanoporetech.com) or via LiveChat in the Nanopore Community.

Fewer pores at the start of sequencing than after Flow Cell Check

Observation Possible cause Comments and actions
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check An air bubble was introduced into the nanopore array After the Flow Cell Check it is essential to remove any air bubbles near the priming port before priming the flow cell. If not removed, the air bubble can travel to the nanopore array and irreversibly damage the nanopores that have been exposed to air. The best practice to prevent this from happening is demonstrated in this video.
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check The flow cell is not correctly inserted into the device Stop the sequencing run, remove the flow cell from the sequencing device and insert it again, checking that the flow cell is firmly seated in the device and that it has reached the target temperature. If applicable, try a different position on the device (GridION/PromethION).
MinKNOW reported a lower number of pores at the start of sequencing than the number reported by the Flow Cell Check Contaminations in the library damaged or blocked the pores The pore count during the Flow Cell Check is performed using the QC DNA molecules present in the flow cell storage buffer. At the start of sequencing, the library itself is used to estimate the number of active pores. Because of this, variability of about 10% in the number of pores is expected. A significantly lower pore count reported at the start of sequencing can be due to contaminants in the library that have damaged the membranes or blocked the pores. Alternative DNA/RNA extraction or purification methods may be needed to improve the purity of the input material. The effects of contaminants are shown in the Contaminants Know-how piece. Please try an alternative extraction method that does not result in contaminant carryover.

MinKNOW script failed

Observation Possible cause Comments and actions
MinKNOW shows "Script failed"
Restart the computer and then restart MinKNOW. If the issue persists, please collect the MinKNOW log files and contact Technical Support. If you do not have another sequencing device available, we recommend storing the flow cell and the loaded library at 4°C and contact Technical Support for further storage guidance.

Pore occupancy below 40%

Observation Possible cause Comments and actions
Pore occupancy <40% Not enough library was loaded on the flow cell Ensure you load the recommended amount of good quality library in the relevant library prep protocol onto your flow cell. Please quantify the library before loading and calculate mols using tools like the Promega Biomath Calculator, choosing "dsDNA: µg to pmol"
Pore occupancy close to 0 The Ligation Sequencing Kit was used, and sequencing adapters did not ligate to the DNA Make sure to use the NEBNext Quick Ligation Module (E6056) and Oxford Nanopore Technologies Ligation Buffer (LNB, provided in the sequencing kit) at the sequencing adapter ligation step, and use the correct amount of each reagent. A Lambda control library can be prepared to test the integrity of the third-party reagents.
Pore occupancy close to 0 The Ligation Sequencing Kit was used, and ethanol was used instead of LFB or SFB at the wash step after sequencing adapter ligation Ethanol can denature the motor protein on the sequencing adapters. Make sure the LFB or SFB buffer was used after ligation of sequencing adapters.
Pore occupancy close to 0 No tether on the flow cell Tethers are adding during flow cell priming (FLT/FCT tube). Make sure FLT/FCT was added to FB/FCF before priming.

Shorter than expected read length

Observation Possible cause Comments and actions
Shorter than expected read length Unwanted fragmentation of DNA sample Read length reflects input DNA fragment length. Input DNA can be fragmented during extraction and library prep.

1. Please review the Extraction Methods in the Nanopore Community for best practice for extraction.

2. Visualise the input DNA fragment length distribution on an agarose gel before proceeding to the library prep. DNA gel2 In the image above, Sample 1 is of high molecular weight, whereas Sample 2 has been fragmented.

3. During library prep, avoid pipetting and vortexing when mixing reagents. Flicking or inverting the tube is sufficient.

Large proportion of unavailable pores

Observation Possible cause Comments and actions
Large proportion of unavailable pores (shown as blue in the channels panel and pore activity plot)

image2022-3-25 10-43-25 The pore activity plot above shows an increasing proportion of "unavailable" pores over time.
Contaminants are present in the sample Some contaminants can be cleared from the pores by the unblocking function built into MinKNOW. If this is successful, the pore status will change to "sequencing pore". If the portion of unavailable pores stays large or increases:

1. A nuclease flush using the Flow Cell Wash Kit (EXP-WSH004) can be performed, or
2. Run several cycles of PCR to try and dilute any contaminants that may be causing problems.

Large proportion of inactive pores

Observation Possible cause Comments and actions
Large proportion of inactive/unavailable pores (shown as light blue in the channels panel and pore activity plot. Pores or membranes are irreversibly damaged) Air bubbles have been introduced into the flow cell Air bubbles introduced through flow cell priming and library loading can irreversibly damage the pores. Watch the Priming and loading your flow cell video for best practice
Large proportion of inactive/unavailable pores Certain compounds co-purified with DNA Known compounds, include polysaccharides, typically associate with plant genomic DNA.

1. Please refer to the Plant leaf DNA extraction method.
2. Clean-up using the QIAGEN PowerClean Pro kit.
3. Perform a whole genome amplification with the original gDNA sample using the QIAGEN REPLI-g kit.
Large proportion of inactive/unavailable pores Contaminants are present in the sample The effects of contaminants are shown in the Contaminants Know-how piece. Please try an alternative extraction method that does not result in contaminant carryover.

Reduction in sequencing speed and q-score later into the run

Observation Possible cause Comments and actions
Reduction in sequencing speed and q-score later into the run For Kit 9 chemistry (e.g. SQK-LSK109), fast fuel consumption is typically seen when the flow cell is overloaded with library (please see the appropriate protocol for your DNA library to see the recommendation). Add more fuel to the flow cell by following the instructions in the MinKNOW protocol. In future experiments, load lower amounts of library to the flow cell.

Temperature fluctuation

Observation Possible cause Comments and actions
Temperature fluctuation The flow cell has lost contact with the device Check that there is a heat pad covering the metal plate on the back of the flow cell. Re-insert the flow cell and press it down to make sure the connector pins are firmly in contact with the device. If the problem persists, please contact Technical Services.

Failed to reach target temperature

Observation Possible cause Comments and actions
MinKNOW shows "Failed to reach target temperature" The instrument was placed in a location that is colder than normal room temperature, or a location with poor ventilation (which leads to the flow cells overheating) MinKNOW has a default timeframe for the flow cell to reach the target temperature. Once the timeframe is exceeded, an error message will appear and the sequencing experiment will continue. However, sequencing at an incorrect temperature may lead to a decrease in throughput and lower q-scores. Please adjust the location of the sequencing device to ensure that it is placed at room temperature with good ventilation, then re-start the process in MinKNOW. Please refer to this FAQ for more information on MinION Mk 1B temperature control.

Guppy – no input .fast5 was found or basecalled

Observation Possible cause Comments and actions
No input .fast5 was found or basecalled input_path did not point to the .fast5 file location The --input_path has to be followed by the full file path to the .fast5 files to be basecalled, and the location has to be accessible either locally or remotely through SSH.
No input .fast5 was found or basecalled The .fast5 files were in a subfolder at the input_path location To allow Guppy to look into subfolders, add the --recursive flag to the command

Guppy – no Pass or Fail folders were generated after basecalling

Observation Possible cause Comments and actions
No Pass or Fail folders were generated after basecalling The --qscore_filtering flag was not included in the command The --qscore_filtering flag enables filtering of reads into Pass and Fail folders inside the output folder, based on their strand q-score. When performing live basecalling in MinKNOW, a q-score of 7 (corresponding to a basecall accuracy of ~80%) is used to separate reads into Pass and Fail folders.

Guppy – unusually slow processing on a GPU computer

Observation Possible cause Comments and actions
Unusually slow processing on a GPU computer The --device flag wasn't included in the command The --device flag specifies a GPU device to use for accelerate basecalling. If not included in the command, GPU will not be used. GPUs are counted from zero. An example is --device cuda:0 cuda:1, when 2 GPUs are specified to use by the Guppy command.

Last updated: 10/23/2023